Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems

Abstract Background To implement a tangential treatment technique for whole breast irradiation using the Varian Halcyon and to compare it with Elekta Synergy Agility plans. Methods For 20 patients two comparable treatment plans with respect to dose coverage and normal tissue sparing were generated....

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Anne Richter, Sonja Wegener, Kathrin Breuer, Gary Razinskas, Stefan Weick, Florian Exner, Klaus Bratengeier, Michael Flentje, Otto Sauer, Bülent Polat
Formato: article
Lenguaje:EN
Publicado: BMC 2021
Materias:
Acceso en línea:https://doaj.org/article/7b65f004c87240f1952754e3729a8009
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:7b65f004c87240f1952754e3729a8009
record_format dspace
spelling oai:doaj.org-article:7b65f004c87240f1952754e3729a80092021-11-08T10:44:59ZComparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems10.1186/s13014-021-01942-y1748-717Xhttps://doaj.org/article/7b65f004c87240f1952754e3729a80092021-11-01T00:00:00Zhttps://doi.org/10.1186/s13014-021-01942-yhttps://doaj.org/toc/1748-717XAbstract Background To implement a tangential treatment technique for whole breast irradiation using the Varian Halcyon and to compare it with Elekta Synergy Agility plans. Methods For 20 patients two comparable treatment plans with respect to dose coverage and normal tissue sparing were generated. Tangential field-in-field treatment plans (Pinnacle/Synergy) were replanned using the sliding window technique (Eclipse/Halcyon). Plan specific QA was performed using the portal Dosimetry and the ArcCHECK phantom. Imaging and treatment dose were evaluated for treatment delivery on both systems using a modified CIRS Phantom. Results The mean number of monitor units for a fraction dose of 2.67 Gy was 515 MUs and 260 MUs for Halcyon and Synergy Agility plans, respectively. The homogeneity index and dose coverage were similar for both treatment units. The plan specific QA showed good agreement between measured and calculated plans. All Halcyon plans passed portal dosimetry QA (3%/2 mm) with 100% points passing and ArcCheck QA (3%/2 mm) with 99.5%. Measurement of the cumulated treatment and imaging dose with the CIRS phantom resulted in lower dose to the contralateral breast for the Halcyon plans. Conclusions For the Varian Halcyon a plan quality similar to the Elekta Synergy device was achieved. For the Halcyon plans the dose contribution from the treatment fields to the contralateral breast was even lower due to less interleaf transmission of the Halcyon MLC and a lower contribution of scattered dose from the collimator system.Anne RichterSonja WegenerKathrin BreuerGary RazinskasStefan WeickFlorian ExnerKlaus BratengeierMichael FlentjeOtto SauerBülent PolatBMCarticleWhole breast irradiationHalcyonIGRTDose to OARsMedical physics. Medical radiology. Nuclear medicineR895-920Neoplasms. Tumors. Oncology. Including cancer and carcinogensRC254-282ENRadiation Oncology, Vol 16, Iss 1, Pp 1-9 (2021)
institution DOAJ
collection DOAJ
language EN
topic Whole breast irradiation
Halcyon
IGRT
Dose to OARs
Medical physics. Medical radiology. Nuclear medicine
R895-920
Neoplasms. Tumors. Oncology. Including cancer and carcinogens
RC254-282
spellingShingle Whole breast irradiation
Halcyon
IGRT
Dose to OARs
Medical physics. Medical radiology. Nuclear medicine
R895-920
Neoplasms. Tumors. Oncology. Including cancer and carcinogens
RC254-282
Anne Richter
Sonja Wegener
Kathrin Breuer
Gary Razinskas
Stefan Weick
Florian Exner
Klaus Bratengeier
Michael Flentje
Otto Sauer
Bülent Polat
Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems
description Abstract Background To implement a tangential treatment technique for whole breast irradiation using the Varian Halcyon and to compare it with Elekta Synergy Agility plans. Methods For 20 patients two comparable treatment plans with respect to dose coverage and normal tissue sparing were generated. Tangential field-in-field treatment plans (Pinnacle/Synergy) were replanned using the sliding window technique (Eclipse/Halcyon). Plan specific QA was performed using the portal Dosimetry and the ArcCHECK phantom. Imaging and treatment dose were evaluated for treatment delivery on both systems using a modified CIRS Phantom. Results The mean number of monitor units for a fraction dose of 2.67 Gy was 515 MUs and 260 MUs for Halcyon and Synergy Agility plans, respectively. The homogeneity index and dose coverage were similar for both treatment units. The plan specific QA showed good agreement between measured and calculated plans. All Halcyon plans passed portal dosimetry QA (3%/2 mm) with 100% points passing and ArcCheck QA (3%/2 mm) with 99.5%. Measurement of the cumulated treatment and imaging dose with the CIRS phantom resulted in lower dose to the contralateral breast for the Halcyon plans. Conclusions For the Varian Halcyon a plan quality similar to the Elekta Synergy device was achieved. For the Halcyon plans the dose contribution from the treatment fields to the contralateral breast was even lower due to less interleaf transmission of the Halcyon MLC and a lower contribution of scattered dose from the collimator system.
format article
author Anne Richter
Sonja Wegener
Kathrin Breuer
Gary Razinskas
Stefan Weick
Florian Exner
Klaus Bratengeier
Michael Flentje
Otto Sauer
Bülent Polat
author_facet Anne Richter
Sonja Wegener
Kathrin Breuer
Gary Razinskas
Stefan Weick
Florian Exner
Klaus Bratengeier
Michael Flentje
Otto Sauer
Bülent Polat
author_sort Anne Richter
title Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems
title_short Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems
title_full Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems
title_fullStr Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems
title_full_unstemmed Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems
title_sort comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the halcyon and synergy agility systems
publisher BMC
publishDate 2021
url https://doaj.org/article/7b65f004c87240f1952754e3729a8009
work_keys_str_mv AT annerichter comparisonofslidingwindowandfieldinfieldtechniquesfortangentialwholebreastirradiationusingthehalcyonandsynergyagilitysystems
AT sonjawegener comparisonofslidingwindowandfieldinfieldtechniquesfortangentialwholebreastirradiationusingthehalcyonandsynergyagilitysystems
AT kathrinbreuer comparisonofslidingwindowandfieldinfieldtechniquesfortangentialwholebreastirradiationusingthehalcyonandsynergyagilitysystems
AT garyrazinskas comparisonofslidingwindowandfieldinfieldtechniquesfortangentialwholebreastirradiationusingthehalcyonandsynergyagilitysystems
AT stefanweick comparisonofslidingwindowandfieldinfieldtechniquesfortangentialwholebreastirradiationusingthehalcyonandsynergyagilitysystems
AT florianexner comparisonofslidingwindowandfieldinfieldtechniquesfortangentialwholebreastirradiationusingthehalcyonandsynergyagilitysystems
AT klausbratengeier comparisonofslidingwindowandfieldinfieldtechniquesfortangentialwholebreastirradiationusingthehalcyonandsynergyagilitysystems
AT michaelflentje comparisonofslidingwindowandfieldinfieldtechniquesfortangentialwholebreastirradiationusingthehalcyonandsynergyagilitysystems
AT ottosauer comparisonofslidingwindowandfieldinfieldtechniquesfortangentialwholebreastirradiationusingthehalcyonandsynergyagilitysystems
AT bulentpolat comparisonofslidingwindowandfieldinfieldtechniquesfortangentialwholebreastirradiationusingthehalcyonandsynergyagilitysystems
_version_ 1718442630597050368