Is There a Difference in the Resonance Frequency Values Obtained Directly on the Implants and Abutments?
The aim was to evaluate the differences of implant stability quotient (ISQ) between implants with external hexagon and Morse taper connectors. The study had a split mouth design, composed by 10 patients who received hybrid protocol prosthesis. In total, 40 implants (3.75 x 13 mm) were installed: on...
Guardado en:
Autores principales: | , , , , , |
---|---|
Lenguaje: | English |
Publicado: |
Universidad de La Frontera. Facultad de Medicina
2015
|
Materias: | |
Acceso en línea: | http://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0718-381X2015000300020 |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
id |
oai:scielo:S0718-381X2015000300020 |
---|---|
record_format |
dspace |
spelling |
oai:scielo:S0718-381X20150003000202016-03-22Is There a Difference in the Resonance Frequency Values Obtained Directly on the Implants and Abutments?Pozzer,LeandroGabrielli,Marisa Aparecida CabriniTorres Neto,Renato AugustoSpin-Neto,RubensAsprino,LucianaPereira-Filho,Valfrido Antonio Implant stability implant stability quotient dental implant The aim was to evaluate the differences of implant stability quotient (ISQ) between implants with external hexagon and Morse taper connectors. The study had a split mouth design, composed by 10 patients who received hybrid protocol prosthesis. In total, 40 implants (3.75 x 13 mm) were installed: on the right side, 20 external hexagon, and on the left, 20 Morse taper. After two years in function, the stability test was applied by using the MRI machine Osstell ISQ directly on the implants and on the abutments. Considering the measurements made on the implants, there were differences between HE and CM mesial (p= 0.011), lingual (p= 0.003) and distal (p= 0.006). Considering the measurements made on the abutments, there were differences between HE and CM on the buccal (p= 0.020) and lingual (p= 0.004). The trend and higher values are for the CM group. The values for Morse taper implants were always higher in a statistically significant manner, when compared to the external hexagon.info:eu-repo/semantics/openAccessUniversidad de La Frontera. Facultad de MedicinaInternational journal of odontostomatology v.9 n.3 20152015-12-01text/htmlhttp://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0718-381X2015000300020en10.4067/S0718-381X2015000300020 |
institution |
Scielo Chile |
collection |
Scielo Chile |
language |
English |
topic |
Implant stability implant stability quotient dental implant |
spellingShingle |
Implant stability implant stability quotient dental implant Pozzer,Leandro Gabrielli,Marisa Aparecida Cabrini Torres Neto,Renato Augusto Spin-Neto,Rubens Asprino,Luciana Pereira-Filho,Valfrido Antonio Is There a Difference in the Resonance Frequency Values Obtained Directly on the Implants and Abutments? |
description |
The aim was to evaluate the differences of implant stability quotient (ISQ) between implants with external hexagon and Morse taper connectors. The study had a split mouth design, composed by 10 patients who received hybrid protocol prosthesis. In total, 40 implants (3.75 x 13 mm) were installed: on the right side, 20 external hexagon, and on the left, 20 Morse taper. After two years in function, the stability test was applied by using the MRI machine Osstell ISQ directly on the implants and on the abutments. Considering the measurements made on the implants, there were differences between HE and CM mesial (p= 0.011), lingual (p= 0.003) and distal (p= 0.006). Considering the measurements made on the abutments, there were differences between HE and CM on the buccal (p= 0.020) and lingual (p= 0.004). The trend and higher values are for the CM group. The values for Morse taper implants were always higher in a statistically significant manner, when compared to the external hexagon. |
author |
Pozzer,Leandro Gabrielli,Marisa Aparecida Cabrini Torres Neto,Renato Augusto Spin-Neto,Rubens Asprino,Luciana Pereira-Filho,Valfrido Antonio |
author_facet |
Pozzer,Leandro Gabrielli,Marisa Aparecida Cabrini Torres Neto,Renato Augusto Spin-Neto,Rubens Asprino,Luciana Pereira-Filho,Valfrido Antonio |
author_sort |
Pozzer,Leandro |
title |
Is There a Difference in the Resonance Frequency Values Obtained Directly on the Implants and Abutments? |
title_short |
Is There a Difference in the Resonance Frequency Values Obtained Directly on the Implants and Abutments? |
title_full |
Is There a Difference in the Resonance Frequency Values Obtained Directly on the Implants and Abutments? |
title_fullStr |
Is There a Difference in the Resonance Frequency Values Obtained Directly on the Implants and Abutments? |
title_full_unstemmed |
Is There a Difference in the Resonance Frequency Values Obtained Directly on the Implants and Abutments? |
title_sort |
is there a difference in the resonance frequency values obtained directly on the implants and abutments? |
publisher |
Universidad de La Frontera. Facultad de Medicina |
publishDate |
2015 |
url |
http://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0718-381X2015000300020 |
work_keys_str_mv |
AT pozzerleandro isthereadifferenceintheresonancefrequencyvaluesobtaineddirectlyontheimplantsandabutments AT gabriellimarisaaparecidacabrini isthereadifferenceintheresonancefrequencyvaluesobtaineddirectlyontheimplantsandabutments AT torresnetorenatoaugusto isthereadifferenceintheresonancefrequencyvaluesobtaineddirectlyontheimplantsandabutments AT spinnetorubens isthereadifferenceintheresonancefrequencyvaluesobtaineddirectlyontheimplantsandabutments AT asprinoluciana isthereadifferenceintheresonancefrequencyvaluesobtaineddirectlyontheimplantsandabutments AT pereirafilhovalfridoantonio isthereadifferenceintheresonancefrequencyvaluesobtaineddirectlyontheimplantsandabutments |
_version_ |
1714203973585469440 |