Comparison between two devices in the detection of corneal thickness changes after cataract surgery
Abstract This study compared corneal thickness (CT) changes obtained with specular microscopy (SM) and a rotating Scheimpflug camera (RSC) after conventional phacoemulsification surgery (PS). One hundred sixty six eyes of 83 patients were analyzed before and one month after PS. One eye underwent PS,...
Guardado en:
Autores principales: | , , , , , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Nature Portfolio
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/c51f0aa8f4ab48ccb8c16accb662ea46 |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
id |
oai:doaj.org-article:c51f0aa8f4ab48ccb8c16accb662ea46 |
---|---|
record_format |
dspace |
spelling |
oai:doaj.org-article:c51f0aa8f4ab48ccb8c16accb662ea462021-12-02T17:04:34ZComparison between two devices in the detection of corneal thickness changes after cataract surgery10.1038/s41598-021-86158-62045-2322https://doaj.org/article/c51f0aa8f4ab48ccb8c16accb662ea462021-03-01T00:00:00Zhttps://doi.org/10.1038/s41598-021-86158-6https://doaj.org/toc/2045-2322Abstract This study compared corneal thickness (CT) changes obtained with specular microscopy (SM) and a rotating Scheimpflug camera (RSC) after conventional phacoemulsification surgery (PS). One hundred sixty six eyes of 83 patients were analyzed before and one month after PS. One eye underwent PS, while the fellow phakic one was used as control. CT was measured with SM at the center of the cornea and with RSC at the pupil center, at the corneal apex and at the thinnest point. In the operated eye, SM showed a larger CT mean increase than those one detected at the three different measurements’ points evaluated by RSC. Inversely, in the fellow phakic eye, SM showed a greater CT mean decrease than those one registered by RSC at its three measurement’s points. Thus, one month after surgery, even if cornea appears clear at the slit-lamp, a significant thickness increase is still present. This is even more evident if the slight decrease of the fellow phakic eye is considered. The differences between the two devices are probably related to the different measured areas.Maddalena De BernardoLivio VitielloGiulia AbbinanteIlaria De PascaleLuigi CapassoGiuseppe MarottaNicola RosaNature PortfolioarticleMedicineRScienceQENScientific Reports, Vol 11, Iss 1, Pp 1-8 (2021) |
institution |
DOAJ |
collection |
DOAJ |
language |
EN |
topic |
Medicine R Science Q |
spellingShingle |
Medicine R Science Q Maddalena De Bernardo Livio Vitiello Giulia Abbinante Ilaria De Pascale Luigi Capasso Giuseppe Marotta Nicola Rosa Comparison between two devices in the detection of corneal thickness changes after cataract surgery |
description |
Abstract This study compared corneal thickness (CT) changes obtained with specular microscopy (SM) and a rotating Scheimpflug camera (RSC) after conventional phacoemulsification surgery (PS). One hundred sixty six eyes of 83 patients were analyzed before and one month after PS. One eye underwent PS, while the fellow phakic one was used as control. CT was measured with SM at the center of the cornea and with RSC at the pupil center, at the corneal apex and at the thinnest point. In the operated eye, SM showed a larger CT mean increase than those one detected at the three different measurements’ points evaluated by RSC. Inversely, in the fellow phakic eye, SM showed a greater CT mean decrease than those one registered by RSC at its three measurement’s points. Thus, one month after surgery, even if cornea appears clear at the slit-lamp, a significant thickness increase is still present. This is even more evident if the slight decrease of the fellow phakic eye is considered. The differences between the two devices are probably related to the different measured areas. |
format |
article |
author |
Maddalena De Bernardo Livio Vitiello Giulia Abbinante Ilaria De Pascale Luigi Capasso Giuseppe Marotta Nicola Rosa |
author_facet |
Maddalena De Bernardo Livio Vitiello Giulia Abbinante Ilaria De Pascale Luigi Capasso Giuseppe Marotta Nicola Rosa |
author_sort |
Maddalena De Bernardo |
title |
Comparison between two devices in the detection of corneal thickness changes after cataract surgery |
title_short |
Comparison between two devices in the detection of corneal thickness changes after cataract surgery |
title_full |
Comparison between two devices in the detection of corneal thickness changes after cataract surgery |
title_fullStr |
Comparison between two devices in the detection of corneal thickness changes after cataract surgery |
title_full_unstemmed |
Comparison between two devices in the detection of corneal thickness changes after cataract surgery |
title_sort |
comparison between two devices in the detection of corneal thickness changes after cataract surgery |
publisher |
Nature Portfolio |
publishDate |
2021 |
url |
https://doaj.org/article/c51f0aa8f4ab48ccb8c16accb662ea46 |
work_keys_str_mv |
AT maddalenadebernardo comparisonbetweentwodevicesinthedetectionofcornealthicknesschangesaftercataractsurgery AT liviovitiello comparisonbetweentwodevicesinthedetectionofcornealthicknesschangesaftercataractsurgery AT giuliaabbinante comparisonbetweentwodevicesinthedetectionofcornealthicknesschangesaftercataractsurgery AT ilariadepascale comparisonbetweentwodevicesinthedetectionofcornealthicknesschangesaftercataractsurgery AT luigicapasso comparisonbetweentwodevicesinthedetectionofcornealthicknesschangesaftercataractsurgery AT giuseppemarotta comparisonbetweentwodevicesinthedetectionofcornealthicknesschangesaftercataractsurgery AT nicolarosa comparisonbetweentwodevicesinthedetectionofcornealthicknesschangesaftercataractsurgery |
_version_ |
1718381874714247168 |