Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study

The aim of this study was to assess the effect of application of a recently developed bio-adhesive (Impladhesive) to abutment screw threads on the removal torque value and rotational misfit at the implant–abutment junction. This in vitro study evaluated 20 implant fixtures and 20 straight abutments....

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Mahnaz Arshad, Sina Khayat Zadeh, Mohammad Atai, Gholamreza Shirani, Georgios E. Romanos, Seyed Hossein Bassir
Formato: article
Lenguaje:EN
Publicado: MDPI AG 2021
Materias:
T
Acceso en línea:https://doaj.org/article/2c75a496c3e64eec95113b5347d5e0b7
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:2c75a496c3e64eec95113b5347d5e0b7
record_format dspace
spelling oai:doaj.org-article:2c75a496c3e64eec95113b5347d5e0b72021-11-25T18:13:59ZEffect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study10.3390/ma142268321996-1944https://doaj.org/article/2c75a496c3e64eec95113b5347d5e0b72021-11-01T00:00:00Zhttps://www.mdpi.com/1996-1944/14/22/6832https://doaj.org/toc/1996-1944The aim of this study was to assess the effect of application of a recently developed bio-adhesive (Impladhesive) to abutment screw threads on the removal torque value and rotational misfit at the implant–abutment junction. This in vitro study evaluated 20 implant fixtures and 20 straight abutments. Specimens were randomly divided into two groups (<i>n</i> = 10) with/without adhesive application. In the adhesive group, the abutment was dipped in Impladhesive before torquing. In the control group, the abutment was torqued conventionally without adhesive application. The removal torque value was recorded after completion of the cyclic loading of 500,000 cycles with 2 Hz frequency and 75 N load. Rotational misfit was recorded using a video measuring machine. After applying the torque, the change in the bisector angle on the abutment hex was recorded for each implant. The biocompatibility of Impladhesive was evaluated using a MTT cell vitality assay. Normal distribution of data was assessed using the Kolmogorov–Smirnov test. Data were analyzed using a <i>t</i>-test and Pearson’s correlation coefficient The application of Impladhesive at the implant–abutment interface resulted in significantly greater mean removal torque value compared to the control group (<i>p</i> = 0.008). In addition, the mean rotational misfit at the implant–abutment interface was significantly lower in the use of Impladhesive compared to the control group (<i>p</i> = 0.001). In addition, the cell vitality was found to be greater than 80% at all evaluated time points. It can be concluded that the application of Impladhesive on the abutment screw significantly decreased rotational misfit and increased the removal torque value. Future studies are needed to evaluate the efficacy of this bio-adhesive an in vivo setting.Mahnaz ArshadSina Khayat ZadehMohammad AtaiGholamreza ShiraniGeorgios E. RomanosSeyed Hossein BassirMDPI AGarticleadhesivesdental implantsdental implant–abutment designin vitrotorqueTechnologyTElectrical engineering. Electronics. Nuclear engineeringTK1-9971Engineering (General). Civil engineering (General)TA1-2040MicroscopyQH201-278.5Descriptive and experimental mechanicsQC120-168.85ENMaterials, Vol 14, Iss 6832, p 6832 (2021)
institution DOAJ
collection DOAJ
language EN
topic adhesives
dental implants
dental implant–abutment design
in vitro
torque
Technology
T
Electrical engineering. Electronics. Nuclear engineering
TK1-9971
Engineering (General). Civil engineering (General)
TA1-2040
Microscopy
QH201-278.5
Descriptive and experimental mechanics
QC120-168.85
spellingShingle adhesives
dental implants
dental implant–abutment design
in vitro
torque
Technology
T
Electrical engineering. Electronics. Nuclear engineering
TK1-9971
Engineering (General). Civil engineering (General)
TA1-2040
Microscopy
QH201-278.5
Descriptive and experimental mechanics
QC120-168.85
Mahnaz Arshad
Sina Khayat Zadeh
Mohammad Atai
Gholamreza Shirani
Georgios E. Romanos
Seyed Hossein Bassir
Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study
description The aim of this study was to assess the effect of application of a recently developed bio-adhesive (Impladhesive) to abutment screw threads on the removal torque value and rotational misfit at the implant–abutment junction. This in vitro study evaluated 20 implant fixtures and 20 straight abutments. Specimens were randomly divided into two groups (<i>n</i> = 10) with/without adhesive application. In the adhesive group, the abutment was dipped in Impladhesive before torquing. In the control group, the abutment was torqued conventionally without adhesive application. The removal torque value was recorded after completion of the cyclic loading of 500,000 cycles with 2 Hz frequency and 75 N load. Rotational misfit was recorded using a video measuring machine. After applying the torque, the change in the bisector angle on the abutment hex was recorded for each implant. The biocompatibility of Impladhesive was evaluated using a MTT cell vitality assay. Normal distribution of data was assessed using the Kolmogorov–Smirnov test. Data were analyzed using a <i>t</i>-test and Pearson’s correlation coefficient The application of Impladhesive at the implant–abutment interface resulted in significantly greater mean removal torque value compared to the control group (<i>p</i> = 0.008). In addition, the mean rotational misfit at the implant–abutment interface was significantly lower in the use of Impladhesive compared to the control group (<i>p</i> = 0.001). In addition, the cell vitality was found to be greater than 80% at all evaluated time points. It can be concluded that the application of Impladhesive on the abutment screw significantly decreased rotational misfit and increased the removal torque value. Future studies are needed to evaluate the efficacy of this bio-adhesive an in vivo setting.
format article
author Mahnaz Arshad
Sina Khayat Zadeh
Mohammad Atai
Gholamreza Shirani
Georgios E. Romanos
Seyed Hossein Bassir
author_facet Mahnaz Arshad
Sina Khayat Zadeh
Mohammad Atai
Gholamreza Shirani
Georgios E. Romanos
Seyed Hossein Bassir
author_sort Mahnaz Arshad
title Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study
title_short Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study
title_full Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study
title_fullStr Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study
title_full_unstemmed Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study
title_sort effect of application of a bio-adhesive on the removal torque value and rotational misfit at the implant–abutment junction: an in vitro study
publisher MDPI AG
publishDate 2021
url https://doaj.org/article/2c75a496c3e64eec95113b5347d5e0b7
work_keys_str_mv AT mahnazarshad effectofapplicationofabioadhesiveontheremovaltorquevalueandrotationalmisfitattheimplantabutmentjunctionaninvitrostudy
AT sinakhayatzadeh effectofapplicationofabioadhesiveontheremovaltorquevalueandrotationalmisfitattheimplantabutmentjunctionaninvitrostudy
AT mohammadatai effectofapplicationofabioadhesiveontheremovaltorquevalueandrotationalmisfitattheimplantabutmentjunctionaninvitrostudy
AT gholamrezashirani effectofapplicationofabioadhesiveontheremovaltorquevalueandrotationalmisfitattheimplantabutmentjunctionaninvitrostudy
AT georgioseromanos effectofapplicationofabioadhesiveontheremovaltorquevalueandrotationalmisfitattheimplantabutmentjunctionaninvitrostudy
AT seyedhosseinbassir effectofapplicationofabioadhesiveontheremovaltorquevalueandrotationalmisfitattheimplantabutmentjunctionaninvitrostudy
_version_ 1718411427129065472