Corneal Curvature Measurements Utilizing a New Swept-source Optical Coherence Tomography Tomey OA-2000® and Comparison With IOL Master® 500 in Pterygium Patients

Introduction: Corneal curvature (CC) is an important anterior segment parameter. This study compared CC measurements conducted with two optical devices in pterygium eyes. Methods: Sixty pterygium eyes of 30 patients were enrolled in this study. CC was measured three times with the optical biometer...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Mohd Radzi Hilmi, Khairidzan Mohd Kamal, Mohd Zulfaezal Che Azemin, Mohd Hafidz Ithnin, Md Muziman Syah Md Mustafa, Firdaus Yusof@Alias, Norsham Ahmad
Formato: article
Lenguaje:EN
Publicado: Universitas Sultan Agung Semarang 2018
Materias:
Acceso en línea:https://doaj.org/article/f3cae608de2349bb8c53db3c718fec9f
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:f3cae608de2349bb8c53db3c718fec9f
record_format dspace
spelling oai:doaj.org-article:f3cae608de2349bb8c53db3c718fec9f2021-11-25T13:16:01ZCorneal Curvature Measurements Utilizing a New Swept-source Optical Coherence Tomography Tomey OA-2000® and Comparison With IOL Master® 500 in Pterygium Patients2085-15452339-093X10.26532/sainsmed.v9i1.2918https://doaj.org/article/f3cae608de2349bb8c53db3c718fec9f2018-11-01T00:00:00Zhttp://jurnal.unissula.ac.id/index.php/sainsmedika/article/view/2918https://doaj.org/toc/2085-1545https://doaj.org/toc/2339-093XIntroduction: Corneal curvature (CC) is an important anterior segment parameter. This study compared CC measurements conducted with two optical devices in pterygium eyes. Methods: Sixty pterygium eyes of 30 patients were enrolled in this study. CC was measured three times with the optical biometer and topography-keratometer Tomey OA-2000 (Tomey Corporation, Nagoya, Japan), then with partial optical coherence interferometry (PCI) IOL Master 500 (Carl Zeiss Meditec, Dublin, CA) and data were statistically analysed. Results: The measurements revealed in a mean CC of 43.86 ± 1.57 D with Tomey OA-2000 and 43.84 ± 1.55 D with IOL Master. Distribution of data is normal, and no significance difference in CC values was detected (P = 0.952) between the two devices. Correlation between CC measurements was highly significant (r = 0. 99; P < 0.0001). The mean difference of CC values between devices was 0.017 D and 95% limit of agreement was -0.088 to 0.12. Duration taken for measurements with the standard biometer IOL Master was longer (55.17 ± 2.24 seconds) than with Tomey OA-2000 (39.88 ± 2.38 seconds) in automatic mode. Duration of manual measurement with Tomey OA-2000 in manual mode was shorter (28.57 ± 2.71 seconds). Conclusion: In pterygium eyes, CC measured with Tomey OA-2000 and IOL Master showed similar values, and high correlation was observed between these two devices. This shows that both devices can be used interchangeably. Tomey OA-2000 is better in terms of faster to operate and has its own topography systems.Mohd Radzi HilmiKhairidzan Mohd KamalMohd Zulfaezal Che AzeminMohd Hafidz IthninMd Muziman Syah Md MustafaFirdaus Yusof@AliasNorsham AhmadUniversitas Sultan Agung Semarangarticlepterygiumcorneal curvatureiol master 500tomey oa-2000corneal topographyMedicine (General)R5-920ENSains Medika, Vol 9, Iss 1 (2018)
institution DOAJ
collection DOAJ
language EN
topic pterygium
corneal curvature
iol master 500
tomey oa-2000
corneal topography
Medicine (General)
R5-920
spellingShingle pterygium
corneal curvature
iol master 500
tomey oa-2000
corneal topography
Medicine (General)
R5-920
Mohd Radzi Hilmi
Khairidzan Mohd Kamal
Mohd Zulfaezal Che Azemin
Mohd Hafidz Ithnin
Md Muziman Syah Md Mustafa
Firdaus Yusof@Alias
Norsham Ahmad
Corneal Curvature Measurements Utilizing a New Swept-source Optical Coherence Tomography Tomey OA-2000® and Comparison With IOL Master® 500 in Pterygium Patients
description Introduction: Corneal curvature (CC) is an important anterior segment parameter. This study compared CC measurements conducted with two optical devices in pterygium eyes. Methods: Sixty pterygium eyes of 30 patients were enrolled in this study. CC was measured three times with the optical biometer and topography-keratometer Tomey OA-2000 (Tomey Corporation, Nagoya, Japan), then with partial optical coherence interferometry (PCI) IOL Master 500 (Carl Zeiss Meditec, Dublin, CA) and data were statistically analysed. Results: The measurements revealed in a mean CC of 43.86 ± 1.57 D with Tomey OA-2000 and 43.84 ± 1.55 D with IOL Master. Distribution of data is normal, and no significance difference in CC values was detected (P = 0.952) between the two devices. Correlation between CC measurements was highly significant (r = 0. 99; P < 0.0001). The mean difference of CC values between devices was 0.017 D and 95% limit of agreement was -0.088 to 0.12. Duration taken for measurements with the standard biometer IOL Master was longer (55.17 ± 2.24 seconds) than with Tomey OA-2000 (39.88 ± 2.38 seconds) in automatic mode. Duration of manual measurement with Tomey OA-2000 in manual mode was shorter (28.57 ± 2.71 seconds). Conclusion: In pterygium eyes, CC measured with Tomey OA-2000 and IOL Master showed similar values, and high correlation was observed between these two devices. This shows that both devices can be used interchangeably. Tomey OA-2000 is better in terms of faster to operate and has its own topography systems.
format article
author Mohd Radzi Hilmi
Khairidzan Mohd Kamal
Mohd Zulfaezal Che Azemin
Mohd Hafidz Ithnin
Md Muziman Syah Md Mustafa
Firdaus Yusof@Alias
Norsham Ahmad
author_facet Mohd Radzi Hilmi
Khairidzan Mohd Kamal
Mohd Zulfaezal Che Azemin
Mohd Hafidz Ithnin
Md Muziman Syah Md Mustafa
Firdaus Yusof@Alias
Norsham Ahmad
author_sort Mohd Radzi Hilmi
title Corneal Curvature Measurements Utilizing a New Swept-source Optical Coherence Tomography Tomey OA-2000® and Comparison With IOL Master® 500 in Pterygium Patients
title_short Corneal Curvature Measurements Utilizing a New Swept-source Optical Coherence Tomography Tomey OA-2000® and Comparison With IOL Master® 500 in Pterygium Patients
title_full Corneal Curvature Measurements Utilizing a New Swept-source Optical Coherence Tomography Tomey OA-2000® and Comparison With IOL Master® 500 in Pterygium Patients
title_fullStr Corneal Curvature Measurements Utilizing a New Swept-source Optical Coherence Tomography Tomey OA-2000® and Comparison With IOL Master® 500 in Pterygium Patients
title_full_unstemmed Corneal Curvature Measurements Utilizing a New Swept-source Optical Coherence Tomography Tomey OA-2000® and Comparison With IOL Master® 500 in Pterygium Patients
title_sort corneal curvature measurements utilizing a new swept-source optical coherence tomography tomey oa-2000® and comparison with iol master® 500 in pterygium patients
publisher Universitas Sultan Agung Semarang
publishDate 2018
url https://doaj.org/article/f3cae608de2349bb8c53db3c718fec9f
work_keys_str_mv AT mohdradzihilmi cornealcurvaturemeasurementsutilizinganewsweptsourceopticalcoherencetomographytomeyoa2000andcomparisonwithiolmaster500inpterygiumpatients
AT khairidzanmohdkamal cornealcurvaturemeasurementsutilizinganewsweptsourceopticalcoherencetomographytomeyoa2000andcomparisonwithiolmaster500inpterygiumpatients
AT mohdzulfaezalcheazemin cornealcurvaturemeasurementsutilizinganewsweptsourceopticalcoherencetomographytomeyoa2000andcomparisonwithiolmaster500inpterygiumpatients
AT mohdhafidzithnin cornealcurvaturemeasurementsutilizinganewsweptsourceopticalcoherencetomographytomeyoa2000andcomparisonwithiolmaster500inpterygiumpatients
AT mdmuzimansyahmdmustafa cornealcurvaturemeasurementsutilizinganewsweptsourceopticalcoherencetomographytomeyoa2000andcomparisonwithiolmaster500inpterygiumpatients
AT firdausyusofalias cornealcurvaturemeasurementsutilizinganewsweptsourceopticalcoherencetomographytomeyoa2000andcomparisonwithiolmaster500inpterygiumpatients
AT norshamahmad cornealcurvaturemeasurementsutilizinganewsweptsourceopticalcoherencetomographytomeyoa2000andcomparisonwithiolmaster500inpterygiumpatients
_version_ 1718413535203033088