A constant-force technique to measure corneal biomechanical changes after collagen cross-linking.

<h4>Purpose</h4>To introduce a constant-force technique for the analysis of corneal biomechanical changes induced after collagen cross-linking (CXL) that is better adapted to the natural loading in the eye than previous methods.<h4>Methods</h4>For the biomechanical testing, a...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Olivier Richoz, Sabine Kling, Souska Zandi, Arthur Hammer, Eberhard Spoerl, Farhad Hafezi
Formato: article
Lenguaje:EN
Publicado: Public Library of Science (PLoS) 2014
Materias:
R
Q
Acceso en línea:https://doaj.org/article/20f32e8431464bb5a45253dd68612ea6
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
Descripción
Sumario:<h4>Purpose</h4>To introduce a constant-force technique for the analysis of corneal biomechanical changes induced after collagen cross-linking (CXL) that is better adapted to the natural loading in the eye than previous methods.<h4>Methods</h4>For the biomechanical testing, a total of 50 freshly enucleated eyes were obtained and subdivided in groups of 5 eyes each. A Zwicki-Line Testing Machine was used to analyze the strain of 11 mm long and 5 mm wide porcine corneal strips, with and without CXL. Before material testing, the corneal tissues were pre-stressed with 0.02 N until force stabilization. Standard strip extensiometry was performed as control technique. For the constant-force technique, tissue elongation (Δ strain, %) was analyzed for 180 seconds while different constant forces (0.25 N, 0.5 N, 1 N, 5 N) were applied.<h4>Results</h4>Using a constant force of 0.5 N, we observed a significant difference in Δstrain between 0.26±0.01% in controls and 0.12±0.03% in the CXL-treated group (p = 0.003) over baseline. Similarly, using a constant force of 1 N, Δstrain was 0.31±0.03% in controls and 0.19±0.02% after CXL treatment (p = 0.008). No significant differences were observed between CXL-treated groups and controls with 0.25 N or 5 N constant forces. Standard stress-strain extensiometry failed to show significant differences between CXL-treated groups and controls at all percentages of strains tested.<h4>Conclusion</h4>We propose a constant-force technique to measure corneal biomechanics in a more physiologic way. When compared to standard stress-strain extensiometry, the constant-force technique provides less variability and thus reaches significant results with a lower sample number.