Experiencia en el uso de un nuevo catéter electrodo de estimulación endocavitaria: Seguimiento a un año

Background: Since February 2002, Flextend® electrode catheters are used at the Cardiovascular Unit of the Catholic University Clinical Hospital. These transvenous catheters have an IS-1 connector, silicone coating, active fixation and retractile helix with dexamethasone acetate. Aim: To report early...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Frangini S,Patricia, Vergara S,Ismael, González A,Rolando, Baeza L,Mariana, Cambón R,Ana María
Lenguaje:Spanish / Castilian
Publicado: Sociedad Médica de Santiago 2006
Materias:
Acceso en línea:http://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0034-98872006000600011
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
id oai:scielo:S0034-98872006000600011
record_format dspace
spelling oai:scielo:S0034-988720060006000112006-08-14Experiencia en el uso de un nuevo catéter electrodo de estimulación endocavitaria: Seguimiento a un añoFrangini S,PatriciaVergara S,IsmaelGonzález A,RolandoBaeza L,MarianaCambón R,Ana María Pacemaker artificial Electrodos, immplanted Background: Since February 2002, Flextend® electrode catheters are used at the Cardiovascular Unit of the Catholic University Clinical Hospital. These transvenous catheters have an IS-1 connector, silicone coating, active fixation and retractile helix with dexamethasone acetate. Aim: To report early and one year results using Flextend® catheters. Material and methods: Retrospective analysis of all patients that received an implant of a Flextend® catheter in the Unit. Results: Forty one Flextend® catheters were implanted in 24 patients, without acute displacement or clinical pericarditis. In 18 electrode catheters located in the right atrium the mean values for p wave, stimulation threshold and impedance at the moment of placement, were 2.9±1.4 mV, 0.8±0.4 V and 522±86 Ohms, respectively. The figures 24 hours after placement were 3.6±2.1 mV, 0.8±0.3 V and 612±69 Ohms, respectively. In 23 electrode catheters in the right ventricle, mean values at the moment of placement for R wave, stimulation threshold and impedance were 11.3±3.6 mV, 0.8±0.2 V and 756±108 Ohms, respectively. The figures 24 hours after placement were 3.6±2.1 mV, 0.8±0.3 V and 612±69 Ohms, respectively. In one patient, the stimulation threshold increased two months after placement and required a new intervention. Conclusions: Flextend® catheters ca be placed successfully with a low rate of complications and stable function parameters on follow upinfo:eu-repo/semantics/openAccessSociedad Médica de SantiagoRevista médica de Chile v.134 n.6 20062006-06-01text/htmlhttp://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0034-98872006000600011es10.4067/S0034-98872006000600011
institution Scielo Chile
collection Scielo Chile
language Spanish / Castilian
topic Pacemaker
artificial
Electrodos, immplanted
spellingShingle Pacemaker
artificial
Electrodos, immplanted
Frangini S,Patricia
Vergara S,Ismael
González A,Rolando
Baeza L,Mariana
Cambón R,Ana María
Experiencia en el uso de un nuevo catéter electrodo de estimulación endocavitaria: Seguimiento a un año
description Background: Since February 2002, Flextend® electrode catheters are used at the Cardiovascular Unit of the Catholic University Clinical Hospital. These transvenous catheters have an IS-1 connector, silicone coating, active fixation and retractile helix with dexamethasone acetate. Aim: To report early and one year results using Flextend® catheters. Material and methods: Retrospective analysis of all patients that received an implant of a Flextend® catheter in the Unit. Results: Forty one Flextend® catheters were implanted in 24 patients, without acute displacement or clinical pericarditis. In 18 electrode catheters located in the right atrium the mean values for p wave, stimulation threshold and impedance at the moment of placement, were 2.9±1.4 mV, 0.8±0.4 V and 522±86 Ohms, respectively. The figures 24 hours after placement were 3.6±2.1 mV, 0.8±0.3 V and 612±69 Ohms, respectively. In 23 electrode catheters in the right ventricle, mean values at the moment of placement for R wave, stimulation threshold and impedance were 11.3±3.6 mV, 0.8±0.2 V and 756±108 Ohms, respectively. The figures 24 hours after placement were 3.6±2.1 mV, 0.8±0.3 V and 612±69 Ohms, respectively. In one patient, the stimulation threshold increased two months after placement and required a new intervention. Conclusions: Flextend® catheters ca be placed successfully with a low rate of complications and stable function parameters on follow up
author Frangini S,Patricia
Vergara S,Ismael
González A,Rolando
Baeza L,Mariana
Cambón R,Ana María
author_facet Frangini S,Patricia
Vergara S,Ismael
González A,Rolando
Baeza L,Mariana
Cambón R,Ana María
author_sort Frangini S,Patricia
title Experiencia en el uso de un nuevo catéter electrodo de estimulación endocavitaria: Seguimiento a un año
title_short Experiencia en el uso de un nuevo catéter electrodo de estimulación endocavitaria: Seguimiento a un año
title_full Experiencia en el uso de un nuevo catéter electrodo de estimulación endocavitaria: Seguimiento a un año
title_fullStr Experiencia en el uso de un nuevo catéter electrodo de estimulación endocavitaria: Seguimiento a un año
title_full_unstemmed Experiencia en el uso de un nuevo catéter electrodo de estimulación endocavitaria: Seguimiento a un año
title_sort experiencia en el uso de un nuevo catéter electrodo de estimulación endocavitaria: seguimiento a un año
publisher Sociedad Médica de Santiago
publishDate 2006
url http://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0034-98872006000600011
work_keys_str_mv AT franginispatricia experienciaenelusodeunnuevocateterelectrododeestimulacionendocavitariaseguimientoaunano
AT vergarasismael experienciaenelusodeunnuevocateterelectrododeestimulacionendocavitariaseguimientoaunano
AT gonzalezarolando experienciaenelusodeunnuevocateterelectrododeestimulacionendocavitariaseguimientoaunano
AT baezalmariana experienciaenelusodeunnuevocateterelectrododeestimulacionendocavitariaseguimientoaunano
AT cambonranamaria experienciaenelusodeunnuevocateterelectrododeestimulacionendocavitariaseguimientoaunano
_version_ 1718436266188472320