Accuracy of BIS monitoring using a novel interface device connecting conventional needle-electrodes and BIS sensors during frontal neurosurgical procedures.

<h4>Background</h4>Bispectral index (BIS) monitoring is a widely used non-invasive method to monitor the depth of anesthesia. However, in the event of surgeries requiring a frontal approach, placement of the electrode may be impossible at the designated area to achieve a proper BIS measu...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Hideki Harada, Seiya Muta, Tatsuyuki Kakuma, Misa Ukeda, So Ota, Maiko Hirata, Hiroshi Fujioka, Osamu Nakashima, Barbara Dietel, Miyuki Tauchi
Formato: article
Lenguaje:EN
Publicado: Public Library of Science (PLoS) 2021
Materias:
R
Q
Acceso en línea:https://doaj.org/article/4c7e27a2b23c448c8da3187b642203bf
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
Descripción
Sumario:<h4>Background</h4>Bispectral index (BIS) monitoring is a widely used non-invasive method to monitor the depth of anesthesia. However, in the event of surgeries requiring a frontal approach, placement of the electrode may be impossible at the designated area to achieve a proper BIS measurement.<h4>Methods</h4>We developed an investigational interface device to connect needle-electrodes to BIS sensors. The safety and clinical performance were investigated in patients who underwent surgery. Direct BIS values from a disposable BIS electrode and indirect values via the interface device were simultaneously recorded from the same areas of electrode placement in a single patient. The agreement between the direct and indirect BIS values was statistically analyzed.<h4>Results</h4>The interface device with a silver electrode demonstrated sufficient electric conduction to transmit electroencephalogram signals. The overall BIS curves were similar to those of direct BIS monitoring. Direct and indirect BIS values from 18 patients were statistically analyzed using a linear mixed model and a significant concordance was confirmed (indirect BIS = 7.0405 + 0.8286 * direct BIS, p<0.0001). Most observed data (2582/2787 data points, 92.64%) had BIS unit differences of 10 or less.<h4>Conclusions</h4>The interface device provides an opportunity for intraoperative BIS monitoring of patients, whose clinical situation does not permit the placement of conventional adhesive sensors at the standard location.