Accuracy of two pulse-oximetry measurements for INTELLiVENT-ASV in mechanically ventilated patients: a prospective observational study

Abstract Recently, maintaining a certain oxygen saturation measured by pulse oximetry (SpO2) range in mechanically ventilated patients was recommended; attaching the INTELLiVENT-ASV to ventilators might be beneficial. We evaluated the SpO2 measurement accuracy of a Nihon Kohden and a Masimo monitor...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Shinshu Katayama, Jun Shima, Ken Tonai, Kansuke Koyama, Shin Nunomiya
Formato: article
Lenguaje:EN
Publicado: Nature Portfolio 2021
Materias:
R
Q
Acceso en línea:https://doaj.org/article/b56b877f037347dcb9a42cfd749cb215
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:b56b877f037347dcb9a42cfd749cb215
record_format dspace
spelling oai:doaj.org-article:b56b877f037347dcb9a42cfd749cb2152021-12-02T16:55:46ZAccuracy of two pulse-oximetry measurements for INTELLiVENT-ASV in mechanically ventilated patients: a prospective observational study10.1038/s41598-021-88608-72045-2322https://doaj.org/article/b56b877f037347dcb9a42cfd749cb2152021-04-01T00:00:00Zhttps://doi.org/10.1038/s41598-021-88608-7https://doaj.org/toc/2045-2322Abstract Recently, maintaining a certain oxygen saturation measured by pulse oximetry (SpO2) range in mechanically ventilated patients was recommended; attaching the INTELLiVENT-ASV to ventilators might be beneficial. We evaluated the SpO2 measurement accuracy of a Nihon Kohden and a Masimo monitor compared to actual arterial oxygen saturation (SaO2). SpO2 was simultaneously measured by a Nihon Kohden and Masimo monitor in patients consecutively admitted to a general intensive care unit and mechanically ventilated. Bland–Altman plots were used to compare measured SpO2 with actual SaO2. One hundred mechanically ventilated patients and 1497 arterial blood gas results were reviewed. Mean SaO2 values, Nihon Kohden SpO2 measurements, and Masimo SpO2 measurements were 95.7%, 96.4%, and 96.9%, respectively. The Nihon Kohden SpO2 measurements were less biased than Masimo measurements; their precision was not significantly different. Nihon Kohden and Masimo SpO2 measurements were not significantly different in the “SaO2 < 94%” group (P = 0.083). In the “94% ≤ SaO2 < 98%” and “SaO2 ≥ 98%” groups, there were significant differences between the Nihon Kohden and Masimo SpO2 measurements (P < 0.0001; P = 0.006; respectively). Therefore, when using automatically controlling oxygenation with INTELLiVENT-ASV in mechanically ventilated patients, the Nihon Kohden SpO2 sensor is preferable. Trial registration UMIN000027671. Registered 7 June 2017.Shinshu KatayamaJun ShimaKen TonaiKansuke KoyamaShin NunomiyaNature PortfolioarticleMedicineRScienceQENScientific Reports, Vol 11, Iss 1, Pp 1-9 (2021)
institution DOAJ
collection DOAJ
language EN
topic Medicine
R
Science
Q
spellingShingle Medicine
R
Science
Q
Shinshu Katayama
Jun Shima
Ken Tonai
Kansuke Koyama
Shin Nunomiya
Accuracy of two pulse-oximetry measurements for INTELLiVENT-ASV in mechanically ventilated patients: a prospective observational study
description Abstract Recently, maintaining a certain oxygen saturation measured by pulse oximetry (SpO2) range in mechanically ventilated patients was recommended; attaching the INTELLiVENT-ASV to ventilators might be beneficial. We evaluated the SpO2 measurement accuracy of a Nihon Kohden and a Masimo monitor compared to actual arterial oxygen saturation (SaO2). SpO2 was simultaneously measured by a Nihon Kohden and Masimo monitor in patients consecutively admitted to a general intensive care unit and mechanically ventilated. Bland–Altman plots were used to compare measured SpO2 with actual SaO2. One hundred mechanically ventilated patients and 1497 arterial blood gas results were reviewed. Mean SaO2 values, Nihon Kohden SpO2 measurements, and Masimo SpO2 measurements were 95.7%, 96.4%, and 96.9%, respectively. The Nihon Kohden SpO2 measurements were less biased than Masimo measurements; their precision was not significantly different. Nihon Kohden and Masimo SpO2 measurements were not significantly different in the “SaO2 < 94%” group (P = 0.083). In the “94% ≤ SaO2 < 98%” and “SaO2 ≥ 98%” groups, there were significant differences between the Nihon Kohden and Masimo SpO2 measurements (P < 0.0001; P = 0.006; respectively). Therefore, when using automatically controlling oxygenation with INTELLiVENT-ASV in mechanically ventilated patients, the Nihon Kohden SpO2 sensor is preferable. Trial registration UMIN000027671. Registered 7 June 2017.
format article
author Shinshu Katayama
Jun Shima
Ken Tonai
Kansuke Koyama
Shin Nunomiya
author_facet Shinshu Katayama
Jun Shima
Ken Tonai
Kansuke Koyama
Shin Nunomiya
author_sort Shinshu Katayama
title Accuracy of two pulse-oximetry measurements for INTELLiVENT-ASV in mechanically ventilated patients: a prospective observational study
title_short Accuracy of two pulse-oximetry measurements for INTELLiVENT-ASV in mechanically ventilated patients: a prospective observational study
title_full Accuracy of two pulse-oximetry measurements for INTELLiVENT-ASV in mechanically ventilated patients: a prospective observational study
title_fullStr Accuracy of two pulse-oximetry measurements for INTELLiVENT-ASV in mechanically ventilated patients: a prospective observational study
title_full_unstemmed Accuracy of two pulse-oximetry measurements for INTELLiVENT-ASV in mechanically ventilated patients: a prospective observational study
title_sort accuracy of two pulse-oximetry measurements for intellivent-asv in mechanically ventilated patients: a prospective observational study
publisher Nature Portfolio
publishDate 2021
url https://doaj.org/article/b56b877f037347dcb9a42cfd749cb215
work_keys_str_mv AT shinshukatayama accuracyoftwopulseoximetrymeasurementsforintelliventasvinmechanicallyventilatedpatientsaprospectiveobservationalstudy
AT junshima accuracyoftwopulseoximetrymeasurementsforintelliventasvinmechanicallyventilatedpatientsaprospectiveobservationalstudy
AT kentonai accuracyoftwopulseoximetrymeasurementsforintelliventasvinmechanicallyventilatedpatientsaprospectiveobservationalstudy
AT kansukekoyama accuracyoftwopulseoximetrymeasurementsforintelliventasvinmechanicallyventilatedpatientsaprospectiveobservationalstudy
AT shinnunomiya accuracyoftwopulseoximetrymeasurementsforintelliventasvinmechanicallyventilatedpatientsaprospectiveobservationalstudy
_version_ 1718382821376000000