INCLUSION OF GLUTATHIONE PEROXIDASE ON POLYPYRROLE MODIFIED ELECTRODE, TO BE ASSAYED AS GLUTATHIONE SENSOR

In this work it was obtained a modified electrode by inclusion of the enzyme glutathione peroxidase (GPx) on a conductive polymer, polypyrrole (PPy), deposited on a graphite electrode (GC|PPy,GPx) for the amperometric detection of reduced glutathione (GSH) in presence of hydrogen peroxide. Also, wor...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: del VALLE,M. A, CHEN,Y-.Y, ALARCÓN,A, RAMOS,A, HERNÁNDEZ,L. A, CANALES,C, GONZÁLEZ,B
Lenguaje:English
Publicado: Sociedad Chilena de Química 2013
Materias:
Acceso en línea:http://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0717-97072013000400018
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
Descripción
Sumario:In this work it was obtained a modified electrode by inclusion of the enzyme glutathione peroxidase (GPx) on a conductive polymer, polypyrrole (PPy), deposited on a graphite electrode (GC|PPy,GPx) for the amperometric detection of reduced glutathione (GSH) in presence of hydrogen peroxide. Also, working conditions were optimized in order to prepare the electrode GC|PPy,GPx, which were tested for different concentrations of monomer and enzyme, electrochemical type of perturbance (by fixed potential or cyclic voltammetry) deposition potential and time (PPy film thickness, GPx), etc. by testing the response of the electrodes versus GSH in presence of hydrogen peroxide at physiological pH and temperature. At optimal experimental conditions, the achieved electrode gives a characteristic response for the analyte, which means that this modified electrode can be proposed as a selective GSH sensor, since it was shown that the enzyme maintains its activity during experimental manipulations and responds in a highly reproducible mode, so its use is proposed as an amperometric biosensor which is highly selective and stable.