Reliability of qualitative occlusal tooth wear evaluation using an intraoral scanner: A pilot study.
Dental wear analysis through the use of an intraoral scanner is a reality of modern dentistry. This study aimed to investigate the reliability of qualitative tooth wear evaluation through three-dimensional images captured with an intraoral scanner and compared to clinical and photographic examinatio...
Guardado en:
Autores principales: | , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Public Library of Science (PLoS)
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/a6a363e702dd4dee9e644ccfc060896d |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
Sumario: | Dental wear analysis through the use of an intraoral scanner is a reality of modern dentistry. This study aimed to investigate the reliability of qualitative tooth wear evaluation through three-dimensional images captured with an intraoral scanner and compared to clinical and photographic examinations. Eighteen adult volunteers of both genders (18 to 55 years old) were submitted to clinical exams, intraoral photographs and intraoral scanning protocol using an optical scanner (TRIOS® Pod, 3Shape, Copenhagen, Denmark). Occlusal tooth wear, from second to second premolars, was measured by two evaluators and reevaluated after 30 days, according to a slight modification of the method described by Mockers et al. Weighted Kappa was used to measure intra and inter-examiner agreement. The Friedman test was used to verify the differences among methods. Random and systematic errors were assessed using Bland-Altman plots. All statistical analysis was performed with p<0.05. There was a substantive agreement for clinical (K = 0.75) and photographic exams (K = 0.79) and a moderate agreement for intraoral scanner analysis (K = 0.60) for inter-examiner evaluation. A substantial intra-examiner agreement was obtained for both evaluators. No significant difference between the methods was observed (p = 0.7343 for examiner 1 and 0.8007 for examiner 2). The Bland-Altman plot confirmed no systematic errors between the methods and a random error of 0.25 with the scanner method when compared to clinical assessment. All three methods showed reliability in qualitative occlusal tooth wear evaluation. Intraoral scanning seems to be a sound and reliable tool to evaluate tooth wear when compared to traditional methods, considering the lower inter-examiner agreement and the inherent limitations of this pilot study. Further research will be necessary in order to achieve more robust evidence. |
---|