STRESS DEGRADATION STUDIES OF DRONEDARONE IN PHARMACEUTICAL DOSAGE FORM BY A VALIDATED STABILITY-INDICATING LC METHOD

A simple, rapid, and precise method is developed for the quantitative estimation of dronedarone hydrochloride in presence of its degradants which are formed due to stress conditions. A chromatographic separation of the dronedarone and its degradants were achieved with Zorbax XDB C8 150 x 4.6 mm, 3.5...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: AHIRRAO,V. K, PATIL,C. S, BEMBALKAR,S. R, KATARIYA,M. V, SONNEKAR,V. S, MARATHE,R. P, NAWALE,R. B, PAWAR,R. P
Lenguaje:English
Publicado: Sociedad Chilena de Química 2012
Materias:
Acceso en línea:http://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0717-97072012000300016
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
Descripción
Sumario:A simple, rapid, and precise method is developed for the quantitative estimation of dronedarone hydrochloride in presence of its degradants which are formed due to stress conditions. A chromatographic separation of the dronedarone and its degradants were achieved with Zorbax XDB C8 150 x 4.6 mm, 3.5µm analytical column using aqueous solution of 0.05M ammonium acetate and acetonitrile (40:60 v/v). The instrumental settings are flow rate of 1.0 mL/min, column temperature at 30ºC, and detector wavelength of 220 nm using a photodiode array detector. Observed theoretical plates and tailing factor for dronedarone were 6577 and 1.09 respectively. Dronedarone was exposed to thermal, photolytic, hydrolytic and oxidative stress conditions, and the stressed samples were analysed by the proposed method. Peak homogeneity data of dronedarone is obtained using photodiode array detector. The stressed sample chromatograms demonstrate the specificity of the method for the estimation dronedarone in presence of its degradants. The described method showed excellent linearity over a range of 10 - 300 µg/mL for dronedarone. The correlation coefficient is 0.999. The relative standard deviation of peak area for six measurements is always found to be less than 2%. The proposed method was found to be suitable and accurate for quantitative determination and stability study of dronedarone.