Accuracy comparison study of new smartphone-based semen analyzer versus laboratory sperm quality analyzer

Purpose: This study aimed to test the clinical efficacy of a portable smartphone-based App assisted semen analysis (SA) system, O'VIEW-M PRO® to clinically accurate in comparison with results of laboratory-based conventional semen analyses including manual microscopic and computer-assisted seme...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Min Jung Park, Mi Young Lim, Hyun Jun Park, Nam Cheol Park
Formato: article
Lenguaje:EN
Publicado: Korean Urological Association 2021
Materias:
Acceso en línea:https://doaj.org/article/ba8ede503d6e4bdbbb8ea00513c4f47e
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:ba8ede503d6e4bdbbb8ea00513c4f47e
record_format dspace
spelling oai:doaj.org-article:ba8ede503d6e4bdbbb8ea00513c4f47e2021-11-10T04:59:18ZAccuracy comparison study of new smartphone-based semen analyzer versus laboratory sperm quality analyzer10.4111/icu.202102662466-04932466-054Xhttps://doaj.org/article/ba8ede503d6e4bdbbb8ea00513c4f47e2021-11-01T00:00:00Zhttps://icurology.org/pdf/10.4111/icu.20210266https://doaj.org/toc/2466-0493https://doaj.org/toc/2466-054XPurpose: This study aimed to test the clinical efficacy of a portable smartphone-based App assisted semen analysis (SA) system, O'VIEW-M PRO® to clinically accurate in comparison with results of laboratory-based conventional semen analyses including manual microscopic and computer-assisted semen analysis (CASA) for self-evaluation of seminal parameters. Materials and Methods: From January to May 2021, a total of 39 semen samples were analyzed for the sperm concentration and motility with new smartphone-based App assisted semen analyzer, O'VIEW-M PRO®, and results compared with those from laboratory-based manual microscopic SA with Makler Counting Chamber and CASA. Results: The coefficient factors among the results of the measurement with Makler chamber and laboratory-based CASA comparing to O'VIEW-M PRO® were 0.666 and 0.655 for sperm density, 0.662 and 0.658 for sperm motility, respectively. There were no particular problems with clinical use of the O'VIEW-M PRO®. Device performance in classifying samples is positive (<15×106 sperm/mL) and negative (>15×106 sperm/mL) for sperm concentration criteria, and positive (<40%) and negative (>40%) for sperm motility criteria. The smartphone-based App assisted SA O'VIEW-M PRO® showed a sensitivity of 92.6%, a specificity of 66.7%, and overall accuracy rate of 84.6%. Conclusions: This study shows a novel smartphone-based App assisted SA system. O'VIEW-M PRO® can easily obtain semen parameter information through self-diagnosis at home and induce infertile men's treatment and help patients after receiving infertile men's treatment before receiving treatment.Min Jung ParkMi Young LimHyun Jun Park Nam Cheol ParkKorean Urological Associationarticlemale infertilitysemen analysissmartphonesperm countDiseases of the genitourinary system. UrologyRC870-923ENInvestigative and Clinical Urology, Vol 62, Iss 6, Pp 672-680 (2021)
institution DOAJ
collection DOAJ
language EN
topic male infertility
semen analysis
smartphone
sperm count
Diseases of the genitourinary system. Urology
RC870-923
spellingShingle male infertility
semen analysis
smartphone
sperm count
Diseases of the genitourinary system. Urology
RC870-923
Min Jung Park
Mi Young Lim
Hyun Jun Park
Nam Cheol Park
Accuracy comparison study of new smartphone-based semen analyzer versus laboratory sperm quality analyzer
description Purpose: This study aimed to test the clinical efficacy of a portable smartphone-based App assisted semen analysis (SA) system, O'VIEW-M PRO® to clinically accurate in comparison with results of laboratory-based conventional semen analyses including manual microscopic and computer-assisted semen analysis (CASA) for self-evaluation of seminal parameters. Materials and Methods: From January to May 2021, a total of 39 semen samples were analyzed for the sperm concentration and motility with new smartphone-based App assisted semen analyzer, O'VIEW-M PRO®, and results compared with those from laboratory-based manual microscopic SA with Makler Counting Chamber and CASA. Results: The coefficient factors among the results of the measurement with Makler chamber and laboratory-based CASA comparing to O'VIEW-M PRO® were 0.666 and 0.655 for sperm density, 0.662 and 0.658 for sperm motility, respectively. There were no particular problems with clinical use of the O'VIEW-M PRO®. Device performance in classifying samples is positive (<15×106 sperm/mL) and negative (>15×106 sperm/mL) for sperm concentration criteria, and positive (<40%) and negative (>40%) for sperm motility criteria. The smartphone-based App assisted SA O'VIEW-M PRO® showed a sensitivity of 92.6%, a specificity of 66.7%, and overall accuracy rate of 84.6%. Conclusions: This study shows a novel smartphone-based App assisted SA system. O'VIEW-M PRO® can easily obtain semen parameter information through self-diagnosis at home and induce infertile men's treatment and help patients after receiving infertile men's treatment before receiving treatment.
format article
author Min Jung Park
Mi Young Lim
Hyun Jun Park
Nam Cheol Park
author_facet Min Jung Park
Mi Young Lim
Hyun Jun Park
Nam Cheol Park
author_sort Min Jung Park
title Accuracy comparison study of new smartphone-based semen analyzer versus laboratory sperm quality analyzer
title_short Accuracy comparison study of new smartphone-based semen analyzer versus laboratory sperm quality analyzer
title_full Accuracy comparison study of new smartphone-based semen analyzer versus laboratory sperm quality analyzer
title_fullStr Accuracy comparison study of new smartphone-based semen analyzer versus laboratory sperm quality analyzer
title_full_unstemmed Accuracy comparison study of new smartphone-based semen analyzer versus laboratory sperm quality analyzer
title_sort accuracy comparison study of new smartphone-based semen analyzer versus laboratory sperm quality analyzer
publisher Korean Urological Association
publishDate 2021
url https://doaj.org/article/ba8ede503d6e4bdbbb8ea00513c4f47e
work_keys_str_mv AT minjungpark accuracycomparisonstudyofnewsmartphonebasedsemenanalyzerversuslaboratoryspermqualityanalyzer
AT miyounglim accuracycomparisonstudyofnewsmartphonebasedsemenanalyzerversuslaboratoryspermqualityanalyzer
AT hyunjunpark accuracycomparisonstudyofnewsmartphonebasedsemenanalyzerversuslaboratoryspermqualityanalyzer
AT namcheolpark accuracycomparisonstudyofnewsmartphonebasedsemenanalyzerversuslaboratoryspermqualityanalyzer
_version_ 1718440592051011584