Anterior segment biometry using ultrasound biomicroscopy and the Artemis-2 very high frequency ultrasound scanner
Haya M Al-Farhan, Reem N AlMutairiKing Saud University, College of Applied Medical Sciences, Department of Optometry and Vision Sciences, Riyadh, Kingdom of Saudi ArabiaPurpose: To compare the precision of anterior chamber angle (ACA) and anterior chamber depth (ACD) measurements taken with ultrasou...
Guardado en:
Autores principales: | , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Dove Medical Press
2013
|
Materias: | |
Acceso en línea: | https://doaj.org/article/462282163d544a538c7f15dcd4eb647f |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
id |
oai:doaj.org-article:462282163d544a538c7f15dcd4eb647f |
---|---|
record_format |
dspace |
spelling |
oai:doaj.org-article:462282163d544a538c7f15dcd4eb647f2021-12-02T00:44:02ZAnterior segment biometry using ultrasound biomicroscopy and the Artemis-2 very high frequency ultrasound scanner1177-54671177-5483https://doaj.org/article/462282163d544a538c7f15dcd4eb647f2013-01-01T00:00:00Zhttp://www.dovepress.com/anterior-segment-biometry-using-ultrasound-biomicroscopy-and-the-artem-a11972https://doaj.org/toc/1177-5467https://doaj.org/toc/1177-5483Haya M Al-Farhan, Reem N AlMutairiKing Saud University, College of Applied Medical Sciences, Department of Optometry and Vision Sciences, Riyadh, Kingdom of Saudi ArabiaPurpose: To compare the precision of anterior chamber angle (ACA) and anterior chamber depth (ACD) measurements taken with ultrasound biomicroscopy (UBM) and the Artemis-2 Very High Frequency Ultrasound Scanner (VHFUS) in normal subjects.Design: Prospective study.Methods: We randomly selected one eye from each of 59 normal subjects in this study. Two subjects dropped out of the study; the associated data were excluded from analysis. ACA and ACD measurements were obtained using the VHFUS and the UBM. The results were compared statistically using repeated-measures analysis of variance for the intraobserver repeatability, unpaired t-test, and limits of agreement.Results: The average ACA values for the UBM and the VHFUS (±standard deviation) were 41.83° ± 5.03° and 33.36° ± 6.03°, respectively. The average ACD values were 2.96 ± 0.34 mm and 2.87 ± 0.31 mm. The intraobserver repeatability analysis of variance P-values for ACA and ACD measurements using UBM were 0.10 and 0.68, respectively; for the Artemis-2 VHFUS, the respective values were 0.68 and 0.09. The difference in ACA measurements was statistically significant (t = 8.41; P < 0.0001), while the difference in ACD values was not (t = 1.51; P < 0.13). The mean ACA difference was 8.50° ± 2.50°, and the limits of agreement were +13.30° to −3.60°. The mean ACD difference was 0.09 ± 0.27 mm, and the limits of agreement ranged from 0.61 mm to −0.43 mm. The mean difference percentage of ACD was 3.1% for both instruments.Conclusion: In case of the ACD, both instruments can be used interchangeably; however, with the ACA instruments, they cannot be used interchangeably.Keywords: anterior chamber angle, anterior chamber depth, Artemis-2 VHF scanner, ultrasound biomicroscope, normal eyesAl-Farhan HMAlMutairi RNDove Medical PressarticleOphthalmologyRE1-994ENClinical Ophthalmology, Vol 2013, Iss default, Pp 141-147 (2013) |
institution |
DOAJ |
collection |
DOAJ |
language |
EN |
topic |
Ophthalmology RE1-994 |
spellingShingle |
Ophthalmology RE1-994 Al-Farhan HM AlMutairi RN Anterior segment biometry using ultrasound biomicroscopy and the Artemis-2 very high frequency ultrasound scanner |
description |
Haya M Al-Farhan, Reem N AlMutairiKing Saud University, College of Applied Medical Sciences, Department of Optometry and Vision Sciences, Riyadh, Kingdom of Saudi ArabiaPurpose: To compare the precision of anterior chamber angle (ACA) and anterior chamber depth (ACD) measurements taken with ultrasound biomicroscopy (UBM) and the Artemis-2 Very High Frequency Ultrasound Scanner (VHFUS) in normal subjects.Design: Prospective study.Methods: We randomly selected one eye from each of 59 normal subjects in this study. Two subjects dropped out of the study; the associated data were excluded from analysis. ACA and ACD measurements were obtained using the VHFUS and the UBM. The results were compared statistically using repeated-measures analysis of variance for the intraobserver repeatability, unpaired t-test, and limits of agreement.Results: The average ACA values for the UBM and the VHFUS (±standard deviation) were 41.83° ± 5.03° and 33.36° ± 6.03°, respectively. The average ACD values were 2.96 ± 0.34 mm and 2.87 ± 0.31 mm. The intraobserver repeatability analysis of variance P-values for ACA and ACD measurements using UBM were 0.10 and 0.68, respectively; for the Artemis-2 VHFUS, the respective values were 0.68 and 0.09. The difference in ACA measurements was statistically significant (t = 8.41; P < 0.0001), while the difference in ACD values was not (t = 1.51; P < 0.13). The mean ACA difference was 8.50° ± 2.50°, and the limits of agreement were +13.30° to −3.60°. The mean ACD difference was 0.09 ± 0.27 mm, and the limits of agreement ranged from 0.61 mm to −0.43 mm. The mean difference percentage of ACD was 3.1% for both instruments.Conclusion: In case of the ACD, both instruments can be used interchangeably; however, with the ACA instruments, they cannot be used interchangeably.Keywords: anterior chamber angle, anterior chamber depth, Artemis-2 VHF scanner, ultrasound biomicroscope, normal eyes |
format |
article |
author |
Al-Farhan HM AlMutairi RN |
author_facet |
Al-Farhan HM AlMutairi RN |
author_sort |
Al-Farhan HM |
title |
Anterior segment biometry using ultrasound biomicroscopy and the Artemis-2 very high frequency ultrasound scanner |
title_short |
Anterior segment biometry using ultrasound biomicroscopy and the Artemis-2 very high frequency ultrasound scanner |
title_full |
Anterior segment biometry using ultrasound biomicroscopy and the Artemis-2 very high frequency ultrasound scanner |
title_fullStr |
Anterior segment biometry using ultrasound biomicroscopy and the Artemis-2 very high frequency ultrasound scanner |
title_full_unstemmed |
Anterior segment biometry using ultrasound biomicroscopy and the Artemis-2 very high frequency ultrasound scanner |
title_sort |
anterior segment biometry using ultrasound biomicroscopy and the artemis-2 very high frequency ultrasound scanner |
publisher |
Dove Medical Press |
publishDate |
2013 |
url |
https://doaj.org/article/462282163d544a538c7f15dcd4eb647f |
work_keys_str_mv |
AT alfarhanhm anteriorsegmentbiometryusingultrasoundbiomicroscopyandtheartemis2veryhighfrequencyultrasoundscanner AT almutairirn anteriorsegmentbiometryusingultrasoundbiomicroscopyandtheartemis2veryhighfrequencyultrasoundscanner |
_version_ |
1718403491274162176 |