Noise performance of textile‐based dry ECG recording electrodes

Abstract This study reports attempts to characterise the noise properties of un‐gelled, textile‐based electrodes intended for use in ECG recording in wearable garments. Measurements were made using very low noise operational amplifiers and with mains power supply interference eliminated from the rec...

Full description

Saved in:
Bibliographic Details
Main Authors: Soumyajyoti Maji, Martin J. Burke
Format: article
Language:EN
Published: Wiley 2021
Subjects:
Online Access:https://doaj.org/article/07fc01a4e09c4425bbca925506ba4649
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Abstract This study reports attempts to characterise the noise properties of un‐gelled, textile‐based electrodes intended for use in ECG recording in wearable garments. Measurements were made using very low noise operational amplifiers and with mains power supply interference eliminated from the recorded signals so that they reflected only genuine noise. Measurements were carried out on four conductive fabric electrodes, one conductive rubber electrode and a standard self‐adhesive, gelled electrode on three male and three female subjects. Curves of the form y = K/f + C were fitted to the recorded noise spectral density functions. Coefficients obtained for the white noise spectral density ranged from 1.1 to 5.6 nV/√Hz and for the flicker (1/f) noise spectral density ranged from 8 to 80 nV√Hz.