Measurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: hospital-based comparative study

Riyam Faihan Rashid, Qasim K Farhood Department of Ophthalmology, College of Medicine, University of Babylon, Babylon, Iraq Background: The measurement of central corneal thickness (CCT) plays an important role in the diagnosis and treatment of glaucoma and many corneal diseases.Objective of the s...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Rashid RF, Farhood QK
Formato: article
Lenguaje:EN
Publicado: Dove Medical Press 2016
Materias:
Acceso en línea:https://doaj.org/article/6e9df81a3e104edbb6f5eb634381ee80
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:6e9df81a3e104edbb6f5eb634381ee80
record_format dspace
spelling oai:doaj.org-article:6e9df81a3e104edbb6f5eb634381ee802021-12-02T00:47:20ZMeasurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: hospital-based comparative study1177-5483https://doaj.org/article/6e9df81a3e104edbb6f5eb634381ee802016-03-01T00:00:00Zhttps://www.dovepress.com/measurement-of-central-corneal-thickness-by-ultrasonic-pachymeter-and--peer-reviewed-article-OPTHhttps://doaj.org/toc/1177-5483Riyam Faihan Rashid, Qasim K Farhood Department of Ophthalmology, College of Medicine, University of Babylon, Babylon, Iraq Background: The measurement of central corneal thickness (CCT) plays an important role in the diagnosis and treatment of glaucoma and many corneal diseases.Objective of the study: To compare the measurement of CCT by ultrasonic pachymeter with that measured by oculus pentacam in both normal subjects and patients with well-controlled glaucoma.Patients and methods: In 173 eyes of both controls and patients with open-angle glaucoma (normal intraocular pressure) attending Ibn Al Haitham Teaching Eye Hospital, CCT was measured by oculus pentacam and then by ultrasound pachymeter at the same morning by the same technicians.Results: The results showed no significant difference in CCT readings measured by the two devices in both glaucoma and control groups (glaucoma group: 547.29±49.32 µm with pentacam vs 547.66±45.24 µm with ultrasound pachymeter; control group: 551.02±36.28 µm with pentacam vs 541.25±34.96 µm with ultrasound pachymeter). P-values were >0.05 in both groups (statistically not significant).Conclusion and recommendation: Ultrasonic pachymeter and oculus pentacam can be used interchangeably in measuring CCT, and we recommend a nontouch method (in this study, pentacam Scheimpflug camera) for measuring CCT during assessment of patients with glaucoma or any ocular disease or surgery. Keywords: CCT, open-angle glaucoma, ultrasound pachymeter, Scheimpflug cameraRashid RFFarhood QKDove Medical PressarticleCentral corneal thicknessultrasonicpentacamglaucomaOphthalmologyRE1-994ENClinical Ophthalmology, Vol 2016, Iss Issue 1, Pp 359-364 (2016)
institution DOAJ
collection DOAJ
language EN
topic Central corneal thickness
ultrasonic
pentacam
glaucoma
Ophthalmology
RE1-994
spellingShingle Central corneal thickness
ultrasonic
pentacam
glaucoma
Ophthalmology
RE1-994
Rashid RF
Farhood QK
Measurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: hospital-based comparative study
description Riyam Faihan Rashid, Qasim K Farhood Department of Ophthalmology, College of Medicine, University of Babylon, Babylon, Iraq Background: The measurement of central corneal thickness (CCT) plays an important role in the diagnosis and treatment of glaucoma and many corneal diseases.Objective of the study: To compare the measurement of CCT by ultrasonic pachymeter with that measured by oculus pentacam in both normal subjects and patients with well-controlled glaucoma.Patients and methods: In 173 eyes of both controls and patients with open-angle glaucoma (normal intraocular pressure) attending Ibn Al Haitham Teaching Eye Hospital, CCT was measured by oculus pentacam and then by ultrasound pachymeter at the same morning by the same technicians.Results: The results showed no significant difference in CCT readings measured by the two devices in both glaucoma and control groups (glaucoma group: 547.29±49.32 µm with pentacam vs 547.66±45.24 µm with ultrasound pachymeter; control group: 551.02±36.28 µm with pentacam vs 541.25±34.96 µm with ultrasound pachymeter). P-values were >0.05 in both groups (statistically not significant).Conclusion and recommendation: Ultrasonic pachymeter and oculus pentacam can be used interchangeably in measuring CCT, and we recommend a nontouch method (in this study, pentacam Scheimpflug camera) for measuring CCT during assessment of patients with glaucoma or any ocular disease or surgery. Keywords: CCT, open-angle glaucoma, ultrasound pachymeter, Scheimpflug camera
format article
author Rashid RF
Farhood QK
author_facet Rashid RF
Farhood QK
author_sort Rashid RF
title Measurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: hospital-based comparative study
title_short Measurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: hospital-based comparative study
title_full Measurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: hospital-based comparative study
title_fullStr Measurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: hospital-based comparative study
title_full_unstemmed Measurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: hospital-based comparative study
title_sort measurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: hospital-based comparative study
publisher Dove Medical Press
publishDate 2016
url https://doaj.org/article/6e9df81a3e104edbb6f5eb634381ee80
work_keys_str_mv AT rashidrf measurementofcentralcornealthicknessbyultrasonicpachymeterandoculuspentacaminpatientswithwellcontrolledglaucomahospitalbasedcomparativestudy
AT farhoodqk measurementofcentralcornealthicknessbyultrasonicpachymeterandoculuspentacaminpatientswithwellcontrolledglaucomahospitalbasedcomparativestudy
_version_ 1718403502522236928