Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry

Abstract The purpose of the current study was to compare the measurements between a new optical biometer based on swept-source optical coherence tomography (SS-OCT), the OA-2000 (Tomey, Japan), and an optical biometer based on optical low coherence reflectometry (OLCR), the Lenstar (Haag-Streit, Swi...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Rongrong Gao, Hao Chen, Giacomo Savini, Yaxin Miao, Xiaorui Wang, Jing Yang, Weiqi Zhao, Qinmei Wang, Jinhai Huang
Formato: article
Lenguaje:EN
Publicado: Nature Portfolio 2017
Materias:
R
Q
Acceso en línea:https://doaj.org/article/3c9e066b13284fa39a816cbd7409c508
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:3c9e066b13284fa39a816cbd7409c508
record_format dspace
spelling oai:doaj.org-article:3c9e066b13284fa39a816cbd7409c5082021-12-02T16:06:17ZComparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry10.1038/s41598-017-02463-z2045-2322https://doaj.org/article/3c9e066b13284fa39a816cbd7409c5082017-05-01T00:00:00Zhttps://doi.org/10.1038/s41598-017-02463-zhttps://doaj.org/toc/2045-2322Abstract The purpose of the current study was to compare the measurements between a new optical biometer based on swept-source optical coherence tomography (SS-OCT), the OA-2000 (Tomey, Japan), and an optical biometer based on optical low coherence reflectometry (OLCR), the Lenstar (Haag-Streit, Switzerland). Ninety-nine eyes of 99 healthy subjects were included. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AD), lens thickness (LT), keratometry (K) readings, including flat K (Kf), steep K (Ks), mean K (Km), astigmatism vectors J0, J45 at diameters of 2.5 and 3.0 mm, and white-to-white diameter (WTW) were measured three times each using both biometer in normal eyes by random sequence. Bland-Altman analysis showed good agreement between the SS-OCT and OLCR devices for AL, AD, ACD, LT, with narrow 95% LoA (−0.05 to 0.07 mm, −0.09 to 0.10 mm, −0.10 to 0.09 mm, and −0.06 to 0.22 mm, respectively), and the P values of ACD were both >0.05. The CCT, Kf, Ks, Km, J0, J45 and WTW values provided by the OA-2000 were in good agreement with the Lenstar, and statistically significant differences were detected for some of them but not clinical differences. The agreement was excellent especially for AL.Rongrong GaoHao ChenGiacomo SaviniYaxin MiaoXiaorui WangJing YangWeiqi ZhaoQinmei WangJinhai HuangNature PortfolioarticleMedicineRScienceQENScientific Reports, Vol 7, Iss 1, Pp 1-12 (2017)
institution DOAJ
collection DOAJ
language EN
topic Medicine
R
Science
Q
spellingShingle Medicine
R
Science
Q
Rongrong Gao
Hao Chen
Giacomo Savini
Yaxin Miao
Xiaorui Wang
Jing Yang
Weiqi Zhao
Qinmei Wang
Jinhai Huang
Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry
description Abstract The purpose of the current study was to compare the measurements between a new optical biometer based on swept-source optical coherence tomography (SS-OCT), the OA-2000 (Tomey, Japan), and an optical biometer based on optical low coherence reflectometry (OLCR), the Lenstar (Haag-Streit, Switzerland). Ninety-nine eyes of 99 healthy subjects were included. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AD), lens thickness (LT), keratometry (K) readings, including flat K (Kf), steep K (Ks), mean K (Km), astigmatism vectors J0, J45 at diameters of 2.5 and 3.0 mm, and white-to-white diameter (WTW) were measured three times each using both biometer in normal eyes by random sequence. Bland-Altman analysis showed good agreement between the SS-OCT and OLCR devices for AL, AD, ACD, LT, with narrow 95% LoA (−0.05 to 0.07 mm, −0.09 to 0.10 mm, −0.10 to 0.09 mm, and −0.06 to 0.22 mm, respectively), and the P values of ACD were both >0.05. The CCT, Kf, Ks, Km, J0, J45 and WTW values provided by the OA-2000 were in good agreement with the Lenstar, and statistically significant differences were detected for some of them but not clinical differences. The agreement was excellent especially for AL.
format article
author Rongrong Gao
Hao Chen
Giacomo Savini
Yaxin Miao
Xiaorui Wang
Jing Yang
Weiqi Zhao
Qinmei Wang
Jinhai Huang
author_facet Rongrong Gao
Hao Chen
Giacomo Savini
Yaxin Miao
Xiaorui Wang
Jing Yang
Weiqi Zhao
Qinmei Wang
Jinhai Huang
author_sort Rongrong Gao
title Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry
title_short Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry
title_full Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry
title_fullStr Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry
title_full_unstemmed Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry
title_sort comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry
publisher Nature Portfolio
publishDate 2017
url https://doaj.org/article/3c9e066b13284fa39a816cbd7409c508
work_keys_str_mv AT rongronggao comparisonofocularbiometricmeasurementsbetweenanewsweptsourceopticalcoherencetomographyandacommonopticallowcoherencereflectometry
AT haochen comparisonofocularbiometricmeasurementsbetweenanewsweptsourceopticalcoherencetomographyandacommonopticallowcoherencereflectometry
AT giacomosavini comparisonofocularbiometricmeasurementsbetweenanewsweptsourceopticalcoherencetomographyandacommonopticallowcoherencereflectometry
AT yaxinmiao comparisonofocularbiometricmeasurementsbetweenanewsweptsourceopticalcoherencetomographyandacommonopticallowcoherencereflectometry
AT xiaoruiwang comparisonofocularbiometricmeasurementsbetweenanewsweptsourceopticalcoherencetomographyandacommonopticallowcoherencereflectometry
AT jingyang comparisonofocularbiometricmeasurementsbetweenanewsweptsourceopticalcoherencetomographyandacommonopticallowcoherencereflectometry
AT weiqizhao comparisonofocularbiometricmeasurementsbetweenanewsweptsourceopticalcoherencetomographyandacommonopticallowcoherencereflectometry
AT qinmeiwang comparisonofocularbiometricmeasurementsbetweenanewsweptsourceopticalcoherencetomographyandacommonopticallowcoherencereflectometry
AT jinhaihuang comparisonofocularbiometricmeasurementsbetweenanewsweptsourceopticalcoherencetomographyandacommonopticallowcoherencereflectometry
_version_ 1718385017149718528