Control of the expiratory flow in a lung model and in healthy volunteers with an adjustable flow regulator: a combined bench and randomized crossover study

Abstract Background Pursed-lips breathing (PLB) is a technique to attenuate small airway collapse by regulating the expiratory flow. During mandatory ventilation, flow-controlled expiration (FLEX), which mimics the expiratory flow course of PLB utilizing a digital system for measurement and control,...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Johannes Schmidt, Anna Martin, Christin Wenzel, Jonas Weber, Steffen Wirth, Stefan Schumann
Formato: article
Lenguaje:EN
Publicado: BMC 2021
Materias:
Acceso en línea:https://doaj.org/article/c3e4e2b9ced14d68bbf389d2582ffc3e
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:c3e4e2b9ced14d68bbf389d2582ffc3e
record_format dspace
spelling oai:doaj.org-article:c3e4e2b9ced14d68bbf389d2582ffc3e2021-11-21T12:31:55ZControl of the expiratory flow in a lung model and in healthy volunteers with an adjustable flow regulator: a combined bench and randomized crossover study10.1186/s12931-021-01886-71465-993Xhttps://doaj.org/article/c3e4e2b9ced14d68bbf389d2582ffc3e2021-11-01T00:00:00Zhttps://doi.org/10.1186/s12931-021-01886-7https://doaj.org/toc/1465-993XAbstract Background Pursed-lips breathing (PLB) is a technique to attenuate small airway collapse by regulating the expiratory flow. During mandatory ventilation, flow-controlled expiration (FLEX), which mimics the expiratory flow course of PLB utilizing a digital system for measurement and control, was shown to exert lung protective effects. However, PLB requires a patient’s participation and coordinated muscular effort and FLEX requires a complex technical setup. Here, we present an adjustable flow regulator to mimic PLB and FLEX, respectively, without the need of a patient’s participation, or a complex technical device. Methods Our study consisted of two parts: First, in a lung model which was ventilated with standard settings (tidal volume 500 ml, respiratory rate 12 min−1, positive end-expiratory pressure (PEEP) 5 cmH2O), the possible reduction of the maximal expiratory flow by utilizing the flow regulator was assessed. Second, with spontaneously breathing healthy volunteers, the short-term effects of medium and strong expiratory flow reduction on airway pressure, the change of end-expiratory lung volume (EELV), and breathing discomfort was investigated. Results In the lung model experiments, expiratory flow could be reduced from − 899 ± 9 ml·s−1 down to − 328 ± 25 ml·s−1. Thereby, inspiratory variables and PEEP were unaffected. In the volunteers, the maximal expiratory flow of − 574 ± 131 ml·s−1 under baseline conditions was reduced to − 395 ± 71 ml·s−1 for medium flow regulation and to − 266 ± 58 ml·s−1 for strong flow regulation, respectively (p < 0.001). Accordingly, mean airway pressure increased from 0.6 ± 0.1 cmH2O to 2.9 ± 0.4 cmH2O with medium flow regulation and to 5.4 ± 2.4 cmH2O with strong flow regulation, respectively (p < 0.001). The EELV increased from baseline by 31 ± 458 ml for medium flow regulation and 320 ± 681 ml for strong flow regulation (p = 0.033). The participants rated breathing with the flow regulator as moderately uncomfortable, but none rated breathing with the flow regulator as intolerable. Conclusions The flow regulator represents an adjustable device for application of a self-regulated expiratory resistive load, representing an alternative for PLB and FLEX. Future applications in spontaneously breathing patients and patients with mandatory ventilation alike may reveal potential benefits. Trial registration: DRKS00015296, registered on 20th August, 2018; URL: https://www.drks.de/drks_web/setLocale_EN.do .Johannes SchmidtAnna MartinChristin WenzelJonas WeberSteffen WirthStefan SchumannBMCarticleExpiratory flow regulationExpiratory resistive loadPursed-lips breathingSpontaneous breathingMandatory ventilationDiseases of the respiratory systemRC705-779ENRespiratory Research, Vol 22, Iss 1, Pp 1-10 (2021)
institution DOAJ
collection DOAJ
language EN
topic Expiratory flow regulation
Expiratory resistive load
Pursed-lips breathing
Spontaneous breathing
Mandatory ventilation
Diseases of the respiratory system
RC705-779
spellingShingle Expiratory flow regulation
Expiratory resistive load
Pursed-lips breathing
Spontaneous breathing
Mandatory ventilation
Diseases of the respiratory system
RC705-779
Johannes Schmidt
Anna Martin
Christin Wenzel
Jonas Weber
Steffen Wirth
Stefan Schumann
Control of the expiratory flow in a lung model and in healthy volunteers with an adjustable flow regulator: a combined bench and randomized crossover study
description Abstract Background Pursed-lips breathing (PLB) is a technique to attenuate small airway collapse by regulating the expiratory flow. During mandatory ventilation, flow-controlled expiration (FLEX), which mimics the expiratory flow course of PLB utilizing a digital system for measurement and control, was shown to exert lung protective effects. However, PLB requires a patient’s participation and coordinated muscular effort and FLEX requires a complex technical setup. Here, we present an adjustable flow regulator to mimic PLB and FLEX, respectively, without the need of a patient’s participation, or a complex technical device. Methods Our study consisted of two parts: First, in a lung model which was ventilated with standard settings (tidal volume 500 ml, respiratory rate 12 min−1, positive end-expiratory pressure (PEEP) 5 cmH2O), the possible reduction of the maximal expiratory flow by utilizing the flow regulator was assessed. Second, with spontaneously breathing healthy volunteers, the short-term effects of medium and strong expiratory flow reduction on airway pressure, the change of end-expiratory lung volume (EELV), and breathing discomfort was investigated. Results In the lung model experiments, expiratory flow could be reduced from − 899 ± 9 ml·s−1 down to − 328 ± 25 ml·s−1. Thereby, inspiratory variables and PEEP were unaffected. In the volunteers, the maximal expiratory flow of − 574 ± 131 ml·s−1 under baseline conditions was reduced to − 395 ± 71 ml·s−1 for medium flow regulation and to − 266 ± 58 ml·s−1 for strong flow regulation, respectively (p < 0.001). Accordingly, mean airway pressure increased from 0.6 ± 0.1 cmH2O to 2.9 ± 0.4 cmH2O with medium flow regulation and to 5.4 ± 2.4 cmH2O with strong flow regulation, respectively (p < 0.001). The EELV increased from baseline by 31 ± 458 ml for medium flow regulation and 320 ± 681 ml for strong flow regulation (p = 0.033). The participants rated breathing with the flow regulator as moderately uncomfortable, but none rated breathing with the flow regulator as intolerable. Conclusions The flow regulator represents an adjustable device for application of a self-regulated expiratory resistive load, representing an alternative for PLB and FLEX. Future applications in spontaneously breathing patients and patients with mandatory ventilation alike may reveal potential benefits. Trial registration: DRKS00015296, registered on 20th August, 2018; URL: https://www.drks.de/drks_web/setLocale_EN.do .
format article
author Johannes Schmidt
Anna Martin
Christin Wenzel
Jonas Weber
Steffen Wirth
Stefan Schumann
author_facet Johannes Schmidt
Anna Martin
Christin Wenzel
Jonas Weber
Steffen Wirth
Stefan Schumann
author_sort Johannes Schmidt
title Control of the expiratory flow in a lung model and in healthy volunteers with an adjustable flow regulator: a combined bench and randomized crossover study
title_short Control of the expiratory flow in a lung model and in healthy volunteers with an adjustable flow regulator: a combined bench and randomized crossover study
title_full Control of the expiratory flow in a lung model and in healthy volunteers with an adjustable flow regulator: a combined bench and randomized crossover study
title_fullStr Control of the expiratory flow in a lung model and in healthy volunteers with an adjustable flow regulator: a combined bench and randomized crossover study
title_full_unstemmed Control of the expiratory flow in a lung model and in healthy volunteers with an adjustable flow regulator: a combined bench and randomized crossover study
title_sort control of the expiratory flow in a lung model and in healthy volunteers with an adjustable flow regulator: a combined bench and randomized crossover study
publisher BMC
publishDate 2021
url https://doaj.org/article/c3e4e2b9ced14d68bbf389d2582ffc3e
work_keys_str_mv AT johannesschmidt controloftheexpiratoryflowinalungmodelandinhealthyvolunteerswithanadjustableflowregulatoracombinedbenchandrandomizedcrossoverstudy
AT annamartin controloftheexpiratoryflowinalungmodelandinhealthyvolunteerswithanadjustableflowregulatoracombinedbenchandrandomizedcrossoverstudy
AT christinwenzel controloftheexpiratoryflowinalungmodelandinhealthyvolunteerswithanadjustableflowregulatoracombinedbenchandrandomizedcrossoverstudy
AT jonasweber controloftheexpiratoryflowinalungmodelandinhealthyvolunteerswithanadjustableflowregulatoracombinedbenchandrandomizedcrossoverstudy
AT steffenwirth controloftheexpiratoryflowinalungmodelandinhealthyvolunteerswithanadjustableflowregulatoracombinedbenchandrandomizedcrossoverstudy
AT stefanschumann controloftheexpiratoryflowinalungmodelandinhealthyvolunteerswithanadjustableflowregulatoracombinedbenchandrandomizedcrossoverstudy
_version_ 1718418969105268736