An alternative way to measure total lung capacity: a pilot study
Total lung capacity is usually measured by a body plethysmograph or helium dilution methods. In this study an alternative approach to obtain total lung capacity of spontaneous breathing subjects is introduced. The device utilises an optoelectronic plethysmograph and a small tube, which measures mout...
Guardado en:
Autores principales: | , , , , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
De Gruyter
2020
|
Materias: | |
Acceso en línea: | https://doaj.org/article/1e7f9c906147416cb0c38220310063fb |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
Sumario: | Total lung capacity is usually measured by a body plethysmograph or helium dilution methods. In this study an alternative approach to obtain total lung capacity of spontaneous breathing subjects is introduced. The device utilises an optoelectronic plethysmograph and a small tube, which measures mouth pressure and allows total lung capacity to be obtained, which differs less than 0.4 L from the total lung capacity of the body plethysmograph. The method shows potential to be a less burdensome method to estimate total lung capacity determination than the body plethysmograph. |
---|