Needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice)
Abstract This method comparison study used radiographs of 20 mice and 20 budgerigars to investigate comparability between computed radiography (CR) and high-resolution screen-film systems and study the effects of reduced radiation doses on image quality of digital radiographs of small patients. Expo...