Roentgen image recording on sillenite-type electrooptical crystals by electrophotographic method

To record Roentgen image on sillenite-type crystals while charging in a corona discharge field, a device has been performed. The resolution not less than 10 pairs of lines per mm, the sensitivity of 25 reciprocal Roentgens, the contrast coefficient not less than 2 at the exposure le...

Description complète

Enregistré dans:
Détails bibliographiques
Auteurs principaux: Dementiev, Igor, Maximov, Eugen, Pocatilov, Evghenii, Taracanova, Larisa
Format: article
Langue:EN
Publié: D.Ghitu Institute of Electronic Engineering and Nanotechnologies 2005
Sujets:
Accès en ligne:https://doaj.org/article/d7d3ab383ddd45ca9c9bc6c0aad9dbe8
Tags: Ajouter un tag
Pas de tags, Soyez le premier à ajouter un tag!
Description
Résumé:To record Roentgen image on sillenite-type crystals while charging in a corona discharge field, a device has been performed. The resolution not less than 10 pairs of lines per mm, the sensitivity of 25 reciprocal Roentgens, the contrast coefficient not less than 2 at the exposure length of 0.6 s were obtained using Bi12GeO20 crystal. The device survives more than 104 of writing-erase cycles. From the sensitivity point of view, the considered registration is just as good as X-Ray films used for radiography and dosimetry.