Intraocular pressure readings obtained through soft contact lenses using four types of tonometer

Joji Takenaka,1 Eriko Kunihara,1 Ulfah Rimayanti,2 Junko Tanaka,3 Makoto Kaneko,4 Yoshiaki Kiuchi1 1Department of Ophthalmology and Visual Science, Graduate School of Biomedical Sciences, Hiroshima University, Hiroshima, Japan; 2Faculty of Medicine and Health Sciences, UIN Alauddin Makassa...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Takenaka J, Kunihara E, Rimayanti U, Tanaka J, Kaneko M, Kiuchi Y
Formato: article
Lenguaje:EN
Publicado: Dove Medical Press 2015
Materias:
Acceso en línea:https://doaj.org/article/7fdc4543167f479b90573cc305a5d455
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:7fdc4543167f479b90573cc305a5d455
record_format dspace
spelling oai:doaj.org-article:7fdc4543167f479b90573cc305a5d4552021-12-02T02:59:36ZIntraocular pressure readings obtained through soft contact lenses using four types of tonometer1177-5483https://doaj.org/article/7fdc4543167f479b90573cc305a5d4552015-10-01T00:00:00Zhttps://www.dovepress.com/intraocular-pressure-readings-obtained-through-soft-contact-lenses-usi-peer-reviewed-article-OPTHhttps://doaj.org/toc/1177-5483Joji Takenaka,1 Eriko Kunihara,1 Ulfah Rimayanti,2 Junko Tanaka,3 Makoto Kaneko,4 Yoshiaki Kiuchi1 1Department of Ophthalmology and Visual Science, Graduate School of Biomedical Sciences, Hiroshima University, Hiroshima, Japan; 2Faculty of Medicine and Health Sciences, UIN Alauddin Makassar, South Sulawesi, Indonesia; 3Department of Epidemiology, Infectious Disease Control and Prevention, Graduate School of Biomedical Sciences, Hiroshima University, Hiroshima, Japan; 4Graduate School of Engineering, Osaka University, Osaka, Japan Purpose: To compare the reliability and accuracy of intraocular pressure (IOP) measured while wearing soft contact lenses (SCLs) using a non-contact tonometer (NCT), Goldmann applanation tonometer (GAT), iCare rebound tonometer (RBT) and the Tono-Pen XL.Methods: Twenty-six healthy subjects were examined. The IOP was measured using NCT, GAT, RBT, and the Tono-Pen XL, while the subjects wore SCLs -5.00 D, -0.50 D and +5.00 D. Bland–Altman plots and a regression analysis were used to compare the IOPs obtained with those instruments and the IOPs of the naked eyes measured using GAT (the standard IOPs in this study).Results: The IOPs obtained by the Tono-Pen XL while the subjects were wearing -5.00 D, -0.50 D, and +5.00 D SCLs were significantly higher than those of the naked eyes obtained using GAT. RBT showed that the IOPs were similar to the GAT standard IOPs under all conditions. The IOPs measured with NCT and GAT while the subjects were wearing -5.00 D and -0.50 D SCLs were similar to the GAT standard IOPs. The IOPs obtained with RBT and NCT while the subjects were wearing -5.00 D and -0.50 D SCLs exhibited a good correlation with the standard IOPs.Conclusion: The NCT and RBT are best when measuring IOP through hydrogel SCLs. Keywords: soft contact lens, intraocular pressure, rebound tonometer, non-contact tonometerTakenaka JKunihara ERimayanti UTanaka JKaneko MKiuchi YDove Medical PressarticleOphthalmologyRE1-994ENClinical Ophthalmology, Vol 2015, Iss default, Pp 1875-1881 (2015)
institution DOAJ
collection DOAJ
language EN
topic Ophthalmology
RE1-994
spellingShingle Ophthalmology
RE1-994
Takenaka J
Kunihara E
Rimayanti U
Tanaka J
Kaneko M
Kiuchi Y
Intraocular pressure readings obtained through soft contact lenses using four types of tonometer
description Joji Takenaka,1 Eriko Kunihara,1 Ulfah Rimayanti,2 Junko Tanaka,3 Makoto Kaneko,4 Yoshiaki Kiuchi1 1Department of Ophthalmology and Visual Science, Graduate School of Biomedical Sciences, Hiroshima University, Hiroshima, Japan; 2Faculty of Medicine and Health Sciences, UIN Alauddin Makassar, South Sulawesi, Indonesia; 3Department of Epidemiology, Infectious Disease Control and Prevention, Graduate School of Biomedical Sciences, Hiroshima University, Hiroshima, Japan; 4Graduate School of Engineering, Osaka University, Osaka, Japan Purpose: To compare the reliability and accuracy of intraocular pressure (IOP) measured while wearing soft contact lenses (SCLs) using a non-contact tonometer (NCT), Goldmann applanation tonometer (GAT), iCare rebound tonometer (RBT) and the Tono-Pen XL.Methods: Twenty-six healthy subjects were examined. The IOP was measured using NCT, GAT, RBT, and the Tono-Pen XL, while the subjects wore SCLs -5.00 D, -0.50 D and +5.00 D. Bland–Altman plots and a regression analysis were used to compare the IOPs obtained with those instruments and the IOPs of the naked eyes measured using GAT (the standard IOPs in this study).Results: The IOPs obtained by the Tono-Pen XL while the subjects were wearing -5.00 D, -0.50 D, and +5.00 D SCLs were significantly higher than those of the naked eyes obtained using GAT. RBT showed that the IOPs were similar to the GAT standard IOPs under all conditions. The IOPs measured with NCT and GAT while the subjects were wearing -5.00 D and -0.50 D SCLs were similar to the GAT standard IOPs. The IOPs obtained with RBT and NCT while the subjects were wearing -5.00 D and -0.50 D SCLs exhibited a good correlation with the standard IOPs.Conclusion: The NCT and RBT are best when measuring IOP through hydrogel SCLs. Keywords: soft contact lens, intraocular pressure, rebound tonometer, non-contact tonometer
format article
author Takenaka J
Kunihara E
Rimayanti U
Tanaka J
Kaneko M
Kiuchi Y
author_facet Takenaka J
Kunihara E
Rimayanti U
Tanaka J
Kaneko M
Kiuchi Y
author_sort Takenaka J
title Intraocular pressure readings obtained through soft contact lenses using four types of tonometer
title_short Intraocular pressure readings obtained through soft contact lenses using four types of tonometer
title_full Intraocular pressure readings obtained through soft contact lenses using four types of tonometer
title_fullStr Intraocular pressure readings obtained through soft contact lenses using four types of tonometer
title_full_unstemmed Intraocular pressure readings obtained through soft contact lenses using four types of tonometer
title_sort intraocular pressure readings obtained through soft contact lenses using four types of tonometer
publisher Dove Medical Press
publishDate 2015
url https://doaj.org/article/7fdc4543167f479b90573cc305a5d455
work_keys_str_mv AT takenakaj intraocularpressurereadingsobtainedthroughsoftcontactlensesusingfourtypesoftonometer
AT kuniharae intraocularpressurereadingsobtainedthroughsoftcontactlensesusingfourtypesoftonometer
AT rimayantiu intraocularpressurereadingsobtainedthroughsoftcontactlensesusingfourtypesoftonometer
AT tanakaj intraocularpressurereadingsobtainedthroughsoftcontactlensesusingfourtypesoftonometer
AT kanekom intraocularpressurereadingsobtainedthroughsoftcontactlensesusingfourtypesoftonometer
AT kiuchiy intraocularpressurereadingsobtainedthroughsoftcontactlensesusingfourtypesoftonometer
_version_ 1718401992889466880