Usefulness of semi-automatic harmonization strategy of standardized uptake values for multicenter PET studies

Abstract This study assessed the possibility of semi-automatic harmonization of standardized uptake values (SUVs) in multicenter studies. Phantom data were acquired using 16 PET/CT scanners (including 3 PET/CT scanners with a silicon photomultiplier detector). PET images obtained using 30-min/bed sc...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Hiromitsu Daisaki, Kazuhiro Kitajima, Masatoyo Nakajo, Tadashi Watabe, Kimiteru Ito, Fumi Sakamoto, Tadaki Nakahara, Mana Ishibashi, Akira Toriihara
Formato: article
Lenguaje:EN
Publicado: Nature Portfolio 2021
Materias:
R
Q
Acceso en línea:https://doaj.org/article/19b1de834b034dc791e419a48eca77e7
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
Descripción
Sumario:Abstract This study assessed the possibility of semi-automatic harmonization of standardized uptake values (SUVs) in multicenter studies. Phantom data were acquired using 16 PET/CT scanners (including 3 PET/CT scanners with a silicon photomultiplier detector). PET images obtained using 30-min/bed scans for optimum harmonization filter calculations and using 90–180-s/bed scans for SUV validation under clinical conditions were obtained. Time of flight and a reconstruction method with point-spread function correction were allowed. The optimal full width at half maximum of the 3D-Gaussian filter that minimizes the root mean square error with the median value of the JSNM harmonization range was calculated semi-automatically. The SUVmax and the SUVpeak of the hot spheres were measured, and the inter-scanner coefficient of variation (COV) was calculated before and after harmonization. The harmonization filter was applied to 11 of the 15 PET/CT scanners in which the SUV calibration accuracy had been verified, but not in the remaining 4 scanners. Under noiseless conditions before harmonization, the inter-scanner COVs of the SUVmax and the SUVpeak were as high as 21.57% and 12.20%, respectively, decreasing to 8.79% and 5.73% after harmonization, respectively. Harmonization brought the SUVmax of all the hot spheres to within the harmonization range. Even under clinical conditions affected by image noise, the inter-scanner COVs for the SUVmax and SUVpeak were as high as 8.83% and 5.18% after harmonization, respectively. By applying an optimal harmonization filter that is calculated semi-automatically, the harmonization of SUVs according to the JSNM strategy is possible in multicenter studies, thereby reducing inter-scanner COVs.