Impact of image averaging on vessel detection using optical coherence tomography angiography in eyes with macular oedema and in healthy eyes
<h4>Purpose</h4> To assess the repeatability of multiple automatic vessel density (VD) measurements and the effect of image averaging on vessel detection by optical coherence tomography angiography (OCTA). <h4>Methods</h4> An observational study was conducted in a series of h...
Guardado en:
Autores principales: | , , , , , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Public Library of Science (PLoS)
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/0fcdd20f73904a9f940ce9eca0e6537d |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
Sumario: | <h4>Purpose</h4> To assess the repeatability of multiple automatic vessel density (VD) measurements and the effect of image averaging on vessel detection by optical coherence tomography angiography (OCTA). <h4>Methods</h4> An observational study was conducted in a series of healthy volunteers and patients with macular oedema. Five sequential OCTA images were acquired for each eye using the OptoVue HD device. The effect of the averaging of the 5 acquisitions on vessel detection was analysed quantitatively using a pixel-by-pixel automated analysis. In addition, two independent retina experts qualitatively assessed the change in vessel detection in averaged images segmented in 9 boxes and compared to the first non-averaged image. <h4>Results</h4> The automatic VD measurement in OCTA images showed a good repeatability with an overall mean intra-class correlation coefficient (ICC) of 0.924. The mean ICC was higher in healthy eyes compared to eyes with macular oedema (0.877 versus 0.960; p < 0.001) and in the superficial vascular plexus versus the deep vascular complex (0.967 versus 0.888; p = 0.001). The quantitative analysis of the effect of the averaging showed that averaged images had a mean gain of 790.4 pixels/box, located around or completing interruptions in the vessel walls, and a mean loss of 727.2 pixels/box. The qualitative analysis of the averaged images showed that 99.6% of boxes in the averaged images had a gain in vessel detection (i.e., vessels detected in the averaged image but not in the non-averaged image). The loss of pixels was due to a reduction in background noise and motion artifacts in all cases and no case of loss of vessel detection was observed. <h4>Conclusion</h4> The automatic VD measurement using the OptoVue HD device showed a good repeatability in 5 acquisitions in a row setting. Averaging images increased vessel detection, and in about a third of boxes, decreased the background noise, both in healthy eyes and, in a greater proportion, in eyes with macular oedema. |
---|