Comparison of a human portable blood glucose meter and automated chemistry analyser for measurement of blood glucose concentrations in healthy dogs
Abstract Blood glucose measurement is one of the most commonly performed clinical diagnostic tests used to monitor glycaemia in several animal diseases. Usually, these laboratory analyses are performed on blood venous samples in remote laboratories, and the results are delayed, at best. The use of p...
Guardado en:
Autores principales: | , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Wiley
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/99d81109f3fe49c9b47dfef71260aa8a |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
Sumario: | Abstract Blood glucose measurement is one of the most commonly performed clinical diagnostic tests used to monitor glycaemia in several animal diseases. Usually, these laboratory analyses are performed on blood venous samples in remote laboratories, and the results are delayed, at best. The use of portable glucometers could evidently solve many constraints but veterinary‐use glucometers are not usually available. The present study aimed to compare blood glucose levels obtained by Bionime glucometer to the reference method using glucose oxidase. Venous blood was collected from a total number of 140 healthy dogs (72 males and 68 females), of different breeds (28 German Shepherd, 27 Pitt bull, 21 Boxer, 24 Rottweiler and 40 cross‐bred dogs) and different ages (range: 3 months–14 years) for glucose measurement using the reference laboratory method. Capillary blood samples were used to conduct a glucose measurement with a human‐use glucometer. Our results revealed that there was no significant difference between the mean capillary blood glucose (CBG) measured with the human‐use glucometer (5.06 ± 0.84 mmol/L) and the mean venous blood glucose (VBG) measured with the laboratory reference method (4.90 ± 0.73 mmol/L) (p = 0.42). Similarly, there was no significant difference of the mean CBG and VBG in male dogs (5.11 ± 0.88 and 4.97 ± 0.75 mmol/L, respectively) and female dogs (5.01 ± 0.81 and 5.07 ± 0.72 mmol/L, respectively) (p = 0.73 and 0.21, respectively), and no correlation to neither age (5.43 ± 0.90 and 5.20 ± 0.70 mmol/L in 3 to 6 month‐old dogs, 5.03 ± 0.82 and 4.94 ± 0.79 mmol/L in 6 months to 1 year‐old, 4.94 ± 0.67 and 5.13 ± 0.66 mmol/L in 1 to 4 year‐old dogs; 4.88 ± 0.94 and 4.80 ± 0.75 mmol/L in dogs older than 4 years, respectively, p < 0.05), nor to breed (4.94 ± 1.01 and 4.99 ± 0.79 mmol/L in German Shepherd, 5.13 ± 0.84 and 4.99 ± 0.79 mmol/L in Pitt Bull, 5.07 ± 0.94 and 5.07 ± 0.77 mmol/L in Boxer, 5.40 ± 0.59 and 5.48 ± 0.55 mmol/L in Rottweiler and 4.89 ± 0.75 and 4.77 ± 0.59 mmol/L in cross‐bred dogs, respectively, p < 0.05). The present study confirms that human glucometer can be used to measure glucose in dogs with a good accuracy. |
---|