Comparison of corneal biomechanics after myopic small-incision lenticule extraction compared to LASIK: an ex vivo study

Anastasios John Kanellopoulos1,2 1Department of Ophthalmology, Laservision Clinical and Research Institute, Athens, Greece; 2Department of Ophthalmology, New York University Medical School, New York, NY, USA Purpose: To investigate ex vivo potentially different corneal biomechanical properties aft...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autor principal: Kanellopoulos AJ
Formato: article
Lenguaje:EN
Publicado: Dove Medical Press 2018
Materias:
Acceso en línea:https://doaj.org/article/bec2d13d8eb746d49a7c5d3a9e4eeded
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:bec2d13d8eb746d49a7c5d3a9e4eeded
record_format dspace
spelling oai:doaj.org-article:bec2d13d8eb746d49a7c5d3a9e4eeded2021-12-02T05:19:42ZComparison of corneal biomechanics after myopic small-incision lenticule extraction compared to LASIK: an ex vivo study1177-5483https://doaj.org/article/bec2d13d8eb746d49a7c5d3a9e4eeded2018-01-01T00:00:00Zhttps://www.dovepress.com/comparison-of-corneal-biomechanics-after-myopic-small-incision-lenticu-peer-reviewed-article-OPTHhttps://doaj.org/toc/1177-5483Anastasios John Kanellopoulos1,2 1Department of Ophthalmology, Laservision Clinical and Research Institute, Athens, Greece; 2Department of Ophthalmology, New York University Medical School, New York, NY, USA Purpose: To investigate ex vivo potentially different corneal biomechanical properties after small-incision lenticule extraction (SMILE) versus LASIK for myopic correction.Methods: Thirty human donor corneas were subjected to either myopic SMILE or femtosecond laser-assisted LASIK. Donor corneas were assigned to six investigative groups: Group A, –3.00 D (diopters) SMILE; Group B, –8.00 D SMILE; Group C, –3.00 D LASIK; and Group D, –8.00 D LASIK. Additionally, two control groups were formed: Group E, SMILE and Group F, LASIK. All groups consisted of five corneas, randomly allocated. The corneas in the control groups were subjected to the corresponding femtosecond-laser lamellar cuts but not to tissue removal. Evaluation of biomechanical tensile strength was conducted by biaxial force application. Primary outcome measures were stress at 10% and 15% strain, and Young’s modulus at 10% and 15% strain.Results: In SMILE, the average relative difference (Δ) of the four outcome measures was –34.46% for –3.00 D correction versus control Group E and –49.34% for –8.00 D correction versus control Group E. In LASIK, average Δ was –24.88% for –3.00 D correction versus control, and –52.73% for –8.00 D correction versus control. All these differences were statistically significant; SMILE compared to LASIK for the same myopic correction appears to result in more biomechanical reduction for –3.00 D corrections by –26%, while a nonstatistically significant difference was noted in –8.00 D corrections.Conclusion: Both SMILE and LASIK procedures do substantially alter corneal biomechanical properties, and the degree of tensile strength reduction is statistically significantly correlated to the extent of myopic correction. Additionally, SMILE procedure seems to result in more tensile strength reduction in lower myopic corrections compared to LASIK, and similar tensile strength reduction to LASIK in higher myopic corrections when compared to LASIK. Keywords: SMILE, femto-second LASIK, myopic LASIK, corneal biomechanics, Young’s modulus, corneal stress–strain, shear strength, tensile strengthKanellopoulos AJDove Medical PressarticleSMILEFemto-second LASIKMyopic LASIKCorneal biomechanicsYoung’s modulusCorneal stress-strainShear strengthTensile strengthOphthalmologyRE1-994ENClinical Ophthalmology, Vol Volume 12, Pp 237-245 (2018)
institution DOAJ
collection DOAJ
language EN
topic SMILE
Femto-second LASIK
Myopic LASIK
Corneal biomechanics
Young’s modulus
Corneal stress-strain
Shear strength
Tensile strength
Ophthalmology
RE1-994
spellingShingle SMILE
Femto-second LASIK
Myopic LASIK
Corneal biomechanics
Young’s modulus
Corneal stress-strain
Shear strength
Tensile strength
Ophthalmology
RE1-994
Kanellopoulos AJ
Comparison of corneal biomechanics after myopic small-incision lenticule extraction compared to LASIK: an ex vivo study
description Anastasios John Kanellopoulos1,2 1Department of Ophthalmology, Laservision Clinical and Research Institute, Athens, Greece; 2Department of Ophthalmology, New York University Medical School, New York, NY, USA Purpose: To investigate ex vivo potentially different corneal biomechanical properties after small-incision lenticule extraction (SMILE) versus LASIK for myopic correction.Methods: Thirty human donor corneas were subjected to either myopic SMILE or femtosecond laser-assisted LASIK. Donor corneas were assigned to six investigative groups: Group A, –3.00 D (diopters) SMILE; Group B, –8.00 D SMILE; Group C, –3.00 D LASIK; and Group D, –8.00 D LASIK. Additionally, two control groups were formed: Group E, SMILE and Group F, LASIK. All groups consisted of five corneas, randomly allocated. The corneas in the control groups were subjected to the corresponding femtosecond-laser lamellar cuts but not to tissue removal. Evaluation of biomechanical tensile strength was conducted by biaxial force application. Primary outcome measures were stress at 10% and 15% strain, and Young’s modulus at 10% and 15% strain.Results: In SMILE, the average relative difference (Δ) of the four outcome measures was –34.46% for –3.00 D correction versus control Group E and –49.34% for –8.00 D correction versus control Group E. In LASIK, average Δ was –24.88% for –3.00 D correction versus control, and –52.73% for –8.00 D correction versus control. All these differences were statistically significant; SMILE compared to LASIK for the same myopic correction appears to result in more biomechanical reduction for –3.00 D corrections by –26%, while a nonstatistically significant difference was noted in –8.00 D corrections.Conclusion: Both SMILE and LASIK procedures do substantially alter corneal biomechanical properties, and the degree of tensile strength reduction is statistically significantly correlated to the extent of myopic correction. Additionally, SMILE procedure seems to result in more tensile strength reduction in lower myopic corrections compared to LASIK, and similar tensile strength reduction to LASIK in higher myopic corrections when compared to LASIK. Keywords: SMILE, femto-second LASIK, myopic LASIK, corneal biomechanics, Young’s modulus, corneal stress–strain, shear strength, tensile strength
format article
author Kanellopoulos AJ
author_facet Kanellopoulos AJ
author_sort Kanellopoulos AJ
title Comparison of corneal biomechanics after myopic small-incision lenticule extraction compared to LASIK: an ex vivo study
title_short Comparison of corneal biomechanics after myopic small-incision lenticule extraction compared to LASIK: an ex vivo study
title_full Comparison of corneal biomechanics after myopic small-incision lenticule extraction compared to LASIK: an ex vivo study
title_fullStr Comparison of corneal biomechanics after myopic small-incision lenticule extraction compared to LASIK: an ex vivo study
title_full_unstemmed Comparison of corneal biomechanics after myopic small-incision lenticule extraction compared to LASIK: an ex vivo study
title_sort comparison of corneal biomechanics after myopic small-incision lenticule extraction compared to lasik: an ex vivo study
publisher Dove Medical Press
publishDate 2018
url https://doaj.org/article/bec2d13d8eb746d49a7c5d3a9e4eeded
work_keys_str_mv AT kanellopoulosaj comparisonofcornealbiomechanicsaftermyopicsmallincisionlenticuleextractioncomparedtolasikanexvivostudy
_version_ 1718400495312175104