Dosimetry, control and irradiation of rat cervical spinal cord

Background and Objective: In radiobiology, the most important physical radiation quantity to predict the effect of irradiation of a biological specimen is the absorbed dose to the tissue of interest. In this project an irradiation set up was designed and verification of set up and dosimetry procedur...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: SR Mahdavi, M Zabih Zadeh, AR Shirazi, H Gharaati
Formato: article
Lenguaje:EN
FA
Publicado: Babol University of Medical Sciences 2005
Materias:
jig
R
Acceso en línea:https://doaj.org/article/bc9cfb368ce64c24b6f5d702877b7e10
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:bc9cfb368ce64c24b6f5d702877b7e10
record_format dspace
spelling oai:doaj.org-article:bc9cfb368ce64c24b6f5d702877b7e102021-11-10T09:14:32ZDosimetry, control and irradiation of rat cervical spinal cord1561-41072251-7170https://doaj.org/article/bc9cfb368ce64c24b6f5d702877b7e102005-04-01T00:00:00Zhttp://jbums.org/article-1-2630-en.htmlhttps://doaj.org/toc/1561-4107https://doaj.org/toc/2251-7170Background and Objective: In radiobiology, the most important physical radiation quantity to predict the effect of irradiation of a biological specimen is the absorbed dose to the tissue of interest. In this project an irradiation set up was designed and verification of set up and dosimetry procedure was performed to deliver a precise X-ray dose to the small field of rat cervical spinal cord. Methods: AAPM-TG61 (American association and physician in medicine radiation therapy committee task group 61) protocol was used for dose measurement and we tailored a rat like phantom from polyethylene for dosimetry verification. The ionization chamber in this study was a farmer type and the X-ray generator was an orthovoltage Siemens machine working at 200 KVp potential. Dosimetry was done in air and phantom. An special jig was also built to fix the animals during irradiation which could help the treatments to be reproducible. Simulation and portal films were obtained to verify the irradiation field. Findings: The average value of dose rate in specified geometries by measurement was 146.54 cGy/min (SD=0.109). While the dose determined by calculation was 95.145 cGy/min (SD=0.105), the comparison between these 2 methods shown a small discrepancy of 0.50% (P<0.001), which lies within the error limit of ±5.3% as mentioned by ICRU. Conclusion: Using protocol of AAPM TG-61 can provide an accurate dosimetry with minimum ambiguity. Application of appropriate correction factors and protocols can increase the accuracy and decrease the irradiation errors in radiobiological studies.SR MahdaviM Zabih ZadehAR ShiraziH GharaatiBabol University of Medical Sciencesarticlex-ray dosimetryradiobiologyrat cervical spinal cord irradiationrat phantomjigMedicineRMedicine (General)R5-920ENFAMajallah-i Dānishgāh-i ̒Ulūm-i Pizishkī-i Bābul, Vol 7, Iss 2, Pp 7-13 (2005)
institution DOAJ
collection DOAJ
language EN
FA
topic x-ray dosimetry
radiobiology
rat cervical spinal cord irradiation
rat phantom
jig
Medicine
R
Medicine (General)
R5-920
spellingShingle x-ray dosimetry
radiobiology
rat cervical spinal cord irradiation
rat phantom
jig
Medicine
R
Medicine (General)
R5-920
SR Mahdavi
M Zabih Zadeh
AR Shirazi
H Gharaati
Dosimetry, control and irradiation of rat cervical spinal cord
description Background and Objective: In radiobiology, the most important physical radiation quantity to predict the effect of irradiation of a biological specimen is the absorbed dose to the tissue of interest. In this project an irradiation set up was designed and verification of set up and dosimetry procedure was performed to deliver a precise X-ray dose to the small field of rat cervical spinal cord. Methods: AAPM-TG61 (American association and physician in medicine radiation therapy committee task group 61) protocol was used for dose measurement and we tailored a rat like phantom from polyethylene for dosimetry verification. The ionization chamber in this study was a farmer type and the X-ray generator was an orthovoltage Siemens machine working at 200 KVp potential. Dosimetry was done in air and phantom. An special jig was also built to fix the animals during irradiation which could help the treatments to be reproducible. Simulation and portal films were obtained to verify the irradiation field. Findings: The average value of dose rate in specified geometries by measurement was 146.54 cGy/min (SD=0.109). While the dose determined by calculation was 95.145 cGy/min (SD=0.105), the comparison between these 2 methods shown a small discrepancy of 0.50% (P<0.001), which lies within the error limit of ±5.3% as mentioned by ICRU. Conclusion: Using protocol of AAPM TG-61 can provide an accurate dosimetry with minimum ambiguity. Application of appropriate correction factors and protocols can increase the accuracy and decrease the irradiation errors in radiobiological studies.
format article
author SR Mahdavi
M Zabih Zadeh
AR Shirazi
H Gharaati
author_facet SR Mahdavi
M Zabih Zadeh
AR Shirazi
H Gharaati
author_sort SR Mahdavi
title Dosimetry, control and irradiation of rat cervical spinal cord
title_short Dosimetry, control and irradiation of rat cervical spinal cord
title_full Dosimetry, control and irradiation of rat cervical spinal cord
title_fullStr Dosimetry, control and irradiation of rat cervical spinal cord
title_full_unstemmed Dosimetry, control and irradiation of rat cervical spinal cord
title_sort dosimetry, control and irradiation of rat cervical spinal cord
publisher Babol University of Medical Sciences
publishDate 2005
url https://doaj.org/article/bc9cfb368ce64c24b6f5d702877b7e10
work_keys_str_mv AT srmahdavi dosimetrycontrolandirradiationofratcervicalspinalcord
AT mzabihzadeh dosimetrycontrolandirradiationofratcervicalspinalcord
AT arshirazi dosimetrycontrolandirradiationofratcervicalspinalcord
AT hgharaati dosimetrycontrolandirradiationofratcervicalspinalcord
_version_ 1718440229566676992