Convenience of Hgb-O detected by optical method in XN-series hematology analyzers in evaluating hemoglobin concentration in samples with chylous turbidity
Abstract The chylous turbidity of blood samples is one of the causes of false-high hemoglobin (Hgb) concentration measurements by the colorimetric method, which has been widely applied in hematology analyzers. In such cases, additional manual procedures are required to correct Hgb concentrations. We...
Guardado en:
Autores principales: | Yu Aruga, Chiaki Ikeda, Arisa Hanai, Sakiko Yoshimura, Momoko Kito, Satoe Miyaki, Misato Tsubokura, Yuka Yasuno, Chiaki Hayashi, Motoi Miyakoshi, Takahiro Nishino, Kimihiko Kawamura, Hiromichi Matsushita |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Nature Portfolio
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/d7d25f2affa04020b008fc99a79005bb |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
Ejemplares similares
-
Chylous ascites after laparoscopic donor nephrectomy: Case report
por: Gaurav Bhandari, et al.
Publicado: (2021) -
Performance of the hematology analyzer XN-31 prototype in the detection of Plasmodium infections in an endemic region of Colombia
por: Lina Zuluaga-Idárraga, et al.
Publicado: (2021) -
Smartphone-based turbidity reader
por: Hatice Ceylan Koydemir, et al.
Publicado: (2019) -
The Arctic Nearshore Turbidity Algorithm (ANTA) - A multi sensor turbidity algorithm for Arctic nearshore environments
por: Konstantin P. Klein, et al.
Publicado: (2021) -
Turbidity maximum zone index: a novel model for remote extraction of the turbidity maximum zone in different estuaries
por: C. Wang, et al.
Publicado: (2021)