Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty
Abstract Our aim was to assess the accuracy of the obtained posterior tibial slope (PTS) with a fixed angle cutting block. 247 TKAs in 213 patients were reviewed. We included 104 Legion Prosthesis, 76 U2 Knee Prosthesis, 46 NexGen LPS-Flex Prosthesis, and 21 Vanguard Knee System products. Preoperati...
Guardado en:
Autores principales: | , , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Nature Portfolio
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/24d993821ac84e72a578ac7bae2c6920 |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
id |
oai:doaj.org-article:24d993821ac84e72a578ac7bae2c6920 |
---|---|
record_format |
dspace |
spelling |
oai:doaj.org-article:24d993821ac84e72a578ac7bae2c69202021-12-02T13:26:25ZEffects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty10.1038/s41598-021-86787-x2045-2322https://doaj.org/article/24d993821ac84e72a578ac7bae2c69202021-03-01T00:00:00Zhttps://doi.org/10.1038/s41598-021-86787-xhttps://doaj.org/toc/2045-2322Abstract Our aim was to assess the accuracy of the obtained posterior tibial slope (PTS) with a fixed angle cutting block. 247 TKAs in 213 patients were reviewed. We included 104 Legion Prosthesis, 76 U2 Knee Prosthesis, 46 NexGen LPS-Flex Prosthesis, and 21 Vanguard Knee System products. Preoperative and postoperative PTS were measured via expanded lateral tibia radiographs. For postoperative PTS, the Legion group had significantly smaller slopes than the U2 Knee group and Vanguard group. However, there was no significant difference between the Legion and NexGen groups, and no significant difference among the NexGen, U2 Knee, and Vanguard groups. Multiple linear regression showed that the different tibial lengths and preoperative PTS had statistically significant effects on postoperative PTS. However, there were weak correlations between the tibial length and PTS, and between preoperative and postoperative PTS. For TKA, although the PTS is not completely consistent with the angle of the cutting block, using conventional tibial bone resection technology with different tibial cutting instrumentations provided by various manufacturers in TKA can obtain safe PTS.Yufeng LuXuechao YuanFeng QiaoYangquan HaoNature PortfolioarticleMedicineRScienceQENScientific Reports, Vol 11, Iss 1, Pp 1-9 (2021) |
institution |
DOAJ |
collection |
DOAJ |
language |
EN |
topic |
Medicine R Science Q |
spellingShingle |
Medicine R Science Q Yufeng Lu Xuechao Yuan Feng Qiao Yangquan Hao Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty |
description |
Abstract Our aim was to assess the accuracy of the obtained posterior tibial slope (PTS) with a fixed angle cutting block. 247 TKAs in 213 patients were reviewed. We included 104 Legion Prosthesis, 76 U2 Knee Prosthesis, 46 NexGen LPS-Flex Prosthesis, and 21 Vanguard Knee System products. Preoperative and postoperative PTS were measured via expanded lateral tibia radiographs. For postoperative PTS, the Legion group had significantly smaller slopes than the U2 Knee group and Vanguard group. However, there was no significant difference between the Legion and NexGen groups, and no significant difference among the NexGen, U2 Knee, and Vanguard groups. Multiple linear regression showed that the different tibial lengths and preoperative PTS had statistically significant effects on postoperative PTS. However, there were weak correlations between the tibial length and PTS, and between preoperative and postoperative PTS. For TKA, although the PTS is not completely consistent with the angle of the cutting block, using conventional tibial bone resection technology with different tibial cutting instrumentations provided by various manufacturers in TKA can obtain safe PTS. |
format |
article |
author |
Yufeng Lu Xuechao Yuan Feng Qiao Yangquan Hao |
author_facet |
Yufeng Lu Xuechao Yuan Feng Qiao Yangquan Hao |
author_sort |
Yufeng Lu |
title |
Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty |
title_short |
Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty |
title_full |
Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty |
title_fullStr |
Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty |
title_full_unstemmed |
Effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty |
title_sort |
effects of different prosthetic instrumentations on tibial bone resection in total knee arthroplasty |
publisher |
Nature Portfolio |
publishDate |
2021 |
url |
https://doaj.org/article/24d993821ac84e72a578ac7bae2c6920 |
work_keys_str_mv |
AT yufenglu effectsofdifferentprostheticinstrumentationsontibialboneresectionintotalkneearthroplasty AT xuechaoyuan effectsofdifferentprostheticinstrumentationsontibialboneresectionintotalkneearthroplasty AT fengqiao effectsofdifferentprostheticinstrumentationsontibialboneresectionintotalkneearthroplasty AT yangquanhao effectsofdifferentprostheticinstrumentationsontibialboneresectionintotalkneearthroplasty |
_version_ |
1718393018196688896 |