Injection time related to intraocular pressure using a CO2 driven preloaded injector: An experimental laboratory study.

<h4>Purpose</h4>Experimental study to measure the intraocular lens (IOL) injection time and injection speed at different intraocular pressure (IOP) settings when using the AutonoMe® injector.<h4>Methods</h4>In this experimental study, following phacoemulsification in porcine...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Jan N Weindler, Tadas Naujokaitis, Sonja K Schickhardt, Ramin Khoramnia, Gerd U Auffarth
Formato: article
Lenguaje:EN
Publicado: Public Library of Science (PLoS) 2021
Materias:
R
Q
Acceso en línea:https://doaj.org/article/7f3d1ca565da40aa91059e8c07323a1e
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
Descripción
Sumario:<h4>Purpose</h4>Experimental study to measure the intraocular lens (IOL) injection time and injection speed at different intraocular pressure (IOP) settings when using the AutonoMe® injector.<h4>Methods</h4>In this experimental study, following phacoemulsification in porcine cadaver eyes, a trocar was inserted at pars plana with a connected infusion and IOPs of 20, 50 and 80 mmHg were generated by altering the infusion height. Twelve CO2 gas-driven injectors were used to implant an IOL via a corneal incision of 2.2 mm. For each IOP setting, the duration of the IOL injection and the injection speed was measured by analyzing a video recording of the procedure.<h4>Results</h4>The mean ±SD injection time (seconds) was 4.47±0.50 at 20 mmHg, 4.98±0.55 at 50 mmHg and 5.47±0.20 at 80 mmHg. The mean ±SD injection speed (millimeters per seconds) was 1.36±0.15 at 20 mmHg, 1.22±0.14 at 50 mmHg and 1.10±0.04 at 80 mmHg. There was a significant (p<0.05) difference between the 20 and 80 mmHg groups in mean injection duration and injection speed.<h4>Conclusion</h4>The CO2 gas driven injector allows a safe IOL injection even at elevated IOP. Although the implantation time is slightly extended at higher IOPs, this does not seem to be clinically relevant. No IOL damage was observed at these pressure settings.