Influence of frosted haptics on rotational stability of toric intraocular lenses
Abstract We investigated the unfolding property and rotational stability of a new toric intraocular lens (IOL); TECNIS toric II (toric-II, ZCW, Johnson & Johnson) that is an improved version of TECNIS toric IOL (toric-I, ZCV). Both IOLs are based on an identical platform, except for the frosted...
Guardado en:
Autores principales: | , , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Nature Portfolio
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/a55127cab60b4766be958ea23b8d81b7 |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
id |
oai:doaj.org-article:a55127cab60b4766be958ea23b8d81b7 |
---|---|
record_format |
dspace |
spelling |
oai:doaj.org-article:a55127cab60b4766be958ea23b8d81b72021-12-02T16:50:24ZInfluence of frosted haptics on rotational stability of toric intraocular lenses10.1038/s41598-021-94293-32045-2322https://doaj.org/article/a55127cab60b4766be958ea23b8d81b72021-07-01T00:00:00Zhttps://doi.org/10.1038/s41598-021-94293-3https://doaj.org/toc/2045-2322Abstract We investigated the unfolding property and rotational stability of a new toric intraocular lens (IOL); TECNIS toric II (toric-II, ZCW, Johnson & Johnson) that is an improved version of TECNIS toric IOL (toric-I, ZCV). Both IOLs are based on an identical platform, except for the frosted haptics with toric-II IOL. The study consisted of two parts; experimental study and clinical, retrospective, case series. Experimental study indicated that the overall time from IOL ejection to unfolding to 11 mm was significantly shorter with toricII than toric-I IOLs (p = 0.032), due to the earlier separation of the haptics from the optic with toric-II IOL. Clinical study included 131 eyes of 99 patients who had undergone phacoemulsification and toric IOL implantation. At 3 months postoperatively, toric-II IOL showed significantly better rotational stability than toric-I IOL, including smaller residual manifest astigmatism (p = 0.018), less amount of axis misalignment from the intended axis (p = 0.04), lower incidence of misalignment > 10º (p = 0.0044), and less degree of prediction errors (p = 0.043). Postoperative uncorrected distance visual acuity tended to be better in the toric-II than in the toric-I groups, with marginal statistical difference (p = 0.057). TECNIS toric II IOL with the frosted haptics showed significantly better rotational stability than its predecessor, probably due to quicker unfolding and greater friction with the capsular bag.Ruriko TakakuShinichiro NakanoMasaharu IidaTetsuro OshikaNature PortfolioarticleMedicineRScienceQENScientific Reports, Vol 11, Iss 1, Pp 1-7 (2021) |
institution |
DOAJ |
collection |
DOAJ |
language |
EN |
topic |
Medicine R Science Q |
spellingShingle |
Medicine R Science Q Ruriko Takaku Shinichiro Nakano Masaharu Iida Tetsuro Oshika Influence of frosted haptics on rotational stability of toric intraocular lenses |
description |
Abstract We investigated the unfolding property and rotational stability of a new toric intraocular lens (IOL); TECNIS toric II (toric-II, ZCW, Johnson & Johnson) that is an improved version of TECNIS toric IOL (toric-I, ZCV). Both IOLs are based on an identical platform, except for the frosted haptics with toric-II IOL. The study consisted of two parts; experimental study and clinical, retrospective, case series. Experimental study indicated that the overall time from IOL ejection to unfolding to 11 mm was significantly shorter with toricII than toric-I IOLs (p = 0.032), due to the earlier separation of the haptics from the optic with toric-II IOL. Clinical study included 131 eyes of 99 patients who had undergone phacoemulsification and toric IOL implantation. At 3 months postoperatively, toric-II IOL showed significantly better rotational stability than toric-I IOL, including smaller residual manifest astigmatism (p = 0.018), less amount of axis misalignment from the intended axis (p = 0.04), lower incidence of misalignment > 10º (p = 0.0044), and less degree of prediction errors (p = 0.043). Postoperative uncorrected distance visual acuity tended to be better in the toric-II than in the toric-I groups, with marginal statistical difference (p = 0.057). TECNIS toric II IOL with the frosted haptics showed significantly better rotational stability than its predecessor, probably due to quicker unfolding and greater friction with the capsular bag. |
format |
article |
author |
Ruriko Takaku Shinichiro Nakano Masaharu Iida Tetsuro Oshika |
author_facet |
Ruriko Takaku Shinichiro Nakano Masaharu Iida Tetsuro Oshika |
author_sort |
Ruriko Takaku |
title |
Influence of frosted haptics on rotational stability of toric intraocular lenses |
title_short |
Influence of frosted haptics on rotational stability of toric intraocular lenses |
title_full |
Influence of frosted haptics on rotational stability of toric intraocular lenses |
title_fullStr |
Influence of frosted haptics on rotational stability of toric intraocular lenses |
title_full_unstemmed |
Influence of frosted haptics on rotational stability of toric intraocular lenses |
title_sort |
influence of frosted haptics on rotational stability of toric intraocular lenses |
publisher |
Nature Portfolio |
publishDate |
2021 |
url |
https://doaj.org/article/a55127cab60b4766be958ea23b8d81b7 |
work_keys_str_mv |
AT rurikotakaku influenceoffrostedhapticsonrotationalstabilityoftoricintraocularlenses AT shinichironakano influenceoffrostedhapticsonrotationalstabilityoftoricintraocularlenses AT masaharuiida influenceoffrostedhapticsonrotationalstabilityoftoricintraocularlenses AT tetsurooshika influenceoffrostedhapticsonrotationalstabilityoftoricintraocularlenses |
_version_ |
1718383055607955456 |