Can the AHCL System Be Used in T1D Patients with Borderline TDDI? A Case Report
(1) Background: Intensive insulin therapy using continuous subcutaneous insulin infusion (CSII) with continuous real-time glucose monitoring (rt CGM) is the best option for patients with T1D. The recent introduction of a technology called Advanced Hybrid Closed Loop (AHCL) represents a new era in th...
Guardado en:
Autores principales: | , , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
MDPI AG
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/fa5b1e3b8d2746118b64fb07ba60670d |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
Sumario: | (1) Background: Intensive insulin therapy using continuous subcutaneous insulin infusion (CSII) with continuous real-time glucose monitoring (rt CGM) is the best option for patients with T1D. The recent introduction of a technology called Advanced Hybrid Closed Loop (AHCL) represents a new era in the treatment of type 1 diabetes, the next step towards better care, as well as improving the effectiveness and safety of therapy. The aim is to present the case of a T1D patient with a borderline total daily dose of insulin being treated with the Medtronic AHCL system in automatic mode. (2) Materials and Methods: A 9-year-old boy, from October 2020, with type 1 diabetes in remission was connected to the Minimed™ 780G (AHCL) system in accordance with the manufacturer’s recommendations (daily insulin dose > 8 units, age > 7). Records of the patient’s history were collected from visits to The Department of Children’s Diabetology, as well as from the Medtronic CareLink™ software and the DPV SWEET program from October 2020 to April 2021. (3) Results: The patient’s total daily insulin requirement decreased in the first 6 weeks after the AHCL was connected, which may reflect the remission phase (tight glycemic control with a healthy lifestyle). The lowest daily insulin requirement of 5.7 units was also recorded. In a three-month follow-up of the patient treated with AHCL, it was found that for almost 38% of the days the insulin dose was less than 8 IU. (4) Conclusions: The AHCL system allows safe and effective insulin therapy in automatic mode, as well as in patients with a lower daily insulin requirement. The AHCL system should be considered a good therapeutic option for patients from the onset of T1D, as well in the remission phase. |
---|