Randomized, masked, in vitro comparison of three commercially available tear film osmometers
Guillermo Rocha,1 Eric Gulliver,1 Armand Borovik,2 Clara C Chan2 1Department of Ophthalmology, University of Manitoba, Winnipeg, 2Department of Ophthalmology and Vision Sciences, University of Toronto, Toronto, Ontario, Canada Purpose: The purpose of this study was to compare the precision and acc...
Guardado en:
Autores principales: | , , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Dove Medical Press
2017
|
Materias: | |
Acceso en línea: | https://doaj.org/article/83cd7925b63144978d129cf7306874c5 |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
id |
oai:doaj.org-article:83cd7925b63144978d129cf7306874c5 |
---|---|
record_format |
dspace |
spelling |
oai:doaj.org-article:83cd7925b63144978d129cf7306874c52021-12-02T07:28:49ZRandomized, masked, in vitro comparison of three commercially available tear film osmometers1177-5483https://doaj.org/article/83cd7925b63144978d129cf7306874c52017-01-01T00:00:00Zhttps://www.dovepress.com/randomized-masked-in-vitro-comparison-of-three-commercially-available--peer-reviewed-article-OPTHhttps://doaj.org/toc/1177-5483Guillermo Rocha,1 Eric Gulliver,1 Armand Borovik,2 Clara C Chan2 1Department of Ophthalmology, University of Manitoba, Winnipeg, 2Department of Ophthalmology and Vision Sciences, University of Toronto, Toronto, Ontario, Canada Purpose: The purpose of this study was to compare the precision and accuracy of commercially available tear film osmometers. Methods: Contrived tear solution target values representing the physiological range of tear osmolarity (normal eyes 297 mOsm/L, moderately dry eyes 342 mOsm/L, and severe dry eyes 383 mOsm/L) were constructed using a mix of mono- and divalent electrolytes, metabolites, serum albumin, and pH balanced to 7.4. Solution values were randomized and masked from the investigators during testing. Osmometers (Wescor 5520 Vapro Pressure Osmometer: device A, TearLab Osmolarity System: device B, and i-Med Pharma i-Pen: device C) were calibrated according to manufacturer instructions. Each level was tested 64× on each osmometer across two sites. Accuracy was reported as a correlation coefficient against expected linear dilutions, precision was calculated as percent coefficient of variation. Results: Device A reported a correlation with known solutions of r2=0.98, with averages of 305.6±4.0, 352.2±5.5, and 389.8±4.0 mOsm/L, and coefficient of variations (CVs) of 1.3%, 1.6%, and 1.0%, respectively. Device B reported an r2=0.96, with averages of 300.6±3.7, 341.4±7.9, and 376.8±5.1 mOsm/L, and CVs of 1.2%, 2.3%, and 1.4%, respectively. Device C reported an r2=0.03, with averages of 336.4±21.5, 342.0±20.7, and 345.7±22.0 mOsm/L, and CVs of 6.4%, 6.1%, and 6.4%, respectively. Conclusion: In this randomized, masked, in vitro study, device A and device B had significantly better accuracy and precision in measuring osmolarity of contrived tear solutions of known target values compared to device C. Device C showed insufficient performance to accurately and precisely delineate osmolarity levels in the physiological range. Furthermore, in vivo studies would be required to compare performance in human subjects. Keywords: tear osmolarity, TearLab Osmometer, i-Pen Osmometer, electrical impedance, Wescor Osmometer, precisionRocha GGulliver EBorovik AChan CCDove Medical PressarticleTear OsmolarityTearLab Osmometeri-Pen OsmometerElectrical ImpedanceWescor OsmometerPrecisionOphthalmologyRE1-994ENClinical Ophthalmology, Vol Volume 11, Pp 243-248 (2017) |
institution |
DOAJ |
collection |
DOAJ |
language |
EN |
topic |
Tear Osmolarity TearLab Osmometer i-Pen Osmometer Electrical Impedance Wescor Osmometer Precision Ophthalmology RE1-994 |
spellingShingle |
Tear Osmolarity TearLab Osmometer i-Pen Osmometer Electrical Impedance Wescor Osmometer Precision Ophthalmology RE1-994 Rocha G Gulliver E Borovik A Chan CC Randomized, masked, in vitro comparison of three commercially available tear film osmometers |
description |
Guillermo Rocha,1 Eric Gulliver,1 Armand Borovik,2 Clara C Chan2 1Department of Ophthalmology, University of Manitoba, Winnipeg, 2Department of Ophthalmology and Vision Sciences, University of Toronto, Toronto, Ontario, Canada Purpose: The purpose of this study was to compare the precision and accuracy of commercially available tear film osmometers. Methods: Contrived tear solution target values representing the physiological range of tear osmolarity (normal eyes 297 mOsm/L, moderately dry eyes 342 mOsm/L, and severe dry eyes 383 mOsm/L) were constructed using a mix of mono- and divalent electrolytes, metabolites, serum albumin, and pH balanced to 7.4. Solution values were randomized and masked from the investigators during testing. Osmometers (Wescor 5520 Vapro Pressure Osmometer: device A, TearLab Osmolarity System: device B, and i-Med Pharma i-Pen: device C) were calibrated according to manufacturer instructions. Each level was tested 64× on each osmometer across two sites. Accuracy was reported as a correlation coefficient against expected linear dilutions, precision was calculated as percent coefficient of variation. Results: Device A reported a correlation with known solutions of r2=0.98, with averages of 305.6±4.0, 352.2±5.5, and 389.8±4.0 mOsm/L, and coefficient of variations (CVs) of 1.3%, 1.6%, and 1.0%, respectively. Device B reported an r2=0.96, with averages of 300.6±3.7, 341.4±7.9, and 376.8±5.1 mOsm/L, and CVs of 1.2%, 2.3%, and 1.4%, respectively. Device C reported an r2=0.03, with averages of 336.4±21.5, 342.0±20.7, and 345.7±22.0 mOsm/L, and CVs of 6.4%, 6.1%, and 6.4%, respectively. Conclusion: In this randomized, masked, in vitro study, device A and device B had significantly better accuracy and precision in measuring osmolarity of contrived tear solutions of known target values compared to device C. Device C showed insufficient performance to accurately and precisely delineate osmolarity levels in the physiological range. Furthermore, in vivo studies would be required to compare performance in human subjects. Keywords: tear osmolarity, TearLab Osmometer, i-Pen Osmometer, electrical impedance, Wescor Osmometer, precision |
format |
article |
author |
Rocha G Gulliver E Borovik A Chan CC |
author_facet |
Rocha G Gulliver E Borovik A Chan CC |
author_sort |
Rocha G |
title |
Randomized, masked, in vitro comparison of three commercially available tear film osmometers |
title_short |
Randomized, masked, in vitro comparison of three commercially available tear film osmometers |
title_full |
Randomized, masked, in vitro comparison of three commercially available tear film osmometers |
title_fullStr |
Randomized, masked, in vitro comparison of three commercially available tear film osmometers |
title_full_unstemmed |
Randomized, masked, in vitro comparison of three commercially available tear film osmometers |
title_sort |
randomized, masked, in vitro comparison of three commercially available tear film osmometers |
publisher |
Dove Medical Press |
publishDate |
2017 |
url |
https://doaj.org/article/83cd7925b63144978d129cf7306874c5 |
work_keys_str_mv |
AT rochag randomizedmaskedinvitrocomparisonofthreecommerciallyavailabletearfilmosmometers AT gullivere randomizedmaskedinvitrocomparisonofthreecommerciallyavailabletearfilmosmometers AT borovika randomizedmaskedinvitrocomparisonofthreecommerciallyavailabletearfilmosmometers AT chancc randomizedmaskedinvitrocomparisonofthreecommerciallyavailabletearfilmosmometers |
_version_ |
1718399412202373120 |