A comparison of NCT, Goldman application tonometry values with and without fluorescein

Hakika Erdogan, Ziya Akingol, Ozlem Cam, Sadik Sencan Department of Ophthalmology, Faculty of Medicine, Maltepe University, Maltepe, Istanbul, Turkey  Purpose: The aim of the study was to statistically compare intraocular pressure (IOP) values measured using noncontact tonometer (N...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Erdogan H, Akingol Z, Cam O, Sencan S
Formato: article
Lenguaje:EN
Publicado: Dove Medical Press 2018
Materias:
Acceso en línea:https://doaj.org/article/ae5c2da0a09b4169bbc9eee84cb16989
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:ae5c2da0a09b4169bbc9eee84cb16989
record_format dspace
spelling oai:doaj.org-article:ae5c2da0a09b4169bbc9eee84cb169892021-12-02T02:18:07ZA comparison of NCT, Goldman application tonometry values with and without fluorescein1177-5483https://doaj.org/article/ae5c2da0a09b4169bbc9eee84cb169892018-10-01T00:00:00Zhttps://www.dovepress.com/a-comparison-of-nct-goldman-application-tonometry-values-with-and-with-peer-reviewed-article-OPTHhttps://doaj.org/toc/1177-5483Hakika Erdogan, Ziya Akingol, Ozlem Cam, Sadik Sencan Department of Ophthalmology, Faculty of Medicine, Maltepe University, Maltepe, Istanbul, Turkey  Purpose: The aim of the study was to statistically compare intraocular pressure (IOP) values measured using noncontact tonometer (NCT), Goldmann applanation tonometry (GAT) with fluorescein (fGAT), and GAT without fluorescein (nGAT). The study was also performed to test whether the values obtained using each technique change in accordance with the central corneal thickness (CCT) and refractive and keratometric values.Study design: This study was a prospective study of 188 eyes of 94 healthy volunteers.Methods: IOP was measured using fGAT, nGAT and NCT. CCT, refractive values, and keratometric values were measured, and the correlations and differences in the IOP for each tonometer were investigated.Results: The mean IOP values obtained with the NCT, nGAT, and fGAT were 17.5±3.7, 12.3±2.7, and 12.5±2 mmHg. The mean CCT was 538.2±34.4 µm, the mean refractive value was 0.9±1.2 D, and the mean keratometric value was 43.5±1.5 D. NCT was positively correlated with fGAT and GAT values and was significantly higher than both the values. There were no differences between fGAT and GAT values. No correlation was observed between the CCT and keratometric and refractive values and the difference between NCT and nGAT or fGAT.Conclusion: Differences in the measurements obtained using nGAT and fGAT were insignificant (P>0.05). Both values were positively correlated with NCT measurements (r=0.354, P<0.05) and were independent of CCT, keratometry, and refraction values. nGAT appears to be suitable for use in routine clinic practice. Keywords: fluorescein, intraocular pressure, Goldmann applanation tonometry, noncontact tonometry, central corneal thickness, refractive valueErdogan HAkingol ZCam OSencan SDove Medical Pressarticlefluoresceinintraocular pressureGoldmann applanation tonometrynoncontact tonometrycentral corneal thicknessrefractive valueOphthalmologyRE1-994ENClinical Ophthalmology, Vol Volume 12, Pp 2183-2188 (2018)
institution DOAJ
collection DOAJ
language EN
topic fluorescein
intraocular pressure
Goldmann applanation tonometry
noncontact tonometry
central corneal thickness
refractive value
Ophthalmology
RE1-994
spellingShingle fluorescein
intraocular pressure
Goldmann applanation tonometry
noncontact tonometry
central corneal thickness
refractive value
Ophthalmology
RE1-994
Erdogan H
Akingol Z
Cam O
Sencan S
A comparison of NCT, Goldman application tonometry values with and without fluorescein
description Hakika Erdogan, Ziya Akingol, Ozlem Cam, Sadik Sencan Department of Ophthalmology, Faculty of Medicine, Maltepe University, Maltepe, Istanbul, Turkey  Purpose: The aim of the study was to statistically compare intraocular pressure (IOP) values measured using noncontact tonometer (NCT), Goldmann applanation tonometry (GAT) with fluorescein (fGAT), and GAT without fluorescein (nGAT). The study was also performed to test whether the values obtained using each technique change in accordance with the central corneal thickness (CCT) and refractive and keratometric values.Study design: This study was a prospective study of 188 eyes of 94 healthy volunteers.Methods: IOP was measured using fGAT, nGAT and NCT. CCT, refractive values, and keratometric values were measured, and the correlations and differences in the IOP for each tonometer were investigated.Results: The mean IOP values obtained with the NCT, nGAT, and fGAT were 17.5±3.7, 12.3±2.7, and 12.5±2 mmHg. The mean CCT was 538.2±34.4 µm, the mean refractive value was 0.9±1.2 D, and the mean keratometric value was 43.5±1.5 D. NCT was positively correlated with fGAT and GAT values and was significantly higher than both the values. There were no differences between fGAT and GAT values. No correlation was observed between the CCT and keratometric and refractive values and the difference between NCT and nGAT or fGAT.Conclusion: Differences in the measurements obtained using nGAT and fGAT were insignificant (P>0.05). Both values were positively correlated with NCT measurements (r=0.354, P<0.05) and were independent of CCT, keratometry, and refraction values. nGAT appears to be suitable for use in routine clinic practice. Keywords: fluorescein, intraocular pressure, Goldmann applanation tonometry, noncontact tonometry, central corneal thickness, refractive value
format article
author Erdogan H
Akingol Z
Cam O
Sencan S
author_facet Erdogan H
Akingol Z
Cam O
Sencan S
author_sort Erdogan H
title A comparison of NCT, Goldman application tonometry values with and without fluorescein
title_short A comparison of NCT, Goldman application tonometry values with and without fluorescein
title_full A comparison of NCT, Goldman application tonometry values with and without fluorescein
title_fullStr A comparison of NCT, Goldman application tonometry values with and without fluorescein
title_full_unstemmed A comparison of NCT, Goldman application tonometry values with and without fluorescein
title_sort comparison of nct, goldman application tonometry values with and without fluorescein
publisher Dove Medical Press
publishDate 2018
url https://doaj.org/article/ae5c2da0a09b4169bbc9eee84cb16989
work_keys_str_mv AT erdoganh acomparisonofnctgoldmanapplicationtonometryvalueswithandwithoutfluorescein
AT akingolz acomparisonofnctgoldmanapplicationtonometryvalueswithandwithoutfluorescein
AT camo acomparisonofnctgoldmanapplicationtonometryvalueswithandwithoutfluorescein
AT sencans acomparisonofnctgoldmanapplicationtonometryvalueswithandwithoutfluorescein
AT erdoganh comparisonofnctgoldmanapplicationtonometryvalueswithandwithoutfluorescein
AT akingolz comparisonofnctgoldmanapplicationtonometryvalueswithandwithoutfluorescein
AT camo comparisonofnctgoldmanapplicationtonometryvalueswithandwithoutfluorescein
AT sencans comparisonofnctgoldmanapplicationtonometryvalueswithandwithoutfluorescein
_version_ 1718402548807761920