Tissue effects in vessel sealing and transection from an ultrasonic device with more intelligent control of energy delivery

Duan Broughton,1 Alissa L Welling,1 Emily H Monroe,1 Kristen Pirozzi,2 John B Schulte,1 Jeffrey W Clymer1 1Ethicon Endo-Surgery, Inc., Cincinnati, OH, 2University of Central Florida, Orlando, FL, USA Background: Ultrasonic surgical devices have been demonstrated to provide excellent hemostasis, effi...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Broughton D, Welling AL, Monroe EH, Pirozzi K, Schulte JB, Clymer JW
Formato: article
Lenguaje:EN
Publicado: Dove Medical Press 2013
Materias:
Acceso en línea:https://doaj.org/article/e2692bc6f7294b758d6d33e1781c9393
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:e2692bc6f7294b758d6d33e1781c9393
record_format dspace
spelling oai:doaj.org-article:e2692bc6f7294b758d6d33e1781c93932021-12-02T02:43:52ZTissue effects in vessel sealing and transection from an ultrasonic device with more intelligent control of energy delivery1179-1470https://doaj.org/article/e2692bc6f7294b758d6d33e1781c93932013-09-01T00:00:00Zhttp://www.dovepress.com/tissue-effects-in-vessel-sealing-and-transection-from-an-ultrasonic-de-a14368https://doaj.org/toc/1179-1470Duan Broughton,1 Alissa L Welling,1 Emily H Monroe,1 Kristen Pirozzi,2 John B Schulte,1 Jeffrey W Clymer1 1Ethicon Endo-Surgery, Inc., Cincinnati, OH, 2University of Central Florida, Orlando, FL, USA Background: Ultrasonic surgical devices have been demonstrated to provide excellent hemostasis, efficient transection, minimal lateral thermal damage, low smoke generation, and no risk of electrical current passage to the patient. These benefits originate from the inherent characteristics of the ultrasonic mechanism, but further improvements were thought possible through optimization of the energy delivery during application. The study reported here compared a new ultrasonic device, Harmonic ACE®+ Shears with Adaptive Tissue Technology, with a commercial predicate device, Harmonic ACE® Shears (both Ethicon Endo-Surgery, Inc., Cincinnati, OH, USA). Methods: Devices were evaluated in an in vivo porcine model intraoperatively and after a 30-day survival period. Both devices were used to seal a variety of vessels 1–5 mm in diameter, and compared for hemostasis, histological thermal damage, and adhesion formation. Sealed vessels were evaluated ex vivo for burst pressure, and visual obstruction caused by plumage from device application was assessed quantitatively. Results: ACE+ produced significantly less thermal damage, had fewer adhesions, offered faster transection, involved less visual obstruction, and had higher burst pressures than the predicate device. All vessel seals, evaluated over the course of a 30-day survival period in pigs, remained intact. Conclusion: The new Adaptive Tissue Technology algorithm assists the surgeon in achieving better control of energy delivery to the tissue, sealing vessels with supra-physiological burst pressures, and low thermal damage. These preclinical results with Harmonic ACE+ may translate into meaningful clinical benefits, providing greater precision along with more efficient and effective cutting and coagulation in open or laparoscopic procedures. Keywords: ultrasonic, Harmonic, ACE, vessel sealing, burst pressure, adhesions, Adaptive Tissue TechnologyBroughton DWelling ALMonroe EHPirozzi KSchulte JBClymer JWDove Medical PressarticleMedical technologyR855-855.5ENMedical Devices: Evidence and Research, Vol 2013, Iss default, Pp 151-154 (2013)
institution DOAJ
collection DOAJ
language EN
topic Medical technology
R855-855.5
spellingShingle Medical technology
R855-855.5
Broughton D
Welling AL
Monroe EH
Pirozzi K
Schulte JB
Clymer JW
Tissue effects in vessel sealing and transection from an ultrasonic device with more intelligent control of energy delivery
description Duan Broughton,1 Alissa L Welling,1 Emily H Monroe,1 Kristen Pirozzi,2 John B Schulte,1 Jeffrey W Clymer1 1Ethicon Endo-Surgery, Inc., Cincinnati, OH, 2University of Central Florida, Orlando, FL, USA Background: Ultrasonic surgical devices have been demonstrated to provide excellent hemostasis, efficient transection, minimal lateral thermal damage, low smoke generation, and no risk of electrical current passage to the patient. These benefits originate from the inherent characteristics of the ultrasonic mechanism, but further improvements were thought possible through optimization of the energy delivery during application. The study reported here compared a new ultrasonic device, Harmonic ACE®+ Shears with Adaptive Tissue Technology, with a commercial predicate device, Harmonic ACE® Shears (both Ethicon Endo-Surgery, Inc., Cincinnati, OH, USA). Methods: Devices were evaluated in an in vivo porcine model intraoperatively and after a 30-day survival period. Both devices were used to seal a variety of vessels 1–5 mm in diameter, and compared for hemostasis, histological thermal damage, and adhesion formation. Sealed vessels were evaluated ex vivo for burst pressure, and visual obstruction caused by plumage from device application was assessed quantitatively. Results: ACE+ produced significantly less thermal damage, had fewer adhesions, offered faster transection, involved less visual obstruction, and had higher burst pressures than the predicate device. All vessel seals, evaluated over the course of a 30-day survival period in pigs, remained intact. Conclusion: The new Adaptive Tissue Technology algorithm assists the surgeon in achieving better control of energy delivery to the tissue, sealing vessels with supra-physiological burst pressures, and low thermal damage. These preclinical results with Harmonic ACE+ may translate into meaningful clinical benefits, providing greater precision along with more efficient and effective cutting and coagulation in open or laparoscopic procedures. Keywords: ultrasonic, Harmonic, ACE, vessel sealing, burst pressure, adhesions, Adaptive Tissue Technology
format article
author Broughton D
Welling AL
Monroe EH
Pirozzi K
Schulte JB
Clymer JW
author_facet Broughton D
Welling AL
Monroe EH
Pirozzi K
Schulte JB
Clymer JW
author_sort Broughton D
title Tissue effects in vessel sealing and transection from an ultrasonic device with more intelligent control of energy delivery
title_short Tissue effects in vessel sealing and transection from an ultrasonic device with more intelligent control of energy delivery
title_full Tissue effects in vessel sealing and transection from an ultrasonic device with more intelligent control of energy delivery
title_fullStr Tissue effects in vessel sealing and transection from an ultrasonic device with more intelligent control of energy delivery
title_full_unstemmed Tissue effects in vessel sealing and transection from an ultrasonic device with more intelligent control of energy delivery
title_sort tissue effects in vessel sealing and transection from an ultrasonic device with more intelligent control of energy delivery
publisher Dove Medical Press
publishDate 2013
url https://doaj.org/article/e2692bc6f7294b758d6d33e1781c9393
work_keys_str_mv AT broughtond tissueeffectsinvesselsealingandtransectionfromanultrasonicdevicewithmoreintelligentcontrolofenergydelivery
AT wellingal tissueeffectsinvesselsealingandtransectionfromanultrasonicdevicewithmoreintelligentcontrolofenergydelivery
AT monroeeh tissueeffectsinvesselsealingandtransectionfromanultrasonicdevicewithmoreintelligentcontrolofenergydelivery
AT pirozzik tissueeffectsinvesselsealingandtransectionfromanultrasonicdevicewithmoreintelligentcontrolofenergydelivery
AT schultejb tissueeffectsinvesselsealingandtransectionfromanultrasonicdevicewithmoreintelligentcontrolofenergydelivery
AT clymerjw tissueeffectsinvesselsealingandtransectionfromanultrasonicdevicewithmoreintelligentcontrolofenergydelivery
_version_ 1718402255554609152