Evolution of visual performance in 70 eyes implanted with the Tecnis® ZMB00 multifocal intraocular lens
Carlos Palomino Bautista David Carmona González Alfredo Castillo GómezHospital Universitario Quirón, Madrid, SpainPurpose: To evaluate the development of visual acuity, dysphotopsia phenomena, and subjective sensations in patients implanted with the Tecni...
Guardado en:
Autores principales: | , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Dove Medical Press
2012
|
Materias: | |
Acceso en línea: | https://doaj.org/article/5f5a2b05988b4699b0f262dc4b4b218b |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
id |
oai:doaj.org-article:5f5a2b05988b4699b0f262dc4b4b218b |
---|---|
record_format |
dspace |
spelling |
oai:doaj.org-article:5f5a2b05988b4699b0f262dc4b4b218b2021-12-02T06:23:59ZEvolution of visual performance in 70 eyes implanted with the Tecnis® ZMB00 multifocal intraocular lens1177-54671177-5483https://doaj.org/article/5f5a2b05988b4699b0f262dc4b4b218b2012-03-01T00:00:00Zhttp://www.dovepress.com/evolution-of-visual-performance-in-70-eyes-implanted-with-the-tecnisre-a9535https://doaj.org/toc/1177-5467https://doaj.org/toc/1177-5483Carlos Palomino Bautista David Carmona González Alfredo Castillo GómezHospital Universitario Quirón, Madrid, SpainPurpose: To evaluate the development of visual acuity, dysphotopsia phenomena, and subjective sensations in patients implanted with the Tecnis® ZMB00 (Advanced Medical Optics Inc, Santa Ana, CA) multifocal intraocular lens.Methods: In a sample of 70 eyes, distance and near visual acuity, with and without correction, contrast sensitivity, and patient satisfaction, were analyzed at 15, 30, and 60 days post-surgery.Results: Mean uncorrected distance visual acuity (logarithm of minimum angle of resolution) at 15, 30, and 60 days was 0.194 ± 0.054, 0.119 ± 0.026, and 0.076 ± 0.014 (P < 0.0001), respectively, and with correction, 0.051 ± 0.007 vs 0.041 ± 0.004 vs 0.022 ± 0.002 (P < 0.0001), respectively). At 60 days, 94.3% of eyes could read 1.00 close-up without correction. Patient satisfaction in terms of dysphotopsia effects and visual acuity was excellent. The mean contrast sensitivity was 1.64 ± 0.10 (logarithmic units) measured with the Pelli–Robson test.Conclusion: This type of multifocal intraocular lens was very effective at the distances analyzed, producing excellent objective and subjective results.Keywords: higher-order aberrations, diffractive multifocal intraocular lens, visual acuity, dysphotopsia Palomino Bautista CCarmona González DCastillo Gómez ADove Medical PressarticleOphthalmologyRE1-994ENClinical Ophthalmology, Vol 2012, Iss default, Pp 403-407 (2012) |
institution |
DOAJ |
collection |
DOAJ |
language |
EN |
topic |
Ophthalmology RE1-994 |
spellingShingle |
Ophthalmology RE1-994 Palomino Bautista C Carmona González D Castillo Gómez A Evolution of visual performance in 70 eyes implanted with the Tecnis® ZMB00 multifocal intraocular lens |
description |
Carlos Palomino Bautista David Carmona González Alfredo Castillo GómezHospital Universitario Quirón, Madrid, SpainPurpose: To evaluate the development of visual acuity, dysphotopsia phenomena, and subjective sensations in patients implanted with the Tecnis® ZMB00 (Advanced Medical Optics Inc, Santa Ana, CA) multifocal intraocular lens.Methods: In a sample of 70 eyes, distance and near visual acuity, with and without correction, contrast sensitivity, and patient satisfaction, were analyzed at 15, 30, and 60 days post-surgery.Results: Mean uncorrected distance visual acuity (logarithm of minimum angle of resolution) at 15, 30, and 60 days was 0.194 ± 0.054, 0.119 ± 0.026, and 0.076 ± 0.014 (P < 0.0001), respectively, and with correction, 0.051 ± 0.007 vs 0.041 ± 0.004 vs 0.022 ± 0.002 (P < 0.0001), respectively). At 60 days, 94.3% of eyes could read 1.00 close-up without correction. Patient satisfaction in terms of dysphotopsia effects and visual acuity was excellent. The mean contrast sensitivity was 1.64 ± 0.10 (logarithmic units) measured with the Pelli–Robson test.Conclusion: This type of multifocal intraocular lens was very effective at the distances analyzed, producing excellent objective and subjective results.Keywords: higher-order aberrations, diffractive multifocal intraocular lens, visual acuity, dysphotopsia  |
format |
article |
author |
Palomino Bautista C Carmona González D Castillo Gómez A |
author_facet |
Palomino Bautista C Carmona González D Castillo Gómez A |
author_sort |
Palomino Bautista C |
title |
Evolution of visual performance in 70 eyes implanted with the Tecnis® ZMB00 multifocal intraocular lens |
title_short |
Evolution of visual performance in 70 eyes implanted with the Tecnis® ZMB00 multifocal intraocular lens |
title_full |
Evolution of visual performance in 70 eyes implanted with the Tecnis® ZMB00 multifocal intraocular lens |
title_fullStr |
Evolution of visual performance in 70 eyes implanted with the Tecnis® ZMB00 multifocal intraocular lens |
title_full_unstemmed |
Evolution of visual performance in 70 eyes implanted with the Tecnis® ZMB00 multifocal intraocular lens |
title_sort |
evolution of visual performance in 70 eyes implanted with the tecnis® zmb00 multifocal intraocular lens |
publisher |
Dove Medical Press |
publishDate |
2012 |
url |
https://doaj.org/article/5f5a2b05988b4699b0f262dc4b4b218b |
work_keys_str_mv |
AT palominobautistac evolutionofvisualperformancein70eyesimplantedwiththetecnisampregzmb00multifocalintraocularlens AT carmonagonzampaacutelezd evolutionofvisualperformancein70eyesimplantedwiththetecnisampregzmb00multifocalintraocularlens AT castillogampoacutemeza evolutionofvisualperformancein70eyesimplantedwiththetecnisampregzmb00multifocalintraocularlens |
_version_ |
1718399934746591232 |