DIAGNOSTIC ACCURACY OF IN-HOUSE BIOCHEMICAL TESTING FOR IDENTIFICATION OF ENTEROCOCCUS SPECIES ISOLATED FROM VARIOUS CLINICAL SPECIMENSAGAINST VITEK 2 SYSTEM

Objective: To determine the diagnostic accuracy of in-house biochemical testing for identification of enterococcus species isolated from various clinical specimens against gold standard i.e., automated Vitek 2 system. This study also includes the antimicrobial susceptibility testing of enterococci a...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Mariam Sarwar, Syed Adeel Hussain Gardezi, Gohar Zaman, Aamer Ikram, Wajid Hussain, Muhammad Tahir Khadim
Formato: article
Lenguaje:EN
Publicado: Army Medical College Rawalpindi 2021
Materias:
R
Acceso en línea:https://doaj.org/article/6bd2b198851748a8a4529c83c806d538
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:6bd2b198851748a8a4529c83c806d538
record_format dspace
spelling oai:doaj.org-article:6bd2b198851748a8a4529c83c806d5382021-12-02T18:01:28ZDIAGNOSTIC ACCURACY OF IN-HOUSE BIOCHEMICAL TESTING FOR IDENTIFICATION OF ENTEROCOCCUS SPECIES ISOLATED FROM VARIOUS CLINICAL SPECIMENSAGAINST VITEK 2 SYSTEM0030-96482411-884210.51253/pafmj.v71i1.3095https://doaj.org/article/6bd2b198851748a8a4529c83c806d5382021-02-01T00:00:00Zhttps://pafmj.org/index.php/PAFMJ/article/view/3095https://doaj.org/toc/0030-9648https://doaj.org/toc/2411-8842Objective: To determine the diagnostic accuracy of in-house biochemical testing for identification of enterococcus species isolated from various clinical specimens against gold standard i.e., automated Vitek 2 system. This study also includes the antimicrobial susceptibility testing of enterococci against various antimicrobials. Study Design: Cross-sectional comparative study. Place and Duration of Study: Department of Microbiology, Armed Forces Institute of Pathology, Rawalpindi, from Apr 2017 to Mar 2018. Methodology: A total of 218 isolates from various clinical specimens suspected to be Enterococcus spp. were checked by in-house biochemical testing including bile esculin, 6.5% NaCl and 1% arabinose and results were compared with Vitek 2 compact system. The frequencies were determined by both systems and antimicrobial susceptibility testing was performed by disk diffusion as per clinical and laboratory standards institute guidelines. Results: Comparing the results of in-house testing with gold standard i.e., Vitek 2 system, the statistical data was calculated. Sensitivity turned out to be 100%, Specificity was found to be 68.75%. Positive and negative predictive values were 97.58% and 100% respectively. Accuracy turned out to be 97.71%. Conclusion: The in-house biochemical testing can be quite a useful method for identification of enterococci in resource-limited settings. However, it requires overight incubation and cannot identify other enterococcal species and non-enterococcal species. Vitek 2 is an automated system that is easy-to-handle, provides a rapid and reasonably accurate identification of enterococci alongwith accurate AST results. Enterococcal isolates from various clinical specimens in our setup showed least resistance to linezolid, followed by teicoplanin and vancomycin. Nitrofurantoin and.............Mariam SarwarSyed Adeel Hussain GardeziGohar ZamanAamer IkramWajid HussainMuhammad Tahir KhadimArmy Medical College Rawalpindiarticleenterococcus sppin-house biochemical testingvitek 2 systemMedicineRMedicine (General)R5-920ENPakistan Armed Forces Medical Journal, Vol 71, Iss 1, Pp 155-61 (2021)
institution DOAJ
collection DOAJ
language EN
topic enterococcus spp
in-house biochemical testing
vitek 2 system
Medicine
R
Medicine (General)
R5-920
spellingShingle enterococcus spp
in-house biochemical testing
vitek 2 system
Medicine
R
Medicine (General)
R5-920
Mariam Sarwar
Syed Adeel Hussain Gardezi
Gohar Zaman
Aamer Ikram
Wajid Hussain
Muhammad Tahir Khadim
DIAGNOSTIC ACCURACY OF IN-HOUSE BIOCHEMICAL TESTING FOR IDENTIFICATION OF ENTEROCOCCUS SPECIES ISOLATED FROM VARIOUS CLINICAL SPECIMENSAGAINST VITEK 2 SYSTEM
description Objective: To determine the diagnostic accuracy of in-house biochemical testing for identification of enterococcus species isolated from various clinical specimens against gold standard i.e., automated Vitek 2 system. This study also includes the antimicrobial susceptibility testing of enterococci against various antimicrobials. Study Design: Cross-sectional comparative study. Place and Duration of Study: Department of Microbiology, Armed Forces Institute of Pathology, Rawalpindi, from Apr 2017 to Mar 2018. Methodology: A total of 218 isolates from various clinical specimens suspected to be Enterococcus spp. were checked by in-house biochemical testing including bile esculin, 6.5% NaCl and 1% arabinose and results were compared with Vitek 2 compact system. The frequencies were determined by both systems and antimicrobial susceptibility testing was performed by disk diffusion as per clinical and laboratory standards institute guidelines. Results: Comparing the results of in-house testing with gold standard i.e., Vitek 2 system, the statistical data was calculated. Sensitivity turned out to be 100%, Specificity was found to be 68.75%. Positive and negative predictive values were 97.58% and 100% respectively. Accuracy turned out to be 97.71%. Conclusion: The in-house biochemical testing can be quite a useful method for identification of enterococci in resource-limited settings. However, it requires overight incubation and cannot identify other enterococcal species and non-enterococcal species. Vitek 2 is an automated system that is easy-to-handle, provides a rapid and reasonably accurate identification of enterococci alongwith accurate AST results. Enterococcal isolates from various clinical specimens in our setup showed least resistance to linezolid, followed by teicoplanin and vancomycin. Nitrofurantoin and.............
format article
author Mariam Sarwar
Syed Adeel Hussain Gardezi
Gohar Zaman
Aamer Ikram
Wajid Hussain
Muhammad Tahir Khadim
author_facet Mariam Sarwar
Syed Adeel Hussain Gardezi
Gohar Zaman
Aamer Ikram
Wajid Hussain
Muhammad Tahir Khadim
author_sort Mariam Sarwar
title DIAGNOSTIC ACCURACY OF IN-HOUSE BIOCHEMICAL TESTING FOR IDENTIFICATION OF ENTEROCOCCUS SPECIES ISOLATED FROM VARIOUS CLINICAL SPECIMENSAGAINST VITEK 2 SYSTEM
title_short DIAGNOSTIC ACCURACY OF IN-HOUSE BIOCHEMICAL TESTING FOR IDENTIFICATION OF ENTEROCOCCUS SPECIES ISOLATED FROM VARIOUS CLINICAL SPECIMENSAGAINST VITEK 2 SYSTEM
title_full DIAGNOSTIC ACCURACY OF IN-HOUSE BIOCHEMICAL TESTING FOR IDENTIFICATION OF ENTEROCOCCUS SPECIES ISOLATED FROM VARIOUS CLINICAL SPECIMENSAGAINST VITEK 2 SYSTEM
title_fullStr DIAGNOSTIC ACCURACY OF IN-HOUSE BIOCHEMICAL TESTING FOR IDENTIFICATION OF ENTEROCOCCUS SPECIES ISOLATED FROM VARIOUS CLINICAL SPECIMENSAGAINST VITEK 2 SYSTEM
title_full_unstemmed DIAGNOSTIC ACCURACY OF IN-HOUSE BIOCHEMICAL TESTING FOR IDENTIFICATION OF ENTEROCOCCUS SPECIES ISOLATED FROM VARIOUS CLINICAL SPECIMENSAGAINST VITEK 2 SYSTEM
title_sort diagnostic accuracy of in-house biochemical testing for identification of enterococcus species isolated from various clinical specimensagainst vitek 2 system
publisher Army Medical College Rawalpindi
publishDate 2021
url https://doaj.org/article/6bd2b198851748a8a4529c83c806d538
work_keys_str_mv AT mariamsarwar diagnosticaccuracyofinhousebiochemicaltestingforidentificationofenterococcusspeciesisolatedfromvariousclinicalspecimensagainstvitek2system
AT syedadeelhussaingardezi diagnosticaccuracyofinhousebiochemicaltestingforidentificationofenterococcusspeciesisolatedfromvariousclinicalspecimensagainstvitek2system
AT goharzaman diagnosticaccuracyofinhousebiochemicaltestingforidentificationofenterococcusspeciesisolatedfromvariousclinicalspecimensagainstvitek2system
AT aamerikram diagnosticaccuracyofinhousebiochemicaltestingforidentificationofenterococcusspeciesisolatedfromvariousclinicalspecimensagainstvitek2system
AT wajidhussain diagnosticaccuracyofinhousebiochemicaltestingforidentificationofenterococcusspeciesisolatedfromvariousclinicalspecimensagainstvitek2system
AT muhammadtahirkhadim diagnosticaccuracyofinhousebiochemicaltestingforidentificationofenterococcusspeciesisolatedfromvariousclinicalspecimensagainstvitek2system
_version_ 1718378994707988480