Optical bench performance of a novel trifocal intraocular lens compared with a multifocal intraocular lens

Shinwook Lee, Myoung Choi, Zaiwei Xu, Zeyu Zhao, Elsinore Alexander, Yueai Liu Alcon Research, Ltd., Fort Worth, TX, USA Purpose: The purpose of this study is to compare the optical characteristics of the novel PanOptix presbyopia-correcting trifocal intraocular lens (IOL) and the multifocal ReSTO...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Lee S, Choi M, Xu ZW, Zhao Z, Alexander E, Liu Y
Formato: article
Lenguaje:EN
Publicado: Dove Medical Press 2016
Materias:
Acceso en línea:https://doaj.org/article/fc4446565d9e492d971fe2d873723cf4
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:doaj.org-article:fc4446565d9e492d971fe2d873723cf4
record_format dspace
spelling oai:doaj.org-article:fc4446565d9e492d971fe2d873723cf42021-12-02T02:56:32ZOptical bench performance of a novel trifocal intraocular lens compared with a multifocal intraocular lens1177-5483https://doaj.org/article/fc4446565d9e492d971fe2d873723cf42016-06-01T00:00:00Zhttps://www.dovepress.com/optical-bench-performance-of-a-novel-trifocal-intraocular-lens-compare-peer-reviewed-article-OPTHhttps://doaj.org/toc/1177-5483Shinwook Lee, Myoung Choi, Zaiwei Xu, Zeyu Zhao, Elsinore Alexander, Yueai Liu Alcon Research, Ltd., Fort Worth, TX, USA Purpose: The purpose of this study is to compare the optical characteristics of the novel PanOptix presbyopia-correcting trifocal intraocular lens (IOL) and the multifocal ReSTOR +3.0 D IOL, through in vitro bench investigations. Methods: The optical characteristics of AcrySof® IQ PanOptix™ (PanOptix) and AcrySof® IQ ReSTOR +3.0 D (ReSTOR +3.0 D) IOLs were evaluated by through-focus Badal images, simulated headlight images, and modulation transfer function (MTF) measurements which determine resolution, photic phenomena, and image quality. Through-focus Badal images of an Early Treatment of Diabetic Retinopathy Study chart were recorded at both photopic and mesopic pupil sizes. Simulated headlight images were taken on an MTF bench with a 50-µm pinhole target and a 5.0 mm pupil at the distance focus of the IOL. MTF curves were measured with a 3.0 mm pupil, and spatial frequencies equivalent to 20/40 and 20/20 visual acuities were recorded to illustrate the through-focus MTF curves. Far-, intermediate-, and near-focus MTF values were obtained. Results: Bench Badal image testing and MTF measurements showed that PanOptix has a near focus at a distance of 42 cm and an additional intermediate focus at a distance of about 60 cm. The near focus for ReSTOR +3.0 D is at 45 cm. PanOptix and ReSTOR +3.0 D have comparable photopic distances and near MTF values. Additionally, PanOptix provided a substantial continuous range of vision from distance to intermediate and to near compared with ReSTOR +3.0 D. The halo propensity for PanOptix was slightly higher than that for ReSTOR +3.0 D. Conclusion: Laboratory-based in vitro simulations showed that PanOptix trifocal IOL has comparable resolution and image quality performance in distance and near foci compared with ReSTOR +3.0 D IOL. PanOptix showed better resolution and image quality performance at the intermediate focus than ReSTOR +3.0 D IOL. Keywords: multifocal, trifocal, modulation transfer function, Badal image, visual acuity, headlight imagesLee SChoi MXu ZWZhao ZAlexander ELiu YDove Medical Pressarticlemultifocaltrifocalmodulation transfer functionBadal imagevisual acuityheadlight imagesOphthalmologyRE1-994ENClinical Ophthalmology, Vol 2016, Iss Issue 1, Pp 1031-1038 (2016)
institution DOAJ
collection DOAJ
language EN
topic multifocal
trifocal
modulation transfer function
Badal image
visual acuity
headlight images
Ophthalmology
RE1-994
spellingShingle multifocal
trifocal
modulation transfer function
Badal image
visual acuity
headlight images
Ophthalmology
RE1-994
Lee S
Choi M
Xu ZW
Zhao Z
Alexander E
Liu Y
Optical bench performance of a novel trifocal intraocular lens compared with a multifocal intraocular lens
description Shinwook Lee, Myoung Choi, Zaiwei Xu, Zeyu Zhao, Elsinore Alexander, Yueai Liu Alcon Research, Ltd., Fort Worth, TX, USA Purpose: The purpose of this study is to compare the optical characteristics of the novel PanOptix presbyopia-correcting trifocal intraocular lens (IOL) and the multifocal ReSTOR +3.0 D IOL, through in vitro bench investigations. Methods: The optical characteristics of AcrySof® IQ PanOptix™ (PanOptix) and AcrySof® IQ ReSTOR +3.0 D (ReSTOR +3.0 D) IOLs were evaluated by through-focus Badal images, simulated headlight images, and modulation transfer function (MTF) measurements which determine resolution, photic phenomena, and image quality. Through-focus Badal images of an Early Treatment of Diabetic Retinopathy Study chart were recorded at both photopic and mesopic pupil sizes. Simulated headlight images were taken on an MTF bench with a 50-µm pinhole target and a 5.0 mm pupil at the distance focus of the IOL. MTF curves were measured with a 3.0 mm pupil, and spatial frequencies equivalent to 20/40 and 20/20 visual acuities were recorded to illustrate the through-focus MTF curves. Far-, intermediate-, and near-focus MTF values were obtained. Results: Bench Badal image testing and MTF measurements showed that PanOptix has a near focus at a distance of 42 cm and an additional intermediate focus at a distance of about 60 cm. The near focus for ReSTOR +3.0 D is at 45 cm. PanOptix and ReSTOR +3.0 D have comparable photopic distances and near MTF values. Additionally, PanOptix provided a substantial continuous range of vision from distance to intermediate and to near compared with ReSTOR +3.0 D. The halo propensity for PanOptix was slightly higher than that for ReSTOR +3.0 D. Conclusion: Laboratory-based in vitro simulations showed that PanOptix trifocal IOL has comparable resolution and image quality performance in distance and near foci compared with ReSTOR +3.0 D IOL. PanOptix showed better resolution and image quality performance at the intermediate focus than ReSTOR +3.0 D IOL. Keywords: multifocal, trifocal, modulation transfer function, Badal image, visual acuity, headlight images
format article
author Lee S
Choi M
Xu ZW
Zhao Z
Alexander E
Liu Y
author_facet Lee S
Choi M
Xu ZW
Zhao Z
Alexander E
Liu Y
author_sort Lee S
title Optical bench performance of a novel trifocal intraocular lens compared with a multifocal intraocular lens
title_short Optical bench performance of a novel trifocal intraocular lens compared with a multifocal intraocular lens
title_full Optical bench performance of a novel trifocal intraocular lens compared with a multifocal intraocular lens
title_fullStr Optical bench performance of a novel trifocal intraocular lens compared with a multifocal intraocular lens
title_full_unstemmed Optical bench performance of a novel trifocal intraocular lens compared with a multifocal intraocular lens
title_sort optical bench performance of a novel trifocal intraocular lens compared with a multifocal intraocular lens
publisher Dove Medical Press
publishDate 2016
url https://doaj.org/article/fc4446565d9e492d971fe2d873723cf4
work_keys_str_mv AT lees opticalbenchperformanceofanoveltrifocalintraocularlenscomparedwithamultifocalintraocularlens
AT choim opticalbenchperformanceofanoveltrifocalintraocularlenscomparedwithamultifocalintraocularlens
AT xuzw opticalbenchperformanceofanoveltrifocalintraocularlenscomparedwithamultifocalintraocularlens
AT zhaoz opticalbenchperformanceofanoveltrifocalintraocularlenscomparedwithamultifocalintraocularlens
AT alexandere opticalbenchperformanceofanoveltrifocalintraocularlenscomparedwithamultifocalintraocularlens
AT liuy opticalbenchperformanceofanoveltrifocalintraocularlenscomparedwithamultifocalintraocularlens
_version_ 1718402079623479296