Micro-Chamber/Thermal Extractor (µ-CTE) as a new sampling system for VOCs emitted by feces
Abstract VOCs (volatile organic compounds) are increasingly wished to be used in diagnosis of diseases. They present strategic advantages, when compared to classical methods used, such as simplicity and current availability of performant non-invasive sample collection methods/systems. However, stand...
Guardado en:
Autores principales: | , , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Nature Portfolio
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/548711760f424381b4283b4a193789dc |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
Sumario: | Abstract VOCs (volatile organic compounds) are increasingly wished to be used in diagnosis of diseases. They present strategic advantages, when compared to classical methods used, such as simplicity and current availability of performant non-invasive sample collection methods/systems. However, standardized sampling methods are required in order to achieve reproducible results. In the current study we developed a method to be used for feces sampling using a Micro-Chamber/Thermal Extractor (µ-CTE). Design Expert software (with Box–Behnken design) was used to predict the solutions. Therefore, by using the simulation experimental plan that was further experimentally verified, extraction time of 19.6 min, at extraction temperature of 30.6 °C by using a flow rate of 48.7 mL/min provided the higher response. The developed method was validated by using correlation tests and Network analysis, which both proved the validity of the developed model. |
---|