Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices

Abstract We assess repeatability of automatic measurements of a new anterior segment optical coherence tomographer and biometer (ANTERION) and their agreement with those provided by an anterior segment-optical coherence tomography device combined with Placido-disk corneal topography (MS-39) and a va...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Domenico Schiano-Lomoriello, Kenneth J. Hoffer, Irene Abicca, Giacomo Savini
Formato: article
Lenguaje:EN
Publicado: Nature Portfolio 2021
Materias:
R
Q
Acceso en línea:https://doaj.org/article/4546be9eaaaa44afa12a178d081b742d
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:4546be9eaaaa44afa12a178d081b742d
record_format dspace
spelling oai:doaj.org-article:4546be9eaaaa44afa12a178d081b742d2021-12-02T14:01:33ZRepeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices10.1038/s41598-020-79674-42045-2322https://doaj.org/article/4546be9eaaaa44afa12a178d081b742d2021-01-01T00:00:00Zhttps://doi.org/10.1038/s41598-020-79674-4https://doaj.org/toc/2045-2322Abstract We assess repeatability of automatic measurements of a new anterior segment optical coherence tomographer and biometer (ANTERION) and their agreement with those provided by an anterior segment-optical coherence tomography device combined with Placido-disk corneal topography (MS-39) and a validated optical biometer (IOLMaster 500). A consecutive series of patients underwent three measurements with ANTERION and one with MS-39. A subgroup of patients underwent biometry also with IOLMaster 500. Repeatability was assessed by means of within-subject standard deviation, coefficient of variation (COV), and intraclass correlation coefficient (ICC). Agreement was investigated with the 95% limits of agreement. Paired t-test and Wilcoxon matched-pairs test were performed to compare the measurements of the different devices. Repeatability of ANTERION measurements was high, with ICC > 0.98 for all parameters except astigmatism (0.963); all parameters apart from those related to astigmatism revealed a COV < 1%. Repeatability of astigmatism improved when only eyes whose keratometric astigmatism was higher than 1.0 D were investigated. Most measurements by ANTERION and MS-39 showed good agreement. No significant differences were found between measurements by ANTERION and IOLMaster, but for corneal diameter. ANTERION revealed high repeatability of automatic measurements and good agreement with both MS-39 and IOLMaster for most parameters.Domenico Schiano-LomorielloKenneth J. HofferIrene AbiccaGiacomo SaviniNature PortfolioarticleMedicineRScienceQENScientific Reports, Vol 11, Iss 1, Pp 1-7 (2021)
institution DOAJ
collection DOAJ
language EN
topic Medicine
R
Science
Q
spellingShingle Medicine
R
Science
Q
Domenico Schiano-Lomoriello
Kenneth J. Hoffer
Irene Abicca
Giacomo Savini
Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices
description Abstract We assess repeatability of automatic measurements of a new anterior segment optical coherence tomographer and biometer (ANTERION) and their agreement with those provided by an anterior segment-optical coherence tomography device combined with Placido-disk corneal topography (MS-39) and a validated optical biometer (IOLMaster 500). A consecutive series of patients underwent three measurements with ANTERION and one with MS-39. A subgroup of patients underwent biometry also with IOLMaster 500. Repeatability was assessed by means of within-subject standard deviation, coefficient of variation (COV), and intraclass correlation coefficient (ICC). Agreement was investigated with the 95% limits of agreement. Paired t-test and Wilcoxon matched-pairs test were performed to compare the measurements of the different devices. Repeatability of ANTERION measurements was high, with ICC > 0.98 for all parameters except astigmatism (0.963); all parameters apart from those related to astigmatism revealed a COV < 1%. Repeatability of astigmatism improved when only eyes whose keratometric astigmatism was higher than 1.0 D were investigated. Most measurements by ANTERION and MS-39 showed good agreement. No significant differences were found between measurements by ANTERION and IOLMaster, but for corneal diameter. ANTERION revealed high repeatability of automatic measurements and good agreement with both MS-39 and IOLMaster for most parameters.
format article
author Domenico Schiano-Lomoriello
Kenneth J. Hoffer
Irene Abicca
Giacomo Savini
author_facet Domenico Schiano-Lomoriello
Kenneth J. Hoffer
Irene Abicca
Giacomo Savini
author_sort Domenico Schiano-Lomoriello
title Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices
title_short Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices
title_full Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices
title_fullStr Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices
title_full_unstemmed Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices
title_sort repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices
publisher Nature Portfolio
publishDate 2021
url https://doaj.org/article/4546be9eaaaa44afa12a178d081b742d
work_keys_str_mv AT domenicoschianolomoriello repeatabilityofautomatedmeasurementsbyanewanteriorsegmentopticalcoherencetomographerandbiometerandagreementwithstandarddevices
AT kennethjhoffer repeatabilityofautomatedmeasurementsbyanewanteriorsegmentopticalcoherencetomographerandbiometerandagreementwithstandarddevices
AT ireneabicca repeatabilityofautomatedmeasurementsbyanewanteriorsegmentopticalcoherencetomographerandbiometerandagreementwithstandarddevices
AT giacomosavini repeatabilityofautomatedmeasurementsbyanewanteriorsegmentopticalcoherencetomographerandbiometerandagreementwithstandarddevices
_version_ 1718392137746219008