Designing and manufacturing a heat control system of X-ray tubes in multiple exposures

Objective: Over 1/3 to half of accurate medical diagnosis are based on x-ray examinations. The x-ray tube is the main part of a radiology set. X-ray production is a phenomenon with low efficiency and over 99.6% of the fast electrons energy changes to heat. This heat is also a limiting factor for the...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: A Shabestani Monfared, H Nezamabadi Pour, A Hashemoghli, A Aliasghar Zadeh
Formato: article
Lenguaje:EN
FA
Publicado: Babol University of Medical Sciences 2001
Materias:
R
Acceso en línea:https://doaj.org/article/d2652a8c805347b9bb0ff0d166bc6d0f
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:d2652a8c805347b9bb0ff0d166bc6d0f
record_format dspace
spelling oai:doaj.org-article:d2652a8c805347b9bb0ff0d166bc6d0f2021-11-10T09:21:13ZDesigning and manufacturing a heat control system of X-ray tubes in multiple exposures1561-41072251-7170https://doaj.org/article/d2652a8c805347b9bb0ff0d166bc6d0f2001-10-01T00:00:00Zhttp://jbums.org/article-1-2845-en.htmlhttps://doaj.org/toc/1561-4107https://doaj.org/toc/2251-7170Objective: Over 1/3 to half of accurate medical diagnosis are based on x-ray examinations. The x-ray tube is the main part of a radiology set. X-ray production is a phenomenon with low efficiency and over 99.6% of the fast electrons energy changes to heat. This heat is also a limiting factor for the system function. The aim of this study was to design and manufacture an automatic heat control system in radiology tubes for multiple exposures. These exposures are allowed with regard to overload system but inappropriate intervaling in exposures causes heat damages in anode. Methods: The hardware of system was designed and manufactured by using of micro-controller system. Its software was written by 8051 Franklin preview software package and on the basis of anode cooling chart data and installed on micro-controller by A1-11 system. Findings: The designed system can calculate and display anode heat at every moment. Functioning test in 100 experimental exposures showed a reliability of 99.92±0.07%. Conclusion: This system can be used as a suitable equipment in protecting and preventing radiology tubes against heat damages after frequent exposures.A Shabestani MonfaredH Nezamabadi PourA HashemoghliA Aliasghar ZadehBabol University of Medical Sciencesarticlex-ray tubeheat damagesanodemultiple exposuresMedicineRMedicine (General)R5-920ENFAMajallah-i Dānishgāh-i ̒Ulūm-i Pizishkī-i Bābul, Vol 3, Iss 4, Pp 7-10 (2001)
institution DOAJ
collection DOAJ
language EN
FA
topic x-ray tube
heat damages
anode
multiple exposures
Medicine
R
Medicine (General)
R5-920
spellingShingle x-ray tube
heat damages
anode
multiple exposures
Medicine
R
Medicine (General)
R5-920
A Shabestani Monfared
H Nezamabadi Pour
A Hashemoghli
A Aliasghar Zadeh
Designing and manufacturing a heat control system of X-ray tubes in multiple exposures
description Objective: Over 1/3 to half of accurate medical diagnosis are based on x-ray examinations. The x-ray tube is the main part of a radiology set. X-ray production is a phenomenon with low efficiency and over 99.6% of the fast electrons energy changes to heat. This heat is also a limiting factor for the system function. The aim of this study was to design and manufacture an automatic heat control system in radiology tubes for multiple exposures. These exposures are allowed with regard to overload system but inappropriate intervaling in exposures causes heat damages in anode. Methods: The hardware of system was designed and manufactured by using of micro-controller system. Its software was written by 8051 Franklin preview software package and on the basis of anode cooling chart data and installed on micro-controller by A1-11 system. Findings: The designed system can calculate and display anode heat at every moment. Functioning test in 100 experimental exposures showed a reliability of 99.92±0.07%. Conclusion: This system can be used as a suitable equipment in protecting and preventing radiology tubes against heat damages after frequent exposures.
format article
author A Shabestani Monfared
H Nezamabadi Pour
A Hashemoghli
A Aliasghar Zadeh
author_facet A Shabestani Monfared
H Nezamabadi Pour
A Hashemoghli
A Aliasghar Zadeh
author_sort A Shabestani Monfared
title Designing and manufacturing a heat control system of X-ray tubes in multiple exposures
title_short Designing and manufacturing a heat control system of X-ray tubes in multiple exposures
title_full Designing and manufacturing a heat control system of X-ray tubes in multiple exposures
title_fullStr Designing and manufacturing a heat control system of X-ray tubes in multiple exposures
title_full_unstemmed Designing and manufacturing a heat control system of X-ray tubes in multiple exposures
title_sort designing and manufacturing a heat control system of x-ray tubes in multiple exposures
publisher Babol University of Medical Sciences
publishDate 2001
url https://doaj.org/article/d2652a8c805347b9bb0ff0d166bc6d0f
work_keys_str_mv AT ashabestanimonfared designingandmanufacturingaheatcontrolsystemofxraytubesinmultipleexposures
AT hnezamabadipour designingandmanufacturingaheatcontrolsystemofxraytubesinmultipleexposures
AT ahashemoghli designingandmanufacturingaheatcontrolsystemofxraytubesinmultipleexposures
AT aaliasgharzadeh designingandmanufacturingaheatcontrolsystemofxraytubesinmultipleexposures
_version_ 1718440141617364992