A SILVER(I) PVC-MEMBRANE SENSOR BASED ON SYNTHESIZED DILAKTAM CROWN ETHER
In this work, we introduce a highly selective and sensitive silver(I) PVC-membrane sensor. Dilaktam Crown ether (DLCE) was used as a membrane-active component to prepare a highly sensitive Ag(I)-selective polymeric membrane electrode. This sensor illustrated very good selectivity and sensitivity tow...
Guardado en:
Autores principales: | , , , , , |
---|---|
Lenguaje: | English |
Publicado: |
Sociedad Chilena de Química
2009
|
Materias: | |
Acceso en línea: | http://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0717-97072009000100015 |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
Sumario: | In this work, we introduce a highly selective and sensitive silver(I) PVC-membrane sensor. Dilaktam Crown ether (DLCE) was used as a membrane-active component to prepare a highly sensitive Ag(I)-selective polymeric membrane electrode. This sensor illustrated very good selectivity and sensitivity towards silver ions over a wide variety of cations, including alkali, alkaline earth, transition and heavy metal ions. The sensor exhibited a Nernstian behavior (with a slope of 59.8 ± 0.2 mV per decade) for a concentration range (1.0 x 10-5-1.0 x 10¹ M) with a detection limit of 6.8xl0-6 M. It displayed a response time in the whole concentration range (~20 s) and its usage exceeded a 75 days period in the pH range of 5.1-7.2. The proposed electrode application was found to be successful as an indicator electrode in the titration with NaCl. |
---|