Testing positive pressure delivered from commercial and WHO-style pediatric bubble CPAP devices

Abstract Background/aim Low-cost commercial bCPAP devices have been deployed in resource-limited settings to treat neonatal respiratory failure. The use of these devices has increased access to pediatric respiratory support for infants. However, constrained resources may result in substitution of re...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Nicholas A. Ettinger, Nathan Serazin, Richard Nguyen, Jennifer Werdenberg, Minke Huibers, Susan Torrey
Formato: article
Lenguaje:EN
Publicado: BMC 2021
Materias:
Acceso en línea:https://doaj.org/article/21dfb57dd42f4e75b7ad6cd18fb822c6
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:21dfb57dd42f4e75b7ad6cd18fb822c6
record_format dspace
spelling oai:doaj.org-article:21dfb57dd42f4e75b7ad6cd18fb822c62021-11-28T12:30:15ZTesting positive pressure delivered from commercial and WHO-style pediatric bubble CPAP devices10.1186/s12887-021-03006-21471-2431https://doaj.org/article/21dfb57dd42f4e75b7ad6cd18fb822c62021-11-01T00:00:00Zhttps://doi.org/10.1186/s12887-021-03006-2https://doaj.org/toc/1471-2431Abstract Background/aim Low-cost commercial bCPAP devices have been deployed in resource-limited settings to treat neonatal respiratory failure. The use of these devices has increased access to pediatric respiratory support for infants. However, constrained resources may result in substitution of recommended consumables and/or use in older age groups. We hypothesized that commercially available bCPAP devices, the standard WHO-style device and various improvised adaptations would all generate effective, safe positive pressure at the patient interface. Methods Performance of 2 commercially available bCPAP devices was tested against the standard WHO-style bCPAP device, as well as several improvised modifications of these devices, by measuring positive pressure delivered at the patient interface. Variables tested included different flow rates, patient interfaces and respiratory circuit tubing. Results Both commercial devices utilized according to manufacturer recommendations generated the expected positive pressure at the patient interface. When testing the recommended WHO-style bCPAP device with recommended materials as well as other improvised modifications, we found variable and potentially unpredictable generation of positive pressure at the patient interface. Conclusions Modified or improvised bCPAP devices should be used with extreme caution as the support provided may be more or less than expected depending on respiratory tubing and flow rates employed. Our data support the effectiveness of bCPAP in newborns and young infants. But, to our knowledge, there are no bCPAP patient interfaces for older children effective with low liter flow devices. Therefore, based on these results, we recommend against using WHO-style bCPAP devices for non-infant patients with respiratory failure and instead recommend using standard oxygen therapy with nasal cannulae or face-masks, as well as early consideration of transfer to a higher level of care.Nicholas A. EttingerNathan SerazinRichard NguyenJennifer WerdenbergMinke HuibersSusan TorreyBMCarticlePediatricsBubble CPAPCritical careGlobal healthResource limited setting; respiratory failurePediatricsRJ1-570ENBMC Pediatrics, Vol 21, Iss 1, Pp 1-10 (2021)
institution DOAJ
collection DOAJ
language EN
topic Pediatrics
Bubble CPAP
Critical care
Global health
Resource limited setting; respiratory failure
Pediatrics
RJ1-570
spellingShingle Pediatrics
Bubble CPAP
Critical care
Global health
Resource limited setting; respiratory failure
Pediatrics
RJ1-570
Nicholas A. Ettinger
Nathan Serazin
Richard Nguyen
Jennifer Werdenberg
Minke Huibers
Susan Torrey
Testing positive pressure delivered from commercial and WHO-style pediatric bubble CPAP devices
description Abstract Background/aim Low-cost commercial bCPAP devices have been deployed in resource-limited settings to treat neonatal respiratory failure. The use of these devices has increased access to pediatric respiratory support for infants. However, constrained resources may result in substitution of recommended consumables and/or use in older age groups. We hypothesized that commercially available bCPAP devices, the standard WHO-style device and various improvised adaptations would all generate effective, safe positive pressure at the patient interface. Methods Performance of 2 commercially available bCPAP devices was tested against the standard WHO-style bCPAP device, as well as several improvised modifications of these devices, by measuring positive pressure delivered at the patient interface. Variables tested included different flow rates, patient interfaces and respiratory circuit tubing. Results Both commercial devices utilized according to manufacturer recommendations generated the expected positive pressure at the patient interface. When testing the recommended WHO-style bCPAP device with recommended materials as well as other improvised modifications, we found variable and potentially unpredictable generation of positive pressure at the patient interface. Conclusions Modified or improvised bCPAP devices should be used with extreme caution as the support provided may be more or less than expected depending on respiratory tubing and flow rates employed. Our data support the effectiveness of bCPAP in newborns and young infants. But, to our knowledge, there are no bCPAP patient interfaces for older children effective with low liter flow devices. Therefore, based on these results, we recommend against using WHO-style bCPAP devices for non-infant patients with respiratory failure and instead recommend using standard oxygen therapy with nasal cannulae or face-masks, as well as early consideration of transfer to a higher level of care.
format article
author Nicholas A. Ettinger
Nathan Serazin
Richard Nguyen
Jennifer Werdenberg
Minke Huibers
Susan Torrey
author_facet Nicholas A. Ettinger
Nathan Serazin
Richard Nguyen
Jennifer Werdenberg
Minke Huibers
Susan Torrey
author_sort Nicholas A. Ettinger
title Testing positive pressure delivered from commercial and WHO-style pediatric bubble CPAP devices
title_short Testing positive pressure delivered from commercial and WHO-style pediatric bubble CPAP devices
title_full Testing positive pressure delivered from commercial and WHO-style pediatric bubble CPAP devices
title_fullStr Testing positive pressure delivered from commercial and WHO-style pediatric bubble CPAP devices
title_full_unstemmed Testing positive pressure delivered from commercial and WHO-style pediatric bubble CPAP devices
title_sort testing positive pressure delivered from commercial and who-style pediatric bubble cpap devices
publisher BMC
publishDate 2021
url https://doaj.org/article/21dfb57dd42f4e75b7ad6cd18fb822c6
work_keys_str_mv AT nicholasaettinger testingpositivepressuredeliveredfromcommercialandwhostylepediatricbubblecpapdevices
AT nathanserazin testingpositivepressuredeliveredfromcommercialandwhostylepediatricbubblecpapdevices
AT richardnguyen testingpositivepressuredeliveredfromcommercialandwhostylepediatricbubblecpapdevices
AT jenniferwerdenberg testingpositivepressuredeliveredfromcommercialandwhostylepediatricbubblecpapdevices
AT minkehuibers testingpositivepressuredeliveredfromcommercialandwhostylepediatricbubblecpapdevices
AT susantorrey testingpositivepressuredeliveredfromcommercialandwhostylepediatricbubblecpapdevices
_version_ 1718407939543269376