The Sonic Resonance Method and the Impulse Excitation Technique: A Comparison Study
In this study, resonant frequencies of flexurally vibrating samples were measured using the sonic resonant method (SRM) and the impulse excitation technique (IET) to assess the equivalency of these two methods. Samples were made from different materials and with two shapes (prism with rectangular cr...
Guardado en:
Autores principales: | , , , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
MDPI AG
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/a9f5781655f743f9b6623288f6c1bd1b |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
Sumario: | In this study, resonant frequencies of flexurally vibrating samples were measured using the sonic resonant method (SRM) and the impulse excitation technique (IET) to assess the equivalency of these two methods. Samples were made from different materials and with two shapes (prism with rectangular cross-section and cylinder with circular cross-section). The mean values and standard deviations of the resonant frequencies were compared using the t-test and the F-test. The tests showed an equivalency of both methods in measuring resonant frequency. The differences between the values measured using SRM and IET were not significant. Graphically, the relationship between the resonant frequencies is a line with a slope of 0.9993 ≈ 1. |
---|