More accurate systolic blood pressure measurement is required for improved hypertension management: a perspective

Meir Nitzan,1 Itzchak Slotki,2 Linda Shavit2 1Department of Applied Physics/Electro-Optics, Jerusalem College of Technology, 2Department of Nephrology, Shaare Zedek Medical Center, Jerusalem, Israel Abstract: The commonly used techniques for systolic blood pressure (SBP) and diastolic blood pressure...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Nitzan M, Slotki I, Shavit L
Formato: article
Lenguaje:EN
Publicado: Dove Medical Press 2017
Materias:
Acceso en línea:https://doaj.org/article/5b8f7ced36cb41e9930c50fbbb27e5d2
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
Descripción
Sumario:Meir Nitzan,1 Itzchak Slotki,2 Linda Shavit2 1Department of Applied Physics/Electro-Optics, Jerusalem College of Technology, 2Department of Nephrology, Shaare Zedek Medical Center, Jerusalem, Israel Abstract: The commonly used techniques for systolic blood pressure (SBP) and diastolic blood pressure (DBP) measurement are the auscultatory Korotkoff-based sphygmomanometry and oscillometry. The former technique is relatively accurate but is limited to a physician’s office because its automatic variant is subject to noise artifacts. Consequently, the Korotkoff-based measurement overestimates the blood pressure in some patients due to white coat effect, and because it is a single measurement, it cannot properly represent the variable blood pressure. Automatic oscillometry can be used at home by the patient and is preferred even in clinics. However, the technique’s accuracy is low and errors of 10–15 mmHg are common. Recently, we have developed an automatic technique for SBP measurement, based on an arm pressure cuff and a finger photoplethysmographic probe. The technique was found to be significantly more accurate than oscillometry, and comparable to the Korotkoff-based technique, the reference-standard for non-invasive blood pressure measurements. The measurement of SBP is a mainstay for the diagnosis and follow-up of hypertension, which is a major risk factor for several adverse events, mainly cardiovascular. Lowering blood pressure evidently reduces the risk, but excessive lowering can result in hypotension and consequently hypoperfusion to vital organs, since blood pressure is the driving force for blood flow. Erroneous measurement by 10 mmHg can lead to a similar unintended reduction of SBP and may adversely affect patients treated to an SBP of 120–130 mmHg. In particular, in elderly patients, unintended excessive reduction of blood pressure due to inaccurate SBP measurement can result in cerebral hypoperfusion and consequent cognitive decline. By using a more accurate technique for automatic SBP measurement (such as the photoplethysmographic-based technique), the optimal blood pressure target can be achieved with lower risk for hypotension and its adverse events. Keywords: oscillometry, hypotension, cognitive impairment, photoplethysmography