Influence of frosted haptics on rotational stability of toric intraocular lenses
Abstract We investigated the unfolding property and rotational stability of a new toric intraocular lens (IOL); TECNIS toric II (toric-II, ZCW, Johnson & Johnson) that is an improved version of TECNIS toric IOL (toric-I, ZCV). Both IOLs are based on an identical platform, except for the frosted...
Guardado en:
Autores principales: | , , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Nature Portfolio
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/a55127cab60b4766be958ea23b8d81b7 |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
Sumario: | Abstract We investigated the unfolding property and rotational stability of a new toric intraocular lens (IOL); TECNIS toric II (toric-II, ZCW, Johnson & Johnson) that is an improved version of TECNIS toric IOL (toric-I, ZCV). Both IOLs are based on an identical platform, except for the frosted haptics with toric-II IOL. The study consisted of two parts; experimental study and clinical, retrospective, case series. Experimental study indicated that the overall time from IOL ejection to unfolding to 11 mm was significantly shorter with toricII than toric-I IOLs (p = 0.032), due to the earlier separation of the haptics from the optic with toric-II IOL. Clinical study included 131 eyes of 99 patients who had undergone phacoemulsification and toric IOL implantation. At 3 months postoperatively, toric-II IOL showed significantly better rotational stability than toric-I IOL, including smaller residual manifest astigmatism (p = 0.018), less amount of axis misalignment from the intended axis (p = 0.04), lower incidence of misalignment > 10º (p = 0.0044), and less degree of prediction errors (p = 0.043). Postoperative uncorrected distance visual acuity tended to be better in the toric-II than in the toric-I groups, with marginal statistical difference (p = 0.057). TECNIS toric II IOL with the frosted haptics showed significantly better rotational stability than its predecessor, probably due to quicker unfolding and greater friction with the capsular bag. |
---|