Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty

Abstract Our aim was to assess the accuracy of the obtained posterior tibial slope (PTS) with a fixed angle cutting block. 247 TKAs in 213 patients were reviewed. We included 104 Legion Prosthesis, 76 U2 Knee Prosthesis, 46 NexGen LPS-Flex Prosthesis, and 21 Vanguard Knee System products. Preoperati...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Yufeng Lu, Xuechao Yuan, Feng Qiao, Yangquan Hao
Formato: article
Lenguaje:EN
Publicado: Nature Portfolio 2021
Materias:
R
Q
Acceso en línea:https://doaj.org/article/24d993821ac84e72a578ac7bae2c6920
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:24d993821ac84e72a578ac7bae2c6920
record_format dspace
spelling oai:doaj.org-article:24d993821ac84e72a578ac7bae2c69202021-12-02T13:26:25ZEffects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty10.1038/s41598-021-86787-x2045-2322https://doaj.org/article/24d993821ac84e72a578ac7bae2c69202021-03-01T00:00:00Zhttps://doi.org/10.1038/s41598-021-86787-xhttps://doaj.org/toc/2045-2322Abstract Our aim was to assess the accuracy of the obtained posterior tibial slope (PTS) with a fixed angle cutting block. 247 TKAs in 213 patients were reviewed. We included 104 Legion Prosthesis, 76 U2 Knee Prosthesis, 46 NexGen LPS-Flex Prosthesis, and 21 Vanguard Knee System products. Preoperative and postoperative PTS were measured via expanded lateral tibia radiographs. For postoperative PTS, the Legion group had significantly smaller slopes than the U2 Knee group and Vanguard group. However, there was no significant difference between the Legion and NexGen groups, and no significant difference among the NexGen, U2 Knee, and Vanguard groups. Multiple linear regression showed that the different tibial lengths and preoperative PTS had statistically significant effects on postoperative PTS. However, there were weak correlations between the tibial length and PTS, and between preoperative and postoperative PTS. For TKA, although the PTS is not completely consistent with the angle of the cutting block, using conventional tibial bone resection technology with different tibial cutting instrumentations provided by various manufacturers in TKA can obtain safe PTS.Yufeng LuXuechao YuanFeng QiaoYangquan HaoNature PortfolioarticleMedicineRScienceQENScientific Reports, Vol 11, Iss 1, Pp 1-9 (2021)
institution DOAJ
collection DOAJ
language EN
topic Medicine
R
Science
Q
spellingShingle Medicine
R
Science
Q
Yufeng Lu
Xuechao Yuan
Feng Qiao
Yangquan Hao
Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty
description Abstract Our aim was to assess the accuracy of the obtained posterior tibial slope (PTS) with a fixed angle cutting block. 247 TKAs in 213 patients were reviewed. We included 104 Legion Prosthesis, 76 U2 Knee Prosthesis, 46 NexGen LPS-Flex Prosthesis, and 21 Vanguard Knee System products. Preoperative and postoperative PTS were measured via expanded lateral tibia radiographs. For postoperative PTS, the Legion group had significantly smaller slopes than the U2 Knee group and Vanguard group. However, there was no significant difference between the Legion and NexGen groups, and no significant difference among the NexGen, U2 Knee, and Vanguard groups. Multiple linear regression showed that the different tibial lengths and preoperative PTS had statistically significant effects on postoperative PTS. However, there were weak correlations between the tibial length and PTS, and between preoperative and postoperative PTS. For TKA, although the PTS is not completely consistent with the angle of the cutting block, using conventional tibial bone resection technology with different tibial cutting instrumentations provided by various manufacturers in TKA can obtain safe PTS.
format article
author Yufeng Lu
Xuechao Yuan
Feng Qiao
Yangquan Hao
author_facet Yufeng Lu
Xuechao Yuan
Feng Qiao
Yangquan Hao
author_sort Yufeng Lu
title Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty
title_short Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty
title_full Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty
title_fullStr Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty
title_full_unstemmed Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty
title_sort effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty
publisher Nature Portfolio
publishDate 2021
url https://doaj.org/article/24d993821ac84e72a578ac7bae2c6920
work_keys_str_mv AT yufenglu effectsofdifferentprostheticinstrumentationsontibialboneresectionintotalkneearthroplasty
AT xuechaoyuan effectsofdifferentprostheticinstrumentationsontibialboneresectionintotalkneearthroplasty
AT fengqiao effectsofdifferentprostheticinstrumentationsontibialboneresectionintotalkneearthroplasty
AT yangquanhao effectsofdifferentprostheticinstrumentationsontibialboneresectionintotalkneearthroplasty
_version_ 1718393018196688896