Investigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin

Peter R Worsley, Nathan D Stanger, Aran K Horrell, Dan L Bader Clinical Academic Facility, Faculty of Health Sciences, University of Southampton, Southampton, UK Background: Research has shown that up to 33% of pressure ulcers (PUs) acquired in hospitals result from the application...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Worsley PR, Stanger ND, Horrell AK, Bader DL
Formato: article
Lenguaje:EN
Publicado: Dove Medical Press 2018
Materias:
Acceso en línea:https://doaj.org/article/a6b885ba32ed4edcbc0eb19f4d13cf11
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:a6b885ba32ed4edcbc0eb19f4d13cf11
record_format dspace
spelling oai:doaj.org-article:a6b885ba32ed4edcbc0eb19f4d13cf112021-12-02T04:49:43ZInvestigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin1179-1470https://doaj.org/article/a6b885ba32ed4edcbc0eb19f4d13cf112018-03-01T00:00:00Zhttps://www.dovepress.com/investigating-the-effects-of-cervical-collar-design-and-fit-on-the-bio-peer-reviewed-article-MDERhttps://doaj.org/toc/1179-1470Peter R Worsley, Nathan D Stanger, Aran K Horrell, Dan L Bader Clinical Academic Facility, Faculty of Health Sciences, University of Southampton, Southampton, UK Background: Research has shown that up to 33% of pressure ulcers (PUs) acquired in hospitals result from the application of a medical device. Cervical collars (C-collars) have been impli-cated in causing PUs, due to the mechanical force they apply to the skin. In order to improve our understanding of collar-related PUs, the present study aimed to assess the biomechanical, biochemical, and microclimate effects of C-collar design and fitting tension. Methods: A cohort of 15 healthy volunteers was fit with two different C-collars according to the manufacturer guidelines. Two further collar tensions were also defined as loose and tight for each device. Each collar condition was applied for 15 minutes, with a 10 minute refractory period. Measurements at the device–skin interface included interface pressures, inflammatory biomarkers, microclimate, range of cervical motion, and comfort scores. Results: The interface pressures at each tissue site increased monotonically with greater collar tension (p<0.01), irrespective of collar design. Biomarker analysis revealed that inflammatory cytokines (IL-1a) were elevated during collar application, with the highest increase during the tight fit condition, representing over a fourfold increase from unloaded conditions. Regardless of collar tension or type, there was an increase in temperature 1.5°C ±0.8°C compared to baseline values. Range of motion significantly decreased with greater strap tension (p<0.05), with an associated increase in discomfort. Conclusion: The present findings revealed that increasing C-collar tensions caused elevated contact pressures at the device–skin interface, with a corresponding inflammatory response at the skin. These peak contact pressures were highest at the occiput, corresponding with reported PU locations. Devices should be designed to uniformly distribute pressures, and appropriate guidance is needed for their application. Keywords: cervical collar, pressure ulcer, device design, biomechanics, biomarkersWorsley PRStanger NDHorrell AKBader DLDove Medical PressarticleCervical collarpressure ulcerdevice designbiomechanicsbiomarkers.Medical technologyR855-855.5ENMedical Devices: Evidence and Research, Vol Volume 11, Pp 87-94 (2018)
institution DOAJ
collection DOAJ
language EN
topic Cervical collar
pressure ulcer
device design
biomechanics
biomarkers.
Medical technology
R855-855.5
spellingShingle Cervical collar
pressure ulcer
device design
biomechanics
biomarkers.
Medical technology
R855-855.5
Worsley PR
Stanger ND
Horrell AK
Bader DL
Investigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin
description Peter R Worsley, Nathan D Stanger, Aran K Horrell, Dan L Bader Clinical Academic Facility, Faculty of Health Sciences, University of Southampton, Southampton, UK Background: Research has shown that up to 33% of pressure ulcers (PUs) acquired in hospitals result from the application of a medical device. Cervical collars (C-collars) have been impli-cated in causing PUs, due to the mechanical force they apply to the skin. In order to improve our understanding of collar-related PUs, the present study aimed to assess the biomechanical, biochemical, and microclimate effects of C-collar design and fitting tension. Methods: A cohort of 15 healthy volunteers was fit with two different C-collars according to the manufacturer guidelines. Two further collar tensions were also defined as loose and tight for each device. Each collar condition was applied for 15 minutes, with a 10 minute refractory period. Measurements at the device–skin interface included interface pressures, inflammatory biomarkers, microclimate, range of cervical motion, and comfort scores. Results: The interface pressures at each tissue site increased monotonically with greater collar tension (p<0.01), irrespective of collar design. Biomarker analysis revealed that inflammatory cytokines (IL-1a) were elevated during collar application, with the highest increase during the tight fit condition, representing over a fourfold increase from unloaded conditions. Regardless of collar tension or type, there was an increase in temperature 1.5°C ±0.8°C compared to baseline values. Range of motion significantly decreased with greater strap tension (p<0.05), with an associated increase in discomfort. Conclusion: The present findings revealed that increasing C-collar tensions caused elevated contact pressures at the device–skin interface, with a corresponding inflammatory response at the skin. These peak contact pressures were highest at the occiput, corresponding with reported PU locations. Devices should be designed to uniformly distribute pressures, and appropriate guidance is needed for their application. Keywords: cervical collar, pressure ulcer, device design, biomechanics, biomarkers
format article
author Worsley PR
Stanger ND
Horrell AK
Bader DL
author_facet Worsley PR
Stanger ND
Horrell AK
Bader DL
author_sort Worsley PR
title Investigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin
title_short Investigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin
title_full Investigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin
title_fullStr Investigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin
title_full_unstemmed Investigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin
title_sort investigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin
publisher Dove Medical Press
publishDate 2018
url https://doaj.org/article/a6b885ba32ed4edcbc0eb19f4d13cf11
work_keys_str_mv AT worsleypr investigatingtheeffectsofcervicalcollardesignandfitonthebiomechanicalandbiomarkerreactionattheskin
AT stangernd investigatingtheeffectsofcervicalcollardesignandfitonthebiomechanicalandbiomarkerreactionattheskin
AT horrellak investigatingtheeffectsofcervicalcollardesignandfitonthebiomechanicalandbiomarkerreactionattheskin
AT baderdl investigatingtheeffectsofcervicalcollardesignandfitonthebiomechanicalandbiomarkerreactionattheskin
_version_ 1718401028525654016