Impact of additional mattresses in emergency CT on the automated patient centering proposed by a 3D camera: a phantom study

Abstract To assess the impact of the use of additional mattresses of different thicknesses on radiation dose and image noise based on the patient centering proposed by a 3D camera for CT. An anthropomorphic phantom was placed on mattresses of different thicknesses (from 3.5 to 13.5 cm) on the table...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Joël Greffier, Julien Frandon, Hélène de Forges, Aymeric Hamard, Asmaa Belaouni, Jean Baptiste Wahl, Djamel Dabli, Jean Paul Beregi
Formato: article
Lenguaje:EN
Publicado: Nature Portfolio 2021
Materias:
R
Q
Acceso en línea:https://doaj.org/article/6e99bd2ef06c4670a5280ece01e594fd
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:6e99bd2ef06c4670a5280ece01e594fd
record_format dspace
spelling oai:doaj.org-article:6e99bd2ef06c4670a5280ece01e594fd2021-12-02T18:02:49ZImpact of additional mattresses in emergency CT on the automated patient centering proposed by a 3D camera: a phantom study10.1038/s41598-021-92637-72045-2322https://doaj.org/article/6e99bd2ef06c4670a5280ece01e594fd2021-06-01T00:00:00Zhttps://doi.org/10.1038/s41598-021-92637-7https://doaj.org/toc/2045-2322Abstract To assess the impact of the use of additional mattresses of different thicknesses on radiation dose and image noise based on the patient centering proposed by a 3D camera for CT. An anthropomorphic phantom was placed on mattresses of different thicknesses (from 3.5 to 13.5 cm) on the table of a CT scanner. The automated patient centering proposed by a 3D camera was analysed as a function of mattress thickness and corrected for table height. For this purpose, the impact on image noise in the lung tissues in the chest area and in the soft tissues in the abdomen-pelvis area, modulated mAs (mAsmod) by the tube current modulation system (TCM) and volume CT dose index (CTDIvol) was assessed slice-by-slice along the z-axis after CT scans. With the use of a mattress, the automated centering proposed by the 3D camera resulted in placement of the phantom above the isocentre. This incorrect positioning led to a significant increase in the mAsmod along the z-axis (p < 0.05) and in the CTDIvol. Image noise was significantly higher (p < 0.05) for automated phantom centering than with manual phantom centering. Differences of image noise between acquisitions with mattresses after automatic and manual phantom centering increased with the mattress thicknesses. The use of an additional mattress placed between the patient’s back and the table-top would require correcting the vertical centering proposed by the 3D camera. This manual correction is essential to avoid increased dose delivered to the patient and higher image noise.Joël GreffierJulien FrandonHélène de ForgesAymeric HamardAsmaa BelaouniJean Baptiste WahlDjamel DabliJean Paul BeregiNature PortfolioarticleMedicineRScienceQENScientific Reports, Vol 11, Iss 1, Pp 1-9 (2021)
institution DOAJ
collection DOAJ
language EN
topic Medicine
R
Science
Q
spellingShingle Medicine
R
Science
Q
Joël Greffier
Julien Frandon
Hélène de Forges
Aymeric Hamard
Asmaa Belaouni
Jean Baptiste Wahl
Djamel Dabli
Jean Paul Beregi
Impact of additional mattresses in emergency CT on the automated patient centering proposed by a 3D camera: a phantom study
description Abstract To assess the impact of the use of additional mattresses of different thicknesses on radiation dose and image noise based on the patient centering proposed by a 3D camera for CT. An anthropomorphic phantom was placed on mattresses of different thicknesses (from 3.5 to 13.5 cm) on the table of a CT scanner. The automated patient centering proposed by a 3D camera was analysed as a function of mattress thickness and corrected for table height. For this purpose, the impact on image noise in the lung tissues in the chest area and in the soft tissues in the abdomen-pelvis area, modulated mAs (mAsmod) by the tube current modulation system (TCM) and volume CT dose index (CTDIvol) was assessed slice-by-slice along the z-axis after CT scans. With the use of a mattress, the automated centering proposed by the 3D camera resulted in placement of the phantom above the isocentre. This incorrect positioning led to a significant increase in the mAsmod along the z-axis (p < 0.05) and in the CTDIvol. Image noise was significantly higher (p < 0.05) for automated phantom centering than with manual phantom centering. Differences of image noise between acquisitions with mattresses after automatic and manual phantom centering increased with the mattress thicknesses. The use of an additional mattress placed between the patient’s back and the table-top would require correcting the vertical centering proposed by the 3D camera. This manual correction is essential to avoid increased dose delivered to the patient and higher image noise.
format article
author Joël Greffier
Julien Frandon
Hélène de Forges
Aymeric Hamard
Asmaa Belaouni
Jean Baptiste Wahl
Djamel Dabli
Jean Paul Beregi
author_facet Joël Greffier
Julien Frandon
Hélène de Forges
Aymeric Hamard
Asmaa Belaouni
Jean Baptiste Wahl
Djamel Dabli
Jean Paul Beregi
author_sort Joël Greffier
title Impact of additional mattresses in emergency CT on the automated patient centering proposed by a 3D camera: a phantom study
title_short Impact of additional mattresses in emergency CT on the automated patient centering proposed by a 3D camera: a phantom study
title_full Impact of additional mattresses in emergency CT on the automated patient centering proposed by a 3D camera: a phantom study
title_fullStr Impact of additional mattresses in emergency CT on the automated patient centering proposed by a 3D camera: a phantom study
title_full_unstemmed Impact of additional mattresses in emergency CT on the automated patient centering proposed by a 3D camera: a phantom study
title_sort impact of additional mattresses in emergency ct on the automated patient centering proposed by a 3d camera: a phantom study
publisher Nature Portfolio
publishDate 2021
url https://doaj.org/article/6e99bd2ef06c4670a5280ece01e594fd
work_keys_str_mv AT joelgreffier impactofadditionalmattressesinemergencyctontheautomatedpatientcenteringproposedbya3dcameraaphantomstudy
AT julienfrandon impactofadditionalmattressesinemergencyctontheautomatedpatientcenteringproposedbya3dcameraaphantomstudy
AT helenedeforges impactofadditionalmattressesinemergencyctontheautomatedpatientcenteringproposedbya3dcameraaphantomstudy
AT aymerichamard impactofadditionalmattressesinemergencyctontheautomatedpatientcenteringproposedbya3dcameraaphantomstudy
AT asmaabelaouni impactofadditionalmattressesinemergencyctontheautomatedpatientcenteringproposedbya3dcameraaphantomstudy
AT jeanbaptistewahl impactofadditionalmattressesinemergencyctontheautomatedpatientcenteringproposedbya3dcameraaphantomstudy
AT djameldabli impactofadditionalmattressesinemergencyctontheautomatedpatientcenteringproposedbya3dcameraaphantomstudy
AT jeanpaulberegi impactofadditionalmattressesinemergencyctontheautomatedpatientcenteringproposedbya3dcameraaphantomstudy
_version_ 1718378872140988416