Validation and Quantification of Domperidone in Spiked Plasma Matrix Using Reversed Phase HPLC-UV Method

Pharmacokinetics studies of domperidone generally analyze plasma matrix samples. The present work aimed to develop and validate a rapid and simple reversed phase-HPLC method for quantifying domperidone in plasma matrices. The chromatographic method implemented: 1. Luna Phenomenex® C18 (250 mm × 4.6...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Sekar Ayu Pawestri, Akhmad Kharis Nugroho, Endang Lukitaningsih, Purwantiningsih Purwantiningsih
Formato: article
Lenguaje:EN
Publicado: Department of Chemistry, Universitas Gadjah Mada 2021
Materias:
Acceso en línea:https://doaj.org/article/2389c520e6f04a299f8a3bf9d0519868
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:2389c520e6f04a299f8a3bf9d0519868
record_format dspace
spelling oai:doaj.org-article:2389c520e6f04a299f8a3bf9d05198682021-12-02T18:02:38ZValidation and Quantification of Domperidone in Spiked Plasma Matrix Using Reversed Phase HPLC-UV Method1411-94202460-157810.22146/ijc.64253https://doaj.org/article/2389c520e6f04a299f8a3bf9d05198682021-08-01T00:00:00Zhttps://jurnal.ugm.ac.id/ijc/article/view/64253https://doaj.org/toc/1411-9420https://doaj.org/toc/2460-1578Pharmacokinetics studies of domperidone generally analyze plasma matrix samples. The present work aimed to develop and validate a rapid and simple reversed phase-HPLC method for quantifying domperidone in plasma matrices. The chromatographic method implemented: 1. Luna Phenomenex® C18 (250 mm × 4.6 mm i.d; 5 µm) column, 2. isocratic mobile phase mixture of phosphate buffer 0.02 M:acetonitrile (70:30, v/v) with a flow rate of 1 mL/min, 3. UV detection at 285 nm. Domperidone and propranolol hydrochloride (as internal standard) were extracted from the deproteinated plasma sample. The method linearity was 0.998 in the range concentration of 15–200 ng/mL. The percentage of accuracy error was between -8.49–4.31%, while the percentage coefficient variation of precision ranged between 5.11–14.24%. This proposed method was simple, rapid (separation time less than 10 min), and selective. The validation parameters responses satisfied the method's requirements to determine domperidone in a plasma sample.Sekar Ayu PawestriAkhmad Kharis NugrohoEndang LukitaningsihPurwantiningsih PurwantiningsihDepartment of Chemistry, Universitas Gadjah Madaarticledomperidonerp-hplcspiked plasmavalidationChemistryQD1-999ENIndonesian Journal of Chemistry, Vol 21, Iss 5, Pp 1180-1186 (2021)
institution DOAJ
collection DOAJ
language EN
topic domperidone
rp-hplc
spiked plasma
validation
Chemistry
QD1-999
spellingShingle domperidone
rp-hplc
spiked plasma
validation
Chemistry
QD1-999
Sekar Ayu Pawestri
Akhmad Kharis Nugroho
Endang Lukitaningsih
Purwantiningsih Purwantiningsih
Validation and Quantification of Domperidone in Spiked Plasma Matrix Using Reversed Phase HPLC-UV Method
description Pharmacokinetics studies of domperidone generally analyze plasma matrix samples. The present work aimed to develop and validate a rapid and simple reversed phase-HPLC method for quantifying domperidone in plasma matrices. The chromatographic method implemented: 1. Luna Phenomenex® C18 (250 mm × 4.6 mm i.d; 5 µm) column, 2. isocratic mobile phase mixture of phosphate buffer 0.02 M:acetonitrile (70:30, v/v) with a flow rate of 1 mL/min, 3. UV detection at 285 nm. Domperidone and propranolol hydrochloride (as internal standard) were extracted from the deproteinated plasma sample. The method linearity was 0.998 in the range concentration of 15–200 ng/mL. The percentage of accuracy error was between -8.49–4.31%, while the percentage coefficient variation of precision ranged between 5.11–14.24%. This proposed method was simple, rapid (separation time less than 10 min), and selective. The validation parameters responses satisfied the method's requirements to determine domperidone in a plasma sample.
format article
author Sekar Ayu Pawestri
Akhmad Kharis Nugroho
Endang Lukitaningsih
Purwantiningsih Purwantiningsih
author_facet Sekar Ayu Pawestri
Akhmad Kharis Nugroho
Endang Lukitaningsih
Purwantiningsih Purwantiningsih
author_sort Sekar Ayu Pawestri
title Validation and Quantification of Domperidone in Spiked Plasma Matrix Using Reversed Phase HPLC-UV Method
title_short Validation and Quantification of Domperidone in Spiked Plasma Matrix Using Reversed Phase HPLC-UV Method
title_full Validation and Quantification of Domperidone in Spiked Plasma Matrix Using Reversed Phase HPLC-UV Method
title_fullStr Validation and Quantification of Domperidone in Spiked Plasma Matrix Using Reversed Phase HPLC-UV Method
title_full_unstemmed Validation and Quantification of Domperidone in Spiked Plasma Matrix Using Reversed Phase HPLC-UV Method
title_sort validation and quantification of domperidone in spiked plasma matrix using reversed phase hplc-uv method
publisher Department of Chemistry, Universitas Gadjah Mada
publishDate 2021
url https://doaj.org/article/2389c520e6f04a299f8a3bf9d0519868
work_keys_str_mv AT sekarayupawestri validationandquantificationofdomperidoneinspikedplasmamatrixusingreversedphasehplcuvmethod
AT akhmadkharisnugroho validationandquantificationofdomperidoneinspikedplasmamatrixusingreversedphasehplcuvmethod
AT endanglukitaningsih validationandquantificationofdomperidoneinspikedplasmamatrixusingreversedphasehplcuvmethod
AT purwantiningsihpurwantiningsih validationandquantificationofdomperidoneinspikedplasmamatrixusingreversedphasehplcuvmethod
_version_ 1718378856770961408