A novel adhesive biosensor system for detecting respiration, cardiac, and limb movement signals during sleep: validation with polysomnography

Elise Jortberg,1 Ikaro Silva,1 Viprali Bhatkar,1 Ryan S McGinnis,2 Ellora Sen-Gupta,1 Briana Morey,1 John A Wright Jr,1 Jesus Pindado,1 Matt T Bianchi3 1MC10, Inc., Lexington, MA 02421, USA; 2Department of Electrical and Biomedical Engineering, University of Vermont, Burlington, VT 05405, USA; 3Neur...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Jortberg E, Silva I, Bhatkar V, McGinnis RS, Sen-Gupta E, Morey B, Wright Jr JA, Pindado J, Bianchi MT
Formato: article
Lenguaje:EN
Publicado: Dove Medical Press 2018
Materias:
Acceso en línea:https://doaj.org/article/1ddc096eb0f34e6f96c002832f130b35
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:1ddc096eb0f34e6f96c002832f130b35
record_format dspace
spelling oai:doaj.org-article:1ddc096eb0f34e6f96c002832f130b352021-12-02T04:03:16ZA novel adhesive biosensor system for detecting respiration, cardiac, and limb movement signals during sleep: validation with polysomnography1179-1608https://doaj.org/article/1ddc096eb0f34e6f96c002832f130b352018-11-01T00:00:00Zhttps://www.dovepress.com/a-novel-adhesive-biosensor-system-for-detecting-respiration-cardiac-an-peer-reviewed-article-NSShttps://doaj.org/toc/1179-1608Elise Jortberg,1 Ikaro Silva,1 Viprali Bhatkar,1 Ryan S McGinnis,2 Ellora Sen-Gupta,1 Briana Morey,1 John A Wright Jr,1 Jesus Pindado,1 Matt T Bianchi3 1MC10, Inc., Lexington, MA 02421, USA; 2Department of Electrical and Biomedical Engineering, University of Vermont, Burlington, VT 05405, USA; 3Neurology Department, Massachusetts General Hospital, Boston, MA 02114, USA Background: Although in-lab polysomnography (PSG) remains the gold standard for objective sleep monitoring, the use of at-home sensor systems has gained popularity in recent years. Two categories of monitoring, autonomic and limb movement physiology, are increasingly recognized as critical for sleep disorder phenotyping, yet at-home options remain limited outside of research protocols. The purpose of this study was to validate the BiostampRC® sensor system for respiration, electrocardiography (ECG), and leg electromyography (EMG) against gold standard PSG recordings.Methods: We report analysis of cardiac and respiratory data from 15 patients and anterior tibialis (AT) data from 19 patients undergoing clinical PSG for any indication who simultaneously wore BiostampRC® sensors on the chest and the bilateral AT muscles. BiostampRC® is a flexible, adhesive, wireless sensor capable of capturing accelerometry, ECG, and EMG. We compared BiostampRC® data and feature extractions with those obtained from PSG.Results: The heart rate extracted from BiostampRC® ECG showed strong agreement with the PSG (cohort root mean square error of 5 beats per minute). We found the thoracic BiostampRC® respiratory waveform, derived from its accelerometer, accurately calculated the respiratory rate (mean average error of 0.26 and root mean square error of 1.84 breaths per minute). The AT EMG signal supported periodic limb movement detection, with area under the curve of the receiver operating characteristic curve equaling 0.88. Upon completion, 88% of subjects indicated willingness to wear BiostampRC® at home on an exit survey.Conclusion: The results demonstrate that BiostampRC® is a tolerable and accurate method for capturing respiration, ECG, and AT EMG time series signals during overnight sleep when compared with simultaneous PSG recordings. The signal quality sufficiently supports analytics of clinical relevance. Larger longitudinal in-home studies are required to support the role of BiostampRC® in clinical management of sleep disorders involving the autonomic nervous system and limb movements. Keywords: electrocardiography, electromyography, respiration, wearableJortberg ESilva IBhatkar VMcGinnis RSSen-Gupta EMorey BWright Jr JAPindado JBianchi MTDove Medical PressarticleelectrocardiographyelectromyographyrespirationwearablePsychiatryRC435-571Neurophysiology and neuropsychologyQP351-495ENNature and Science of Sleep, Vol Volume 10, Pp 397-408 (2018)
institution DOAJ
collection DOAJ
language EN
topic electrocardiography
electromyography
respiration
wearable
Psychiatry
RC435-571
Neurophysiology and neuropsychology
QP351-495
spellingShingle electrocardiography
electromyography
respiration
wearable
Psychiatry
RC435-571
Neurophysiology and neuropsychology
QP351-495
Jortberg E
Silva I
Bhatkar V
McGinnis RS
Sen-Gupta E
Morey B
Wright Jr JA
Pindado J
Bianchi MT
A novel adhesive biosensor system for detecting respiration, cardiac, and limb movement signals during sleep: validation with polysomnography
description Elise Jortberg,1 Ikaro Silva,1 Viprali Bhatkar,1 Ryan S McGinnis,2 Ellora Sen-Gupta,1 Briana Morey,1 John A Wright Jr,1 Jesus Pindado,1 Matt T Bianchi3 1MC10, Inc., Lexington, MA 02421, USA; 2Department of Electrical and Biomedical Engineering, University of Vermont, Burlington, VT 05405, USA; 3Neurology Department, Massachusetts General Hospital, Boston, MA 02114, USA Background: Although in-lab polysomnography (PSG) remains the gold standard for objective sleep monitoring, the use of at-home sensor systems has gained popularity in recent years. Two categories of monitoring, autonomic and limb movement physiology, are increasingly recognized as critical for sleep disorder phenotyping, yet at-home options remain limited outside of research protocols. The purpose of this study was to validate the BiostampRC® sensor system for respiration, electrocardiography (ECG), and leg electromyography (EMG) against gold standard PSG recordings.Methods: We report analysis of cardiac and respiratory data from 15 patients and anterior tibialis (AT) data from 19 patients undergoing clinical PSG for any indication who simultaneously wore BiostampRC® sensors on the chest and the bilateral AT muscles. BiostampRC® is a flexible, adhesive, wireless sensor capable of capturing accelerometry, ECG, and EMG. We compared BiostampRC® data and feature extractions with those obtained from PSG.Results: The heart rate extracted from BiostampRC® ECG showed strong agreement with the PSG (cohort root mean square error of 5 beats per minute). We found the thoracic BiostampRC® respiratory waveform, derived from its accelerometer, accurately calculated the respiratory rate (mean average error of 0.26 and root mean square error of 1.84 breaths per minute). The AT EMG signal supported periodic limb movement detection, with area under the curve of the receiver operating characteristic curve equaling 0.88. Upon completion, 88% of subjects indicated willingness to wear BiostampRC® at home on an exit survey.Conclusion: The results demonstrate that BiostampRC® is a tolerable and accurate method for capturing respiration, ECG, and AT EMG time series signals during overnight sleep when compared with simultaneous PSG recordings. The signal quality sufficiently supports analytics of clinical relevance. Larger longitudinal in-home studies are required to support the role of BiostampRC® in clinical management of sleep disorders involving the autonomic nervous system and limb movements. Keywords: electrocardiography, electromyography, respiration, wearable
format article
author Jortberg E
Silva I
Bhatkar V
McGinnis RS
Sen-Gupta E
Morey B
Wright Jr JA
Pindado J
Bianchi MT
author_facet Jortberg E
Silva I
Bhatkar V
McGinnis RS
Sen-Gupta E
Morey B
Wright Jr JA
Pindado J
Bianchi MT
author_sort Jortberg E
title A novel adhesive biosensor system for detecting respiration, cardiac, and limb movement signals during sleep: validation with polysomnography
title_short A novel adhesive biosensor system for detecting respiration, cardiac, and limb movement signals during sleep: validation with polysomnography
title_full A novel adhesive biosensor system for detecting respiration, cardiac, and limb movement signals during sleep: validation with polysomnography
title_fullStr A novel adhesive biosensor system for detecting respiration, cardiac, and limb movement signals during sleep: validation with polysomnography
title_full_unstemmed A novel adhesive biosensor system for detecting respiration, cardiac, and limb movement signals during sleep: validation with polysomnography
title_sort novel adhesive biosensor system for detecting respiration, cardiac, and limb movement signals during sleep: validation with polysomnography
publisher Dove Medical Press
publishDate 2018
url https://doaj.org/article/1ddc096eb0f34e6f96c002832f130b35
work_keys_str_mv AT jortberge anoveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT silvai anoveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT bhatkarv anoveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT mcginnisrs anoveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT senguptae anoveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT moreyb anoveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT wrightjrja anoveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT pindadoj anoveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT bianchimt anoveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT jortberge noveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT silvai noveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT bhatkarv noveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT mcginnisrs noveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT senguptae noveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT moreyb noveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT wrightjrja noveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT pindadoj noveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
AT bianchimt noveladhesivebiosensorsystemfordetectingrespirationcardiacandlimbmovementsignalsduringsleepvalidationwithpolysomnography
_version_ 1718401417356509184