Randomized, masked, in vitro comparison of three commercially available tear film osmometers

Guillermo Rocha,1 Eric Gulliver,1 Armand Borovik,2 Clara C Chan2 1Department of Ophthalmology, University of Manitoba, Winnipeg, 2Department of Ophthalmology and Vision Sciences, University of Toronto, Toronto, Ontario, Canada Purpose: The purpose of this study was to compare the precision and acc...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Rocha G, Gulliver E, Borovik A, Chan CC
Formato: article
Lenguaje:EN
Publicado: Dove Medical Press 2017
Materias:
Acceso en línea:https://doaj.org/article/83cd7925b63144978d129cf7306874c5
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:83cd7925b63144978d129cf7306874c5
record_format dspace
spelling oai:doaj.org-article:83cd7925b63144978d129cf7306874c52021-12-02T07:28:49ZRandomized, masked, in vitro comparison of three commercially available tear film osmometers1177-5483https://doaj.org/article/83cd7925b63144978d129cf7306874c52017-01-01T00:00:00Zhttps://www.dovepress.com/randomized-masked-in-vitro-comparison-of-three-commercially-available--peer-reviewed-article-OPTHhttps://doaj.org/toc/1177-5483Guillermo Rocha,1 Eric Gulliver,1 Armand Borovik,2 Clara C Chan2 1Department of Ophthalmology, University of Manitoba, Winnipeg, 2Department of Ophthalmology and Vision Sciences, University of Toronto, Toronto, Ontario, Canada Purpose: The purpose of this study was to compare the precision and accuracy of commercially available tear film osmometers. Methods: Contrived tear solution target values representing the physiological range of tear osmolarity (normal eyes 297 mOsm/L, moderately dry eyes 342 mOsm/L, and severe dry eyes 383 mOsm/L) were constructed using a mix of mono- and divalent electrolytes, metabolites, serum albumin, and pH balanced to 7.4. Solution values were randomized and masked from the investigators during testing. Osmometers (Wescor 5520 Vapro Pressure Osmometer: device A, TearLab Osmolarity System: device B, and i-Med Pharma i-Pen: device C) were calibrated according to manufacturer instructions. Each level was tested 64× on each osmometer across two sites. Accuracy was reported as a correlation coefficient against expected linear dilutions, precision was calculated as percent coefficient of variation. Results: Device A reported a correlation with known solutions of r2=0.98, with averages of 305.6±4.0, 352.2±5.5, and 389.8±4.0 mOsm/L, and coefficient of variations (CVs) of 1.3%, 1.6%, and 1.0%, respectively. Device B reported an r2=0.96, with averages of 300.6±3.7, 341.4±7.9, and 376.8±5.1 mOsm/L, and CVs of 1.2%, 2.3%, and 1.4%, respectively. Device C reported an r2=0.03, with averages of 336.4±21.5, 342.0±20.7, and 345.7±22.0 mOsm/L, and CVs of 6.4%, 6.1%, and 6.4%, respectively. Conclusion: In this randomized, masked, in vitro study, device A and device B had significantly better accuracy and precision in measuring osmolarity of contrived tear solutions of known target values compared to device C. Device C showed insufficient performance to accurately and precisely delineate osmolarity levels in the physiological range. Furthermore, in vivo studies would be required to compare performance in human subjects. Keywords: tear osmolarity, TearLab Osmometer, i-Pen Osmometer, electrical impedance, Wescor Osmometer, precisionRocha GGulliver EBorovik AChan CCDove Medical PressarticleTear OsmolarityTearLab Osmometeri-Pen OsmometerElectrical ImpedanceWescor OsmometerPrecisionOphthalmologyRE1-994ENClinical Ophthalmology, Vol Volume 11, Pp 243-248 (2017)
institution DOAJ
collection DOAJ
language EN
topic Tear Osmolarity
TearLab Osmometer
i-Pen Osmometer
Electrical Impedance
Wescor Osmometer
Precision
Ophthalmology
RE1-994
spellingShingle Tear Osmolarity
TearLab Osmometer
i-Pen Osmometer
Electrical Impedance
Wescor Osmometer
Precision
Ophthalmology
RE1-994
Rocha G
Gulliver E
Borovik A
Chan CC
Randomized, masked, in vitro comparison of three commercially available tear film osmometers
description Guillermo Rocha,1 Eric Gulliver,1 Armand Borovik,2 Clara C Chan2 1Department of Ophthalmology, University of Manitoba, Winnipeg, 2Department of Ophthalmology and Vision Sciences, University of Toronto, Toronto, Ontario, Canada Purpose: The purpose of this study was to compare the precision and accuracy of commercially available tear film osmometers. Methods: Contrived tear solution target values representing the physiological range of tear osmolarity (normal eyes 297 mOsm/L, moderately dry eyes 342 mOsm/L, and severe dry eyes 383 mOsm/L) were constructed using a mix of mono- and divalent electrolytes, metabolites, serum albumin, and pH balanced to 7.4. Solution values were randomized and masked from the investigators during testing. Osmometers (Wescor 5520 Vapro Pressure Osmometer: device A, TearLab Osmolarity System: device B, and i-Med Pharma i-Pen: device C) were calibrated according to manufacturer instructions. Each level was tested 64× on each osmometer across two sites. Accuracy was reported as a correlation coefficient against expected linear dilutions, precision was calculated as percent coefficient of variation. Results: Device A reported a correlation with known solutions of r2=0.98, with averages of 305.6±4.0, 352.2±5.5, and 389.8±4.0 mOsm/L, and coefficient of variations (CVs) of 1.3%, 1.6%, and 1.0%, respectively. Device B reported an r2=0.96, with averages of 300.6±3.7, 341.4±7.9, and 376.8±5.1 mOsm/L, and CVs of 1.2%, 2.3%, and 1.4%, respectively. Device C reported an r2=0.03, with averages of 336.4±21.5, 342.0±20.7, and 345.7±22.0 mOsm/L, and CVs of 6.4%, 6.1%, and 6.4%, respectively. Conclusion: In this randomized, masked, in vitro study, device A and device B had significantly better accuracy and precision in measuring osmolarity of contrived tear solutions of known target values compared to device C. Device C showed insufficient performance to accurately and precisely delineate osmolarity levels in the physiological range. Furthermore, in vivo studies would be required to compare performance in human subjects. Keywords: tear osmolarity, TearLab Osmometer, i-Pen Osmometer, electrical impedance, Wescor Osmometer, precision
format article
author Rocha G
Gulliver E
Borovik A
Chan CC
author_facet Rocha G
Gulliver E
Borovik A
Chan CC
author_sort Rocha G
title Randomized, masked, in vitro comparison of three commercially available tear film osmometers
title_short Randomized, masked, in vitro comparison of three commercially available tear film osmometers
title_full Randomized, masked, in vitro comparison of three commercially available tear film osmometers
title_fullStr Randomized, masked, in vitro comparison of three commercially available tear film osmometers
title_full_unstemmed Randomized, masked, in vitro comparison of three commercially available tear film osmometers
title_sort randomized, masked, in vitro comparison of three commercially available tear film osmometers
publisher Dove Medical Press
publishDate 2017
url https://doaj.org/article/83cd7925b63144978d129cf7306874c5
work_keys_str_mv AT rochag randomizedmaskedinvitrocomparisonofthreecommerciallyavailabletearfilmosmometers
AT gullivere randomizedmaskedinvitrocomparisonofthreecommerciallyavailabletearfilmosmometers
AT borovika randomizedmaskedinvitrocomparisonofthreecommerciallyavailabletearfilmosmometers
AT chancc randomizedmaskedinvitrocomparisonofthreecommerciallyavailabletearfilmosmometers
_version_ 1718399412202373120